Anal. Chem. 2008, 80, 7194–7197
Secondary Ion Mass Spectrometry
Jennifer Griffiths
An old dog with some nifty new tricks.
Six years ago, things were looking bleak in the secondary ion
MS (SIMS) community. According to researchers, the field had
reached maturity, and not much was new. The AC SIMS
Product Review that was published around that time (2002,
74, 335 A-341 A) even began with the line, “If it ain’t broke,
don’t fix it.” Joe Gardella of the University at Buffalo, State
University of New York, says that shortly thereafter, at the 15th
International Conference on SIMS in 2003, many in the field
wondered if the technique had run out of novel applications.
“There was a palpable sense at that point that maybe the
conference couldn’t stand alone anymore,” he says. “That is,
maybe the technique was routine enough that you didn’t need
a separate conference to talk about it.”
Then, suddenly, everything turned around. “The field has
certainly undergone a seismic shift, both in intensity and direction,” says Nick Winograd of Pennsylvania State University. “The
seeds of it were there in 2002, but I don’t think many people were
quite prepared for the rapidity of the change.”
Two advances that were just beginning to be widely adopted
that year have taken over and rejuvenated the field: cluster ion
beams and TOFMS analyzers. SIMS traditionally has been limited
to atomic analyses of surfaces, but these two techniques in
combination provide a route to a long-sought-after goal in the
SIMS community: analysis of intact complex organic molecules.
In light of these changes, a Product Review update is in order.
The tables that accompany this article are divided into two
categories: instruments offering quadrupole or magnetic-sector
MS (Table 1) and those offering TOFMS (Table 2). These tables
are not meant to be comprehensive sources of product information; please contact the manufacturers for more details.
THE BASICS
SIMS is a surface analysis technique in which a beam of “primary
ions”straditionally Ar+, Ga+, or alkali metal ionssare shot at a
sample’s surface; the primary ions transfer their energy to
molecules on the surface and dislodge “secondary ions”. (Secondary ions are either positive or negative, depending on the primary
ions’ identity; neutral species are also ejected, but these cannot
be analyzed by MS unless more convoluted measures are taken.)
7194
Analytical Chemistry, Vol. 80, No. 19, October 1, 2008
“It’s rather like playing molecular pool,” says Fraser Reich of Kore
Technology (U.K.). “The cue ball goes into the material, and it
sputterssit lifts off material that’s characteristic of the top surface.”
Once the particles are in flight, an electric field directs them
into an MS analyzer. These come in a variety of flavors: at first,
quadrupole and magnetic-sector instruments dominated the field,
but more recently, TOF instruments have started to take over.
Each type of mass spectrometer has advantages and disadvantages. For example, according to John Eccles of Millbrook
(U.K.), a big advantage of a TOF system is that you can acquire
a full mass spectrum in one operation. “So, rather than having to
scan through the spectrum to build it up, you can acquire it
simultaneously, which means it’s much more efficient than a
quadrupole [or magnetic-sector] system for analyzing small areas.”
Parallel detection also means the sample needs to be hit less often
with the destructive primary ion beam; therefore, TOF SIMS is
much better suited to analyzing fragile complex organic molecules
than are the other techniques.
Quadrupole and magnetic-sector instruments, on the other
hand, are very good for looking at individual atoms. These
10.1021/ac801528u 2008 American Chemical Society
Published on Web 08/29/2008
Table 1. Selected quadrupole and magnetic-sector instruments.a
Product
Company
Approximate
price
(U.S.D.)
Applications
IMS 7f/NanoSIMS50
Cameca Instruments
208-442-6559
www.cameca.com
Contact vendor
Depth profiling, imaging,
layer characterization,
isotope ratios,
trace elements
Primary ion
source
Vacuum
chamber
features
O2, Cs
Mass
analyzer
Magnetic sector
Mass
range
Mass
resolving
power
Detector
0-360 u
Titanium sublimation pump
with ion or turbomolecular
pumps
MAXIM SIMS Workstation
MiniSIMS alpha
Hiden Analytical
www.hidenanalytical.com
Physical Electronics
952-828-6100
www.phi.com
Contact vendor
Depth profiling, imaging,
static SIMS, sputtered neutral
MS (SNMS), desorbed gas analysis,
thin films, glass coatings,
hard coatings, semiconductors,
cleanliness monitoring, failure
analysis, research, and production
control
O2, Cs, Ar, Xe
Static, imaging, or
dynamic SIMS
Semiconductors, glass,
photovoltaics, thin films
Ga+
Spherical, bakeable,
ultrahigh-vacuum (UHV)
chamber using
industry-standard components;
15 ports with spare ports
for customization or addition of
optional detectors;
ion or electron guns
MAXIM SIMS/SNMS analyzer
fitted with
9-mm-triple-filter
Hiden quadrupole
Optional: 1 to 300, 510,
or 1000 u
High-vacuum, automated
sample loading
Cs+ surface ionization source,
O2+ duoplasmatron source
Five-axis stage, three ion gun
ports, charge neutralization
Quadrupole
Quadrupole
3-300 u
1-340 u
Unit mass resolution
Unit mass resolution
Faraday cup for analog and
channel electron multiplier
for pulse counting
Windows-based PC instrument
control with dedicated
SIMetric SIMS depth-profiling
software
390,000
>20,000
Electron multiplier,
Faraday cup
Channeltron
Channeltron
Hardware/
software
PC with MS Windows-based
control and data
treatment software;
imaging software capabilities
Hiden MasSoft running on PC
provides visual build of
experiment flow;
data are easily exported to
standard processing packages
PC- and Windows-based
software for data acquisition,
display, and processing
Options
Charge compensation for
insulating samples; rotating
stage for rough analysis;
low-energy kit for excellent
depth resolution
Positive or negative secondary
ion analysis; low depth
resolution at high sensitivity
High sensitivity for light
elements; excellent lateral
resolution
Special
features
a
ADEPT-1010
Millbrook Instruments
+44-1254-699606
www.minisims.com
200,000
Highly flexible design to
permit a very wide
range of samples
Electron gun for charge
neutralization;
enhanced sample handling for
multiple samples;
spectral library
Tabletop design; automated
operation
Gallium liquid-metal ion gun
(LMIG)
Extremely high vacuum
(<8 × 1011 torr) enables low
detection limits for atmospherics
(C, O, H)
Low extraction field enables
easy charge neutralization
Contact vendors for full product lines.
instruments were developed mainly for use in the semiconductor
industry to detect small amounts of dopant in silicon chips. That
industry required high sensitivity and an instrument capable not
only of analyzing the top surface (static SIMS) but also of acquiring
a “depth profile” down into the sample (dynamic SIMS).
Graham Cooke of Hiden (U.K.) says that when used in
industry, magnetic-sector or quadrupole SIMS is usually one of
an array of surface techniques used to analyze a sample. “It’s not
often that you only use SIMS by itself to solve a problem,” he
says. “Gone are the days where people used to have one analysis
technique in each lab, and they wouldn’t talk to each other.”
Researchers now combine SIMS with X-ray photoelectron spectroscopy, electron microscopy, and perhaps transmission electron
microscopy. “No one technique has [all of] the depth resolution,
the sensitivity, and the lateral resolution that you could possibly
want,” he adds.
Some of these surface analysis techniques are now being
combined into one instrument, and Cooke points out that quadrupole analyzers are the best choice for this purpose. That’s
because they use a relatively weak electric field to funnel ions
into the mass spectrometer. (The stronger electric field of a TOF
instrument could, for example, deflect incoming electrons from
other techniques away from the target surface.)
WHAT’S NEW?
Although magnetic-sector and quadrupole instruments still definitely have important roles to play in industry, researchers in the
field say that the current “hot” SIMS research areas are cluster
ion sources and TOFMSsusually used in combination to look at
complex organic molecules. For example, the two technologies
have caused an explosion in biological SIMS research, says
Nathan Havercroft of ION-TOF. “Six years ago it was never even
Analytical Chemistry, Vol. 80, No. 19, October 1, 2008
7195
Table 2. Selected TOF instruments.a
Product
J105 Chemical
Imager
TOF.SIMS
5 Series
SurfaceSeer
Ionoptika
ION-TOF GmbH
Kore Technology
+44-2380-270-735
+49-251-1622140
+44-01353-653030
www.ionoptika.com
www.iontof.com
www.kore.co.uk
Approximate 1,500,000 (depending on
Contact vendor
290,000
price
configuration)
(U.S.D.)
Applications Static and dynamic SIMS,
Static or dynamic SIMS;
Surface MS; ideal for
3D analysis, identification
surface spectroscopy;
quality control or
and mapping of organic
ion mapping; depth profiling
troubleshooting for the
molecules
and 3D analysis in
performance of materials
fields such as materials
that will be coated,
characterization, polymers,
printed, or bonded
microelectronics,
together, where generally
catalysis, biomaterials,
the condition of the
and failure analysis;
surface of the items
basic research
is of critical
importance
Pulsed 5 keV inert gas
Bismuth field emission
Primary ion Gold LMIG, C60
(Ar+)
cluster ion source,
source
oxygen source,
cesium source, C60
Vacuum
Precision three-axis stage
Oil- and cooling-waterMultiple-port UHV
chamber
with heating and cooling;
free pumping system;
analysis vacuum chamber;
features
ports available for
horizontal and grounded
high-vacuum gauge
add-on techniques
sample stage configuration;
monitors the pressure in
sample size up to 300 mm;
the analytical chamber and
up to three ion columns
shuts down high voltages
with up to five sources can
if the pressure
be fitted
rises beyond a set
point
Mass
TOF with dual time
Nonlinear reflectron
TOF (reflectron)
analyzer
focus
analyzer, full mass
resolution at full
transmission
Mass
e5000 Da
1->10,000 u
>1000 m/z
range
Mass
>10,000
>12,000 at 29 u
1000 m/∆m
resolving
>18,000 at 300 u
power
Detector
Microchannel plate with
Secondary electron
Dual microchannel plate
preamplifier; secondary
detector and secondary
detector
electron detector
ion detectors:
microchannel
plate/scintillator/
photomultiplier
tube type
Hardware/
Full software control of
PC control using
GRAMS/AI software by
software
instrument systems;
Windows XP;
Thermo Scientific with
data acquisition
data-reduction
Kore Technology mass
and display
software for imaging,
spectral extensions
profiling, and 3D
analysis; high-massresolution spectral
library
Options
Bright duoplasmatron ion
Temperature-controlled
Multiple sample holder to
source, 30 kV or floating
heating and cooling in
permit analysis of more
low-energy ion gun
load-lock and analysis
than one sample per
chamber
sample load; optical
(-150 to +600 °C);
microscope/camera;
G-SIMS with dedicated
additional gun for
Bi/Mn emitter;
depth profiling
customized sample
preparation
chambers
Special
3D imaging of
All-purpose Bi source
Compact, very simple to
features
biomolecules inside
with high cluster
use; sample turnaround
cells, MS/MS
currents; modular design
5 min; positive and
open to retrofits;
negative SIMS; sputter
live sample viewing by
cleaning facility
two cameras; raw data
stored for retrospective
analysis
Company
a
MiniSIMS ToF
nanoTOF
Millbrook Instruments
+44-1254-699606
www.minisims.com
400,000
Physical Electronics
952-828-6100
www.phi.com
Contact vendor
Static, imaging, and
dynamic SIMS
Polymers,
semiconductors,
biomedical devices,
pharmaceuticals,
biological tissues,
minerals
Ga+
LMIG
High vacuum,
automated
sample loading
Five-axis stage, four ion
gun ports; charge
neutralization; air shock
vibration isolation
TOF reflectron
(secondary ion
beam pulsing)
TOF
1-1200
1-10,000 u
650 m/∆m
>10,000 m/∆m
Channel plate detector
Dual channel plate
PC- and Windows-based
software for data
acquisition, display,
and processing
Windows-based PC
instrument control with
dedicated WinCadence
TOF SIMS software for
acquiring spectra,
2D and 3D images,
and depth profiles
Electron gun for charge
Hot/cold stage, LMIG
neutralization; enhanced
gold cluster emitter,
sample handling
C60 ion gun, gas ion gun,
oxygen leak, Cs+ ion
for multiple samples;
spectral library
gun, 300 mm wafer stage
Tabletop design,
automated operation
Stage design shuttles
between intro chamber
and analysis position
automatically; air shock
vibration isolation for
high-magnification
imaging
Contact vendors for full product lines.
really talked about. There were a couple of people who were just
starting, but now we are seeing many groups that are involved in
SIMS on biological samplessit’s great!”
7196
Analytical Chemistry, Vol. 80, No. 19, October 1, 2008
Cluster ion sources are exactly what their name implies: rather
than a single atom as a primary ion, these charged particles are made
up of multiple atoms. “The reason that these clusters are so
important... is that [with] the other [monatomic] projectiles, if you
take a molecular surfacesa polymer, biomolecules, any kind of
organic moleculesand you hit it with these energetic beams, it blows
the molecules apart,” says Winograd. “You get some of them
desorbed that you can see in your mass spectrometer, but it leaves
a great deal of chemical damage behind. And eventually your surface
is converted into junk, and the signal goes away.” The cluster ion
beams, on the other hand, can lift molecules with molecular weights
of several thousand daltons off of the surface intact.
Currently, researchers can choose from four types of cluster ion
beams: several different types of Bi or Au polyatomic clusters, C60+,
and SF5+. Each type has advantages and disadvantages, and some
are used more commonly than others. For example, the metal
polyatomic ions are important for the emerging field of SIMS
imaging.
SIMS has been adapted for imaging in much the same way
that MALDI has been. “You can raster scan [the incoming primary
ion beam] over the surface, and in that way you can build up an
image of the distribution of a particular secondary ion generated
from the surface, which is essentially a chemical map,” explains
Eccles. “One of the advantages of SIMS is that you have excellent
S/N in the data, so you can do that process very quickly, and
you can generate images in literally just 5 or 10 seconds.” And
with a TOF analyzer, a single acquired image actually stores a
complete mass spectrum at each pixel, so multiple chemical maps
each showing an individual species can be generated retrospectively.
SIMS and MALDI imaging are quite complementary. For example, using SIMS, “our lateral resolution is much higher. We’re
down in the hundred nanometer range for imaging, whereas MALDI
is still stuck in the tens or twenties of microns,” says Havercroft.
“But we’re never going to have the same upper mass range as
MALDI.” (SIMS can analyze molecules up to ∼10,000 Da, whereas
MALDI’s mass limit is in the hundreds of thousands of daltons.)
As a result, many researchers envision the two techniques
merging into a powerful method to simultaneously obtain information. “MALDI imaging focuses on peptides, and SIMS imaging
focuses on lipid drugs, so there’s a great potential combination
there,” says Winograd. “I think there’s a great opportunity there
for expanding the sphere of influence of SIMS if we can get the
MALDI people interested in including our technology.”
Another new application of cluster ion beams and TOF SIMS
is what is called “molecular” depth profiling. (For this purpose,
the C60+and SF5+ ion sources are more often used.) In depth
profiling, the intensity of the incoming ion beam is increased to
the point that rather than removing just a small amount of material
for analysis, a larger quantity is blasted off the surface layer by
layer. Each layer is analyzed separately by MS, and a profile of
analyte versus depth is constructed.
“Elemental depth profiling is well established and used heavily
in industry, but organic [molecule] depth profiling is new,” says
Scott Bryan of Physical Electronics. “The research efforts now
are to optimize these cluster ion beams for minimizing the damage
to the organics while you are depth profiling.... You have to be
able to maintain enough of the molecular information to monitor
the concentration of these molecules as a function of depth and
not reduce them down to elemental carbon.”
The new ability to depth profile more complex species has
many applications. For example, says Havercroft, academic groups
recently have done some very exciting work on the depth profiling
of polymers. “[This] has been a little bit of a holy grail for SIMS
in general,” he says. “We’re now finding you can actually profile
through polymers and monitor the chemical changes with depth.”
There are also biological applications for molecular depth
profiling. “The other thing that people are looking at with SIMS
is the possibility of actually depth profiling with these cluster ion
sources through tissue,” says Gardella. “So, not only do you get
a spatial distribution that’s subcellular but you could actually look
at different portions of a tissue section.” Winograd also believes
that SIMS imaging and depth profiling could some day be
combined. “I have this dream of taking a single biological cell,
preparing it without changing it so I can put it in the mass
spectrometer, and then lifting it directlysentirely preserving all
the x, y, z informationsso you can get a complete chemical analysis
of the cell in three dimensions,” he says.
LOOKING AHEAD
Despite its recent rejuvenation, the SIMS field still has several
potential areas for improvement if it is to become a more widely
used technique. One big issue for industrial users is the cost and
complexity of the instruments. “It’s hard to convince companies
that this is a really cool technique [when] you need to spend $2
million, and then you need to have three Ph.D.s and other
professionals to keep the infrastructure up,” says Gardella.
Several companies offer instruments that address these issues,
and each costs a fraction of a typical SIMS system. “We’re trying to
do our bit to make SIMS available for customers with less money,”
says Reich. “I think otherwise it will remain a fairly niche technique
for those 15 or 20 [buyers] a year that can afford the big price.”
In academia, where the trend seems to be moving toward
biological applications, researchers say that one big challenge in
the near future is going to come from the huge mound of data
produced by a TOF SIMS imaging experiment. “Normally when
you take an image, it’s 256 × 256 pixelssthat’s 65,536 mass spectra
per image.... Each mass spectrum is ∼100,000 data points... which
is a terabyte [of data], or a fraction thereof,” says Winograd.
“There will be quite a bit of need for data compression and
capabilities of dealing with all this information.” Inroads are
already being made toward solving this problem, which will only
grow in the future.
One final area in which SIMS has recently taken a great leap
forward is tandem MS (which will add to the data-handling challenge). In 2002, no commercially available system offered this
capability, but researchers hoped that one would become available.
“To do bioanalysis, you’ve got to have the tandem MS capabilities,”
Winograd says. Only this year did that dream become a reality with
Ionoptika’s new SIMS instrument that includes MS/MS functionality.
Why did it take so long? “I think it’s only becoming necessary
with the advent of cluster beams,” says Rowland Hill of Ionoptika.
“Before 2000, you could only really look at elementals with SIMS,
because the atomic primary beams smash all the molecules up.
Now that we’ve got cluster beams, there’s a drive to do better
mass spectrometry on organics.” The potential of this type of
analysis will become clear once more of these instruments are
installed in laboratories.
Jennifer Griffiths is a senior associate editor of Analytical Chemistry.
AC801528U
Analytical Chemistry, Vol. 80, No. 19, October 1, 2008
7197