Uncorrected proof
Analytical Data: Reliability and Presentation
NAILA SIDDIQUE*, SHAHIDA WAHEED
Chemistry Division, Pakistan Institute of Nuclear Science and Technology,
P.O Nilore, Islamabad 45650, Pakistan
[email protected]*
(Received on 21st March 2013, accepted in revised form 8th August 2013)
Summary: Chemical analysis whether it is used to determine the composition of a sample or
to devise a procedure for testing or preparation of another sample requires systematic
experiment design and implementation. In order to determine and verify the validity of results
various methods are employed to evaluate the data obtained. This process enables the analyst
not only to understand the results but to find possible reasons for differences and similarities
between samples. A simple scheme for carrying out analysis in order to obtain valid and
reliable results is outlined in this paper. Moreover the importance of using reference and
quality control materials to obtain quantitative results is also highlighted. To evaluate the
performance and capability of a laboratory or an analytical procedure, parameters such as
relative bias, z-scores, u-test, tests for accuracy and precision etc can be used. The use and
significance of these parameters is explained using examples in this manuscript. Uncertainties
and errors in measurement as well as the limits of detection (LOD) of an experimental
procedure can also provide vital information about the data obtained. Simple calculations are
used to explain how these can be obtained and what their magnitudes imply.
Key words: Quality assurance (QA), Quality control (QC), Reference materials (RMs), Relative bias, z-score,
u-test, Precision, Accuracy, Uncertainty, Limits of detection (LODs)
Introduction
A basic requirement of any scientific study
is reliable compositional data. For this purpose
various analytical techniques may be used depending
upon the nature of the results required. For example
if elemental composition is required then techniques
such as inductively coupled plasma atomic emission
spectrometry
(ICP-AES),
atomic
absorption
spectrometry (AAS), neutron activation analysis
(NAA), X-ray fluorescence spectroscopy (XRF),
proton induced X-ray emission (PIXE) etc can be
used.
When
analyzing
organic
samples,
chromatographic
techniques,
such
as,
gas
chromatography (GC), high performance liquid
chromatography (HPLC), or other spectroscopy
techniques such as nuclear magnetic resonance
(NMR), infrared spectroscopy (IR), Raman
spectroscopy etc can be used. The selection of an
analytical technique depends on the type of
information required and at what sensitivity level.
For trace analysis involving small amounts of
samples a sensitive and versatile technique is needed.
If the selected technique does not involve laborious,
costly and time consuming sample preparation steps
prior to measurements then possible contamination or
loss of sample is avoided. No one technique is ideal
and therefore the best suited available technique is
selected for analysis. [1-4]
The neutron activation analysis laboratory
(NAA) at the Miniature Neutron Source Reactor
(MNSR), Chemistry Division, Pakistan Institute of
Nuclear Science and Technology (PINSTECH) was
certified as a testing laboratory by the Pakistan
National Accreditation Council (PNAC) on the 19th
of April 2005. [5] It was re-assessed and its
certification was continued for further 3 years in
2008. This certification implies that the data reported
by the NAA/ MNSR Laboratory is reliable and
acceptable to the PNAC if submitted by any industry
or organization. Moreover reports containing
compositional data are routinely required for trade
and to prove the quality of products.
In scientific literature only data which have
been obtained using reliable and tested procedures
are considered acceptable. To produce reliable results
the analysts must follow systematic procedures. The
procedures used should be tested using calibration
procedures and the analysis of reference or standard
samples. Moreover the analysis should be carried out
efficiently so that the expenditure of chemical
reagents and time is kept to a minimum. This goal is
most easily achieved when fewer and simpler sample
preparation steps are employed.
This manuscript was undertaken to present
the methodology routinely followed for elemental
analysis using NAA at the NAA/MNSR Laboratory.
The information provided in this paper which
includes step-wise procedure from sample’s arrival at
the laboratory to the submission of results is provided
as a guide for other analytical chemists. The basic
aim of this manuscript is to educate, inform or
remind analysts of good experimental design,
sequential and methodical analysis and proper way of
1
Uncorrected proof
reporting the results. Therefore this paper will focus
on the data obtained and how to understand and
present it to show its validity and reliability. The
basic concepts provided here can be applied to the
results obtained using any analytical procedure to
obtain valid results in as short time as possible.
Hence “mock results” have been used in examples to
enable the reader to better understand the information
provided.
The QC material, RM material and blank are
prepared following the procedure employed for test
sample preparation, the only difference being is that
in place of the test sample the QC material, RM
material and substrate (blank) are used respectively.
In the next section the various calculations
undertaken to obtain the results of Fe concentration
in a test sample are given as an example.
Chemical Analysis
Analytical results obtained are only as
reliable as the method and care employed to obtain
them. Therefore analytical procedures are carefully
developed and tested prior to analysis of a new type
of sample. Fig. 1 shows a flow diagram for carrying
out chemical analysis. From this figure it can be seen
that once sampling has been completed and a sample
provided to the analyst, the analyst has to select a
suitable analytical technique and prepare the sample
for analysis. In order to do this, steps may be required
which involve drying or grinding of the sample to
obtain a homogeneous sample which fully represents
the test sample. After this, representative sub-samples
of the test sample are taken and prepared for
measurement. The sample preparation step is
technique dependant and may involve the formation
of the sample in the form of a pellet or disc for XRF
and PIXE analysis or a digestion procedure to obtain
the sample in a liquid form for GC, HPLC, AAS and
ICP-AES. Some non-destructive techniques such as
NAA may not even require a sample preparation step
which makes contribution from blank minimal.[6]
In order to carry out analysis the following
samples are prepared:
1.
2.
3.
4.
Test sample, usually measured in triplicate or
more
At least 2 quality control (QC) materials whose
composition is known may be obtained from
reputable RM producers
Reference materials (RMs) which are used for
calibration of instrument and to obtain
quantitative data. These should consist of ~5
samples of different concentrations of an
element/ compound being studied. The data
obtained for these RMs are used to prepare
calibration plots. Moreover certified RMs along
with synthetic or laboratory prepared RMs can
also be used to increase the number of elements/
compounds determined in a single analytical
procedure.
In cases where sample preparation involves
solvents or substrates a blank is also prepared.
Fig. 1: Schematic diagram to show the steps
involved in analytical analysis.
Results
The analytical techniques used for analysis
will provide results for the RM, QC material, blank
sample and test sample. In order to understand the
tools used to obtain and evaluate the results obtained
and what they signify, the Fe concentration in a test
sample is used as an example. An important point to
note is that the number of significant figures used to
report the data should be realistic and consistent
when performing calculations involving multiple
steps. In order to obtain analytical results the
instrument needs to be calibrated as discussed below.
However before quantitative data can be presented
and explained it is best to first discuss measurement
uncertainty and limits of detection (LODs). These are
2
Uncorrected proof
an essential part of any chemical analysis and are
required when result reports are prepared.
Uncertainty Measurement
All analytical results are reported along with
their uncertainties or errors. These provide an
indication about the spread or variation in the value
of the result provided. Measurement uncertainty
analysis may be performed using the methodology
outlined in JCGM 100: 2008. [7] To calculate the
uncertainty for a technique the uncertainty budget has
to be prepared and all possible type A and type B
sources of uncertainties identified. For NAA, Type A
uncertainties are random errors which occur in any
measurement and may include measurement standard
deviation (SD), uncertainty in peak area, weighing
errors, errors in volume measurement, spectral
interferences,
summing
peaks
corrections,
uncertainty due to matrix effect etc while type B
sources of uncertainties include uncertainty
associated with calibration of instruments such as
weight balance and the detector used and
uncertainties quoted in the RM certificate. Both these
uncertainties can be combined in the following way:
2
2
Unc Combined = k * σ 2 + Unc PA
+ Unc W2 + Unc V2 + Unc B2 + Unc D2 + Unc RM
+ ...
(1)
where σ, UncPA, UncW, UncV, UncB, UncD, UncRM etc
are the variation in measurement (standard
deviation), uncertainty in estimation of peak area,
weighing, volume, balance calibration, HPGe
detector calibration and RMs uncertainties
respectively. The first 4 terms are the type A and the
last 3 are the type B sources of uncertainty. The
uncertainties listed in the above equation are by no
means exhaustive and will differ from technique to
technique. Therefore the analyst has to determine all
possible source of uncertainty in their analytical
procedure. Coverage factor of k=1 to 3 can be used in
the above equation. Values of k=1, 2 and 3 imply
confidence intervals of 68.27%, 95.45% and 99.99%
respectively.
From equation 1, it can be seen that the
measurement uncertainty can be reduced by reducing
all its sources. However limitations are imposed on
analytical results by the instruments used and their
capabilities as well as the standards and reagents used
in carrying out a measurement. It is best to use RM
which have low uncertainties for all possible
elements/ compounds. This may not be possible as
RM producers provide recommended as well as
information values for some elements on the RM
certificates. In order to obtain an estimate of the
uncertainty for an element for which an information
value is given, adopting a worse case scenario
approach, the given value is divided by SQRT(3)
assuming a rectangular distribution. However as the
given information value most probably lies near the
centre as compared to the edges, the information
value should be divided by SQRT(6), assuming a
triangular distribution to obtain a measure of the
uncertainty. The latter approach may be used if the
RMs used are routinely used in analysis and in the
past the information values have provided accurate
and precise results.
Limit of Detection (LOD)
Limit of detection (LOD) is defined in many
ways and may also be referred to the minimum
detectable net concentration or limit of
determination/ limit of decision. In its simplest form
it is” the lowest concentration that can be measured
with reasonable statistical certainty. [7] Generally
LODs are calculated using three standard deviations
as recommended by the Committee of Environmental
Improvement of the American Chemical Society.[8]
Therefore LODs are obtained from %3σ and the
concentration of the element/ compound determined
as described below.
Calibration for quantification of results
Up to 5 samples of different concentrations
of an element/ compound are prepared and used as
RM. Here some results for Fe are given in Table-1.
The data in Table-1 shows how a measurement
parameter such as peak area varies with Fe
concentration. From this data Fig. 2 is plotted and a
straight line fitted. The intercept (a) and the slope (b)
of the line is given by the equations below:
Intercept=
(∑
i =n
i =1
i =n
n * i =n x 2 −
∑i =1 i
(n * ∑
b=
i =n
Slope =
)
y i * ∑i =1 xi2 − ∑i =1 xi * ∑i =1 xi * y i (2)
i =n
a=
i =1
i=n
(∑ x )
i =1
2
i =n
i
xi * y i − ∑i =1 xi * ∑i =1 yi
i =n
2
n*
∑i =1 x −
i =n
(
i =n
)
)
(3)
∑i=1 x
i =n
2
Table-1: Calibration data for Fe synthetic RM.
Fe concentration (mg/kg)
10.0±0.5
50.0±2.5
100.0±5.0
500.0±25.0
1000.0±50.0
Mean peak area of emitted gamma ray
725±36
4325±216
9406±470
45749±2287
120635±6032
3
Uncorrected proof
and a mean value is obtained. This is due to the fact
that each sample is analyzed in triplicate. Apart from
these the different sources of uncertainties are also
given in the same unit (percentage). In order to obtain
the overall combined uncertainty all uncertainties
have to be converted to the same unit (concentration
unit or %). Using equation 6 it can be seen that a
peak area of 25462 in the test sample corresponds to
the Fe concentration [Fe] of 242 µg/g. Similarly the
amount of Fe in the blank sample for a peak area of
400 is 31 µg/g giving an overall Fe concentration for
the test sample of 211 µg/g. Using a coverage factor
of 2 and the data given in Table-2 and 3 the
measurement uncertainty can be obtained to give the
amount of Fe in the test sample as 211±41 µg/g while
that in the blank is 31±5 µg/g.
140000
120000
y = 118.96x - 3,325.35
R² = 0.99
Measured Peak Area
100000
80000
60000
40000
20000
0
0
200
400
600
Fe RM Conc (µg/g)
800
1000
1200
Fig. 2: Calibration plot for Fe.
The standard errors in these parameters are
obtained using the formulae given below:
∑i −1 ((b*xi + a )− yi ))2
i=n
x2
1
SE ( a ) = + i = n
n ∑ ( xi − x )
i =1
(n − 2 )
(4)
∑i −1 ((b*xi + a ) − yi )) 2
From Table-2 and 3 it can be seen that the
uncertainty in the [Fe] of the blank sample is higher
than that of the test sample due to the much higher Fe
content of the sample. Similarly the LOD for the
blank is lower but closer to its Fe content. The closer
a value to the LOD for an element/ compound the
higher will be its uncertainty and the less reliable its
value will be.
i =n
SE (b) =
(n − 2 )
∑ (x
i =n
i =1
i
− x)
2
(5)
Using the plot in Fig. 2 and equations 1 to 4
the line that best fits the data comes out to be;
Peak
Area
(±13.10)*[Fe]+3325.35(±0.00)
(6)
=
18.96
The chi-squared for this plot is 0.99 which
shows that this line fits the observed data very well.
The Fe peak areas for the test and blank
samples are given in Table-2. These are test data
presented here to show how quantitative results are
obtained. Here 3 values are given for the peak areas
The concentrations for all possible elements/
compounds for all samples (test sample, QC material
and blank) may be obtained by repetition of the
calculations shown above for all possible elements/
compounds. It should be noted that the number of
elements measured in the blank sample should be as
few as possible for it to act as a good blank.
Moreover the concentration of any element/
compound measured in the blank should also be
much lower than that measured in the actual sample.
Therefore spec pure reagents are generally used in
sample preparation or sampling media such as filters
etc are used which should not contribute to the
background. However some impurities or species at
trace amounts may be present which have to be
measured and their amounts subtracted to obtain the
actual concentrations.
Table-2: Estimation of measurement uncertainty. (Data cited at 95% confidence interval).
Sample
Test Sample
Blank
Value 1
25210
385
Fe Peak Area
Value 2
Value 3
25437
25739
423
392
Mean
25462
400
SD
265
20
%SD
1.04
5.06
UncPA
0.63
5.00
Uncertainty Budget (%)
UncB UncW UncV UncD
4.00
0.50
1.00
1.00
4.00
0.50
1.00
1.00
UncRM
2.00
Combined Unc (%)
9.74
16.59
Table-3: Fe concentration and LOD (mg/kg) of test and blank samples. (Data cited at 95% confidence interval).
Sample
Test Sample + Blank Sample
Blank
Test Sample
Peak Area
25462
400
25062
[Fe]=(Peak Area+3325.35)/118.96 (mg/kg)
242
31
211
Unc (mg/kg) 3 Sigma (%) LOD (mg/kg)
24
5.00
12.10
5
8.00
2.51
41
9.43
19.91
4
Uncorrected proof
Quality assurance / quality control (QA/QC)
To prove the validity of the results obtained
QC materials such as reference materials (RMs) are
used. Here the data for an RM are shown in Table-4
as an example. In this table recommended values
provide by the RM producers along with data
obtained during a study are given. The variation in
measurements (SD), measurement uncertainties and
LODs are also given. An important point to note is
that the number of elements listed in the QC table
should contain all of the elements quantified in the
test sample.
Close examination of Table-4 shows
experimental values to be in close agreement with the
recommended values. Values for elements that the
RM producer has not recommended but has given as
information values (Br, Hf, Lu, Sc, Ta, Tb, Th and
Tb) are included in this table. In order to see if the
results obtained are a good comparison to the
recommended values the relative SD (%RSD) can be
calculated. If the value of this parameter is ~10% or
less for most of the elements studied the results are
probably reliable. Another quick indicator of data
quality is obtained when ratios of the recommended
values are obtained with the observed values. As can
be seen from Table-4 these should be as close to 1.0
as possible. From this table it can be seen that around
76% data lies in the range 0.9-1.1 while 92% data lies
in the range 0.85-1.15.
Application of t-test
Statistical tools such as t-test may be used to
verify similarities between the 2 data sets given in
Table 4. When t-test is applied to the results obtained
for the RM a value on -0.07 is obtained. At a
significance level of 0.05 the value of t for 48 degrees
of freedom is 2.01. As the calculated t is lower than
this value it shows that the experimental values for
these elements do not differ significantly from the
recommended values.
If R. Bias ≤ MAB (Maximum Acceptable
Bias) implies satisfactory performance and if R. Bias
≥ MAB means unsatisfactory performance [9]
MAB values are given by the RM
manufacturer and generally have values of 20-25%.
These have been obtained and given in Table-5.
Scrutiny of the data in Table-5 shows that all of the
reported data have R. Bias <20% apart from As. The
R. Bias for this analyte is >28% making its value
questionable. As these results were obtained using
NAA it can be speculated that the lower As value
may originate from an over correction due to the
presence of bromine in the sample, which may give
rise to spectral interferences due to inadequate
resolution of the two peaks or limitations with the
evaluation software.
Z-Score
z − score =
(Value
Analyst
− Value RM )
(8)
σ
where σ=12.5% of the consensus/assigned value.[9]
If Z≤ 2 satisfactory performance
2<Z< 3 questionable performance
and Z≥ 3 unsatisfactory performance
z-Scores were calculated and are also given
in Table-5. From this data once again it can be seen
that all of the reported data has Z-scores less than 2
apart from As which has a Z-score >2 but <3 making
its value questionable. Table 5 also shows that all
reported results have acceptable Z-scores. This shows
that the procedures employed in obtaining the given
results are good and produce accurate and precise
results. However care should be exercised when
measuring the As concentration in a sample
u-Test
u − Test =
Value Analyst − Value RM
Unc
2
Analyst
+ Unc
(9)
2
RM
Data Evaluation Parameters
In order to carry out more through studies
and evaluate the results obtained the following
parameters may be calculated.
Relative bias
R.Bias =
(Value
Analyst
− ValueRM )
ValueRM
∗ 100% (7)
If u<2.58 it implies satisfactory performance
for a level of probability at 99%. [9]
u-Test values were calculated and are given
in Table 5. These fulfill the criteria for good reliable
results as u<2.58 for all of reported data including
As. Therefore it can be seen that no one parameter
shows the reliability of a value as u-test shows that
As value is also reliable whereas R.Bias and z-score
show the data for this analyte to be questionable.
5
Uncorrected proof
Table-4: Elemental composition of QC material at 95% confidence interval.
Laboratory Values (mg/kg)
Mean
Unc
RSD %
SD
Al
44336
4440
7.4
3268
As
8.24
1.26
10.7
0.88
#Br
216
66
8.9
19.31
Ce
63.02
15.96
6.0
3.81
Co
8.87
2.14
6.7
0.59
Cr
74.00
16.05
5.3
3.87
Cs
3.65
0.54
4.9
0.18
Eu
0.97
0.30
9.3
0.09
Fe
26081
1388
4.2
1099
#Hf
5.75
0.90
3.5
0.24
K
19057
5988
11.5
2197
La
28.55
5.82
8.7
2.52
#Lu
0.28
0.11
10.7
0.03
Mn
332
46
12.9
43
Na
23756
1891
3.7
890
Rb
73.63
10.36
9.5
7.02
Sb
1.34
0.21
10.4
0.14
#Sc
8.14
1.23
3.7
0.26
Sm
4.62
0.41
6.5
0.25
#Ta
0.93
0.26
10.8
0.10
#Tb
0.70
0.33
5.7
0.04
#Th
8.30
1.27
4.8
0.37
V
63.92
10.43
6.6
4.20
#Yb
1.94
0.57
9.3
0.18
Zn
151
17
9.0
13.46
# Given as information values by the RM producer
Element
Ratio of Recommended
/Lab Values
1.17
1.40
1.04
0.97
1.04
1.01
1.02
1.11
1.01
1.08
1.05
1.06
1.11
1.07
1.00
1.11
1.00
1.02
1.07
1.04
0.90
1.07
1.14
1.07
0.93
Recommended Values (mg/kg)
Mean
SD
51800
6475
11.50
1.44
224.00
28.00
61.10
7.64
9.20
1.15
74.40
9.30
3.73
0.47
1.08
0.13
26300
3288
6.23
0.78
20000
2500
30.20
3.78
0.31
0.04
356
45
23800
2975
82.00
10.25
1.34
0.17
8.32
1.04
4.94
0.62
0.97
0.12
0.63
0.08
8.89
1.11
73.00
9.13
2.08
0.26
140.60
17.58
LOD
1215
0.25
113.96
1.30
0.30
3.26
0.45
0.06
363
0.24
3453
1.97
0.02
2
190
24.35
0.30
0.09
0.07
0.05
0.03
0.22
16.82
0.25
3.60
Table-5: Evaluation of data obtained for QC material (Reference Material).
Element
Rel Bias (%)
z-score
u-test
Al
-14.41
-1.15
-0.95
As
-28.39
-2.27
-1.71
#Br
-3.67
-0.29
-0.11
Ce
3.14
0.25
0.11
Co
-3.55
-0.28
-0.13
Cr
-0.53
-0.04
-0.02
Cs
-2.25
-0.18
-0.12
Eu
-10.17
-0.81
-0.34
Fe
-0.83
-0.07
-0.06
#Hf
-7.76
-0.62
-0.41
K
-4.71
-0.38
-0.15
La
-5.45
-0.44
-0.24
#Lu
-9.58
-0.77
-0.25
Mn
-6.87
-0.55
-0.38
Na
-0.18
-0.01
-0.01
Rb
-10.21
-0.82
-0.57
Sb
-0.06
0.00
0.00
#Sc
-2.20
-0.18
-0.11
Sm
-6.54
-0.52
-0.44
#Ta
-3.96
-0.32
-0.13
#Tb
10.71
0.86
0.20
#Th
-6.63
-0.53
-0.35
V
-12.44
-1.00
-0.66
#Yb
-6.57
-0.53
-0.22
Zn
7.13
0.57
0.41
# Given as information values by the RM producer
A1
7464.17
3.26
8.21
1.92
0.33
0.40
0.08
0.11
219.14
0.48
942.80
1.65
0.03
24.46
43.99
8.37
0.00
0.18
0.32
0.04
0.07
0.59
9.08
0.14
10.02
Trueness
For results to be accurate the requirement is
[9]
Trueness
A2
20256.42
4.94
185.77
45.64
6.27
47.86
1.85
0.84
9207.19
3.06
16742.13
17.91
0.31
164.63
9094.42
37.60
0.70
4.15
1.92
0.74
0.87
4.36
35.75
1.62
62.78
Acceptance
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
P (%)
16.02
19.78
33.19
28.24
27.15
25.03
19.48
33.12
13.59
19.99
33.82
23.92
42.69
18.61
14.82
18.82
20.26
19.58
15.36
30.48
48.60
19.79
20.55
32.02
16.76
Precision
Acceptance
A
A
A
A
A
A
A
A
A
A
A
A
N
A
A
A
A
A
A
A
N
A
A
A
A
Final Score
A
A
A
A
A
A
A
A
A
A
A
A
W
A
A
A
A
A
A
A
W
A
A
A
A
seen that all of the data fulfills the expressions A1≤
A2 meaning that the trueness of accuracy criteria is
fulfilled.
Precision
A1≤ A2
To check the precision of the data the
following parameter is calculated: [9]
where
A1 = Value Analyst − Value RM
And A2 = 2.58 ∗ Unc
2
Analyst
(10)
+ Unc
2
RM
(11)
The values of A1 and A2 were calculated
and are given in Table-5. From these results it can be
Unc
Analyst
P =
Value
Analyst
2
Unc RM
+
Value
RM
2
∗ 100%
(12)
If P ≤ LAP (Limit of Acceptable Precision)
implies satisfactory performance
6
Uncorrected proof
LAP data are given by the RM manufacturer
and generally have magnitudes of 20-25%. However
LAPs may be as high as 40% in some cases. The
parameter P has been obtained for all of the elements
determined in the RM sample and are given in Table.
From these results it can be seen that 23 of the 25
elements determined in the RM sample have P≤
MAB, only Lu and Tb have P >40 %. This may be
due to the higher reported uncertainties for these
elements which can be reduced by greater care in
carrying out analysis as well as using RMs with
lower uncertainties for these elements.
Acceptance Criteria
In order to reach a final decision about each
value in a data set the following criteria are used. If
any of the z or u score criteria are not fulfilled then
the result is declared “Not Acceptable”. However if
all criteria are fulfilled but either trueness or
precision criteria is not fulfilled then a further check
is applied i.e. the reported result relative bias (R.Bias)
is compared with the maximum acceptable bias
(MAB) as defined by the RM producer. If R.Bias ≤
MAB, the final score will be “Warning”. “Warning”
reflects two situations; 1) the result has a small
measurement uncertainty; but its bias is still within
MAB or 2) a result close to the assigned property
value is reported, but the associated uncertainty is
large. If R. Bias > MAB the result will be “Not
Acceptable”. Evaluation of the results for the RM
sample using the treatment outlined above provides
the outcomes given in Table-5. Therefore only the
results for Lu and Tb fall into the “Warning”
category, while the data for the remaining 23
elements are all classified as acceptable. [9].
Laboratory Classification
RM manufacturers, such as the IAEA, uses
the following criteria to evaluate the performance of
laboratories which participate in any intercomparsion
or proficiency test (PT) exercise: [10]
Group 1 laboratories
the data;
Group 2 laboratories
90% of the data;
Group 3 laboratories
75%of the data;
Group 4 laboratories
the data
scoring Z < 3 for ≥ 90% of
scoring Z < 3 for 75% to <
scoring Z < 3 for 50% to <
scoring Z < 3 for < 50% of
If the above criteria is used for self
evaluation by a laboratory or for an analytical
procedure, then as all of the data given in Table-4 and
5 have Z < 3 and the laboratory is placed in Group
1. However taking into account all of the acceptance
criteria it can be seen from Table-5 that 23 of the 25
results reported i.e. 92% are acceptable with only the
results for Lu and Tb being deemed unsatisfactory.
Graphical representation of QA/QC results
Generally it is better to show results
graphically as plots show most trends more clearly
and are easier to read. Here various parameters, as
given in Table-4 and 5, have been plotted to highlight
this point. [10-19] In Fig. 3 the recommended and
observed laboratory values have been plotted side by
side as bars to show direct comparison between the
two data sets. This is shown as a log plot to include
elements with a large range of concentrations. The
uncertainties in both data sets have also been plotted
as error bars. From this plot it can be seen that the
bars for the recommended and laboratory values for
each element have very similar lengths. Table-4 and
Fig. 3 are the simplest ways of comparing RM data
with observed results and can point out outliers and
significantly different data points at a glance.
In Fig. 4 the recommended values have been
plotted against the observed values for the RM
sample. This is another simple way of presenting the
results without much data manipulation. As can be
seen from this figure, all data points lie on the y=x
line with intercept of zero. This shows good
agreement between the 2 datasets. Such plots are
generally presented as log log plots to take into
account the large concentration ranges of elements
present in the RMs. Uncertainties cited by the RM
producer and those measured are also plotted to show
any variations in data.
Another graphical method of data
presentation is by plotting the ratios of the
recommended to the laboratory values. This has been
done in Fig. 5. These were given in Table-4 but in
this plot it can be seen that the elements Al, As, Eu,
Rb, Tb and V lie outside the ±10% range. This
parameter shows the questionable character of As
which is underestimated significantly in this study.
Moreover Tb is over-estimated as it has the lowest
ratio.
Graphically data can be presented by
utilizing equations 7 to 9 and plotted the Relative
Bias, the z-scores and the u-test values. This has been
done in Fig. 6 to 8 respectively. Therefore in Fig. 6
the Relative Bias (Rel.Bias%) has been plotted for
the RM for all the elements measured. From this
figure it can be seen that the values of this parameter
are generally negative in magnitude which means that
the observed values are less than the recommended
values suggesting slight under-estimation. This
7
Uncorrected proof
feature is more evident for the elements As and Tb
for reasons mentioned earlier. In Fig. 7 and 8 the zscores and the u-Test values have been plotted. From
these figures it can be seen that As has the highest
magnitude of both these parameters. However the
magnitudes of both these parameters lie within the
prescribed ranges.
From Fig. 3 to 7 the same results are
presented in various ways to distinguish between
Recommended Results
reliable and less reliable results. The same results are
presented in Table-3 and 4. From these plots and
tables it can be seen that the data obtained for the QC
material RM is in very good agreement with the
recommended values. Hence the methodology used
to obtain the results reported in Table-4 provided
reliable results giving the analyst confidence in the
reported results.
Laboratory Results
Zn
Yb
V
Th
Tb
Ta
Sm
Sc
Sb
Rb
Na
Mn
Lu
La
K
Hf
Fe
Eu
Cs
Cr
Co
Ce
Br
As
Al
0.1
1
10
100
Concentration (mg/kg)
1000
10000
100000
Fig. 3: Plot of recommended and laboratory values for RM sample.
100000
1.60
1.50
1000
100
10
1
0.1
0.1
1
10
100
1000
10000
100000
Recommended Values (µg/g)
Ratios of Recommended/ Laboratary Values
Laboratory Values (µg/g)
10000
1.40
1.30
1.20
1.10
1.00
0.90
0.80
0.70
0.60
0.50
Al As Br Ce Co Cr Cs Eu Fe Hf K La Lu Mn Na Rb Sb Sc Sm Ta Tb Th V Yb Zn
Fig. 4(a): Comparison of recommended data with
the laboratory results for RM sample
Fig. 4(b): Plot of ratios of recommended/ laboratary
values for RM sample
8
Uncorrected proof
30
20
Relative Bias (%)
10
0
Al As Br Ce Co Cr Cs Eu Fe Hf K La Lu Mn Na Rb Sb Sc Sm Ta Tb Th V Yb Zn
-10
data and how to present them so that the reader can
easily understand how they have been obtained. It
also shows the significance of simple evaluation tools
which can be used routinely to evaluate the results
obtained and provide confidence in the reported
results. Such tools can be used to devise and test new
analytical procedures as well as test and evaluate the
performance of individual laboratories or analysts.
-20
References
-30
1.
2.
Fig. 5: Relative bias plot for RM sample.
3.
3
2
4.
Z-Score
1
0
Al As Br Ce Co Cr Cs Eu Fe Hf K La Lu Mn Na Rb Sb Sc Sm Ta Tb Th V Yb Zn
5.
-1
-2
6.
-3
7.
Fig. 6: Z-scored plot for RM sample.
8.
9.
2
1.5
1
u-Test
0.5
0
Al As Br Ce Co Cr Cs Eu Fe Hf K La Lu Mn Na Rb Sb Sc Sm Ta Tb Th V Yb Zn
10.
-0.5
-1
11.
-1.5
-2
12.
13.
Fig. 7: u-Test plot for RM sample.
Conclusion
The information presented in this paper
shows the importance of understanding analytical
14.
S. Waheed, N. Siddique and J. H. Zaidi, Journal
of Radioanalytical and Nuclear Chemistry, 289,
765, (2011).
N. Siddique and S. Waheed, Journal of
Radioanalytical and Nuclear Chemistry, 291,
817 (2012).
N. Siddique, S. Waheed and Y. Faiz, Journal of
Radioanalytical and Nuclear Chemistry, 291,
919 (2012).
N. Siddique, Sabiha-Javied, S. Waheed and M.
Tufail, Journal of Radioanalytical and Nuclear
Chemistry, 292, 445 (2012).
http://www.pnac.org.pk/info/Accreditation%20S
copes/Finalized%20scopes%20(Feb%202007)/PI
NSTECH%20Lab,%20Nilore.pdf
http://www.pnac.org.pk/index.php?PageId=87
International Standard ISO/IEC 17025, Second
edition 2005-05-15 General Requirements for the
Competence of Testing and Calibration
Laboratories ISO, Geneva Switzerland
Quantifying Uncertainty in Nuclear Analytical
Measurements, Eurachem/ Citac Guide, (2000)
R. J. Mesley, W. D. Pocklington and R. F.
Walker, Analyst, 116, 975 (1991).
A. Shakhashiro, A. Trinkl, A. Torvenyi, E.
Zeiller, T. Benesch, U. Sansone. (2006)
IAEA/AL/168, Report on the IAEA-CU-2006-06
Proficiency Test on the determination of major,
minor and trace elements in ancient Chinese
ceramic, Seibersdorf, Austria, November 2006
N. Siddique, S. Waheed, M. Daud, A. Rahman
and S. Ahmad, Journal of Radioanalytical and
Nuclear Chemistry, 274, 181 (2007).
S. Waheed, M. Waseem, A. Rahman and S.
Ahmad, Geostandards Newsletter, 25, 137
(2001).
S. Waheed, S. Ahmad, A. Rahman and I. H.
Qureshi, Journal of Radioanalytical and Nuclear
Chemistry, 250, 97 (2001).
S. Waheed, A. Rahman, M. Daud, J. H. Zaidi
and S. Ahmad, Radiochimica Acta, 92, 939
(2004).
N. Siddique, A. Rahman, S. Waheed, M. Wasim,
M. Daud and S. Ahmad, Journal of
Radioanalytical and Nuclear Chemistry, 268,
579 (2006).
9
Uncorrected proof
15. S. Waheed, A. Rahman, N. Siddique, S. Ahmad
and M. Rossbach, Accreditation and Quality
Assurance, 12, 311 (2007).
16. S. Waheed, N. Siddique, M. Arif, I. Fatima, N.
Khalid, S. Rahman, M. Daud and M. Wasim,
Journal of Radioanalytical and Nuclear
Chemistry, 277, 539 (2008).
17. N. Siddique, A. Rahman, S. Waheed, M. Wasim,
M. Daud and S. Ahmad, Journal of the Chemical
Society of Pakistan, 31, 622 (2009).
18. S. Waheed, A. Rahman, N. Siddique and S
Ahmad, Geostandards and Geoanalytical
Research, 31, 133, (2007).
19. N. Siddique and S. Waheed, Journal of the
Chemical Society of Pakistan, 31, 916, (2009).
10