977
Electrophoresis 2008, 29, 977–993
Dietrich Kohlheyer
Jan C. T. Eijkel
Albert van den Berg
Richard B. M. Schasfoort
MESA1 Institute for
Nanotechnology,
University of Twente,
Enschede, The Netherlands
Received September 28, 2007
Revised November 23, 2007
Accepted November 24, 2007
Review
Miniaturizing free-flow electrophoresis –
a critical review
Free-flow electrophoresis (FFE) separation methods have been developed and investigated
for around 50 years and have been applied not only to many types of analytes for various
biomedical applications, but also for the separation of inorganic and organic substances. Its
continuous sample preparation and mild separation conditions make it also interesting for
online monitoring and detection applications. Since 1994 several microfluidic, miniaturized FFE devices were developed and experimentally characterized. In contrast to their
large-scale counterparts microfluidic FFE (m-FFE) devices offer new possibilities due to the
very rapid separations within several seconds or below and the requirement for sample
volumes in the microliter range. Eventually, these m-FFE systems might find application in
so-called lab-on-a-chip devices for real-time monitoring and separation applications. This
review gives detailed information on the results so far published on m-FFE chips, comprising its four main modes, namely free-flow zone electrophoresis (FFZE), free-flow IEF
(FFIEF), free-flow ITP (FFITP), and free-flow field-step electrophoresis (FFFSE). The principles of the different FFE modes and the basic underlying theory are given and discussed
with special emphasis on miniaturization. Different designs as well as fabrication methods
and applied materials are discussed and evaluated. Furthermore, the separation results
shown indicate that similar separation quality with respect to conventional FFE systems, as
defined by the resolution and peak capacity, can be achieved with m-FFE separations when
applying much lower electrical voltages. Furthermore, innovations still occur and several
approaches for hyphenated, more integrated systems have been proposed so far, some of
which are discussed here. This review is intended as an introduction and early compendium for research and development within this field.
Keywords:
Free-flow electrophoresis / Free-flow isoelectric focusing / Free-flow isotachophoresis / Microfluidic
DOI 10.1002/elps.200700725
1
Introduction
Since the introduction of free-flow electrophoresis (FFE) in
the 1960s [1], this separation method has found a permanent
position among analytical and preparative methods in biochemistry and chemistry for the separation of, e.g., cells,
organelles, peptides, proteins, inorganic, and organic compounds [2–5]. In FFE, analytes are separated continuously in
Correspondence: Dietrich Kohlheyer, MESA1 Institute for Nanotechnology, University of Twente, P. O. Box 217, NL-7500AE
Enschede, The Netherlands
E-mail:
[email protected]
Fax: 131-53-489-3595
Abbreviations: ASS, acetylsalicylic acid; FFE, free-flow electrophoresis; ì-FFE, microfluidic free-flow electrophoresis; FFFSE,
free-flow field step electrophoresis; FFIEF, free-flow IEF; FFITP,
free-flow ITP; FFZE, free-flow zone electrophoresis; LOC, lab-ona-chip; SPR, surface plasmon resonance
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
an electrical field applied perpendicular to a thin pressuredriven carrier electrolyte flow between two insulating plates,
as shown in Fig. 1. The sample mixture is injected into the
carrier electrolyte flow and with increasing residence time
the differently charged components split up into diverging
lanes which can be collected at various device outlets.
Commercially available large-scale FFE systems (e.g., BD
Diagnostics, Germany) were originally developed as standalone sample clean-up devices, e.g., as a prefractionation step
prior to 2-DE or other detection and identification methods.
Size-reduced mini-FFE systems were developed later and
were coupled to MS and LC enabling continuous separation
with online detection possibilities [6, 7]. This combination of
FFE as a continuous separation method with online detection would allow for real-time monitoring of analytes, such
as patients samples, reaction products, and more. These sizereduced systems are still complicated and slow in operation
and downscaling would be very promising. Eventually such
microfluidic FFE (m-FFE) systems might find application in
small portable devices and point-of-care (POC) tools [8]. With
www.electrophoresis-journal.com
978
D. Kohlheyer et al.
Figure 1. Illustrative principle of FFZE and main components.
the upcoming trend of miniaturization and the development
of lab-on-a-chip (LOC) systems [9], also microfluidic, chipbased FFE systems have been developed and demonstrated.
The miniaturization of FFE implies several advantages
especially considering sample volume and separation speed.
In contrast to the tens of milliliters of sample consumed by
conventional large-scale FFE devices, m-FFE systems require
only tens of nanoliters up to hundreds of microliters of
sample. This is especially interesting in clinical analysis
where often only low sample volumes are available. Furthermore, instead of residence times of up to tens of minutes, mFFE devices separate within several seconds. Scaling laws
predict a 100-fold increase in speed for a ten-fold (linear)
downscaling of an FFE experiment [10]. Such short analysis
times would be very beneficial in POC devices. The shallower
separation chambers of several micrometers enable good
heat dissipation allowing higher electrical field strengths
necessary for rapid separations. According to theory, the
quality of the separation, as defined by the resolution does
Electrophoresis 2008, 29, 977–993
not decline with reduced separation geometry [11], as shown
for CIEF. Due to the mentioned advantages of miniaturizing
FFE and due to the availability of new fabrication technologies, the interest in this field has recently grown rapidly
indicated by the relatively large number of m-FFE publications in 2005 and in 2006. This review has its focus mainly
on m-FFE systems, while for a more general overview the
reader is referred to the recent review by Pamme [12] on
continuous flow separations in microfluidic devices.
This review gives detailed information about the recent
developments in the field of m-FFE systems including various modes of FFE. The different technological solutions
and developments are critically analyzed and compared
with respect to design, fabrication methods, and separation
quality. Furthermore, detection methods and hyphenation
aspects are discussed. In this paper, we will first describe
the common separation modes of FFE and discuss basics
of the related theory. In Section 3 we will describe different
technological approaches that have been applied for FFE
microchips. In Section 4 recent separation results are
shown and hyphenation and detection aspects are discussed in Section 5.
2
FFE modes and related theory
Most of the modes of standard CE [13] can be applied in FFE
as well. The four common ones are free-flow zone electrophoresis (FFZE), free-flow IEF (FFIEF), free-flow ITP
(FFITP), and free-flow field step electrophoresis (FFFSE) [14,
15].
2.1 FFZE
FFZE implies the usage of a carrier electrolyte flow of constant composition of pH and electrical conductivity in which
components are separated according to their mobility (determined by the charge-to-size ratio; Fig. 2a). It has been found
Figure 2. Modes of FFE: (a) FFZE, (b) FFIEF, (c) FFITP, and (d) FFFSE.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.electrophoresis-journal.com
Microfluidics and Miniaturization
Electrophoresis 2008, 29, 977–993
that there is a direct correlation between CZE as used for
analytical applications and FFZE as used for preparative
ones. The theory of this correlation can be found in refs. [16,
17]. In FFZE the analyte to be separated is deflected linearly
under a constant angle determined by the electrical field
strength, the analyte mobility, and the flow velocity. The
migration distance d of the analyte moving through the
electrical field is given by
d ¼ mp Et
(1)
where mp is the apparent electrophoretic mobility, E is the
electrical field strength, and t is the residence time of molecules in the separation chamber [18].
The goal of FFZE is to achieve high resolution, which is
reached by analytes migrating in narrow zones with sharp
boundaries. However, several phenomena negatively influence the separation quality. The following sources of band
broadening are common in FFZE: the width of the injected
sample stream leads to a band SD sINJ, and further types of
band broadening are diffusional broadening (sD), hydrodynamic broadening (sHD), electrodynamic broadening
(sED), Joule heating (sJH), and electromigration dispersion
(sEMD) [4, 19]. Not all sources of band broadening are discussed here and the reader is referred to the literature for
more details. The variance s2T of a separated analyte band is
then given by the sum of the variances of all broadening
contributing factors:
s2T ¼ s2INJ þ s2D þ s2HD þ s2ED þ s2JH þ s2EMD
(2)
The variance due to the sample injection width wi is usually
expressed by [20]
s2INJ ¼
w2i
12
(3)
Reducing the sample injection width is often easier in
microfluidic systems than in larger fluidic systems for
example by the precise control of the neighboring laminar
flow streams causing hydrodynamic focusing of the sample
as for example shown in ref. [21].
The variance caused by lateral diffusion is directly related
to the residence time t of the analyte in the separation
chamber, and can be expressed by
s2D ¼ 2Dt
(4)
where D is the analyte diffusion coefficient [22]. Short residence times, as usually involved in m-FFE systems therefore
reduce this effect.
The parabolic flow profile, which is caused by the pressure-driven flow leads to unequal velocity regions inside the
separation chamber. The analytes flowing near to the top and
the bottom plate spend more time in the electrical field, and
thus are deflected more than analytes flowing near the
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
979
chamber center. This effect is called hydrodynamic broadening and leads to a crescent-shaped deformation of the
sample. The variance due to hydrodynamic band broadening
equals
s2HD ¼
h2 t 2 2
E mp
105D
(5)
where h is the chamber height [22]. Equation (5) shows
that a shallower separation chamber, as used in m-FFE devices does strongly reduce the effect of hydrodynamic band
broadening.
The applied electrical field together with an electrical
double layer present at the separation chamber surfaces
leads to EOF inside the separation chamber. Normally, in
an open channel EOF results in a plug-shaped flow profile
from the anode towards the cathode. However, in many
FFE systems the EOF generated at the upper and lower
cover plates results in a hydrodynamic counterflow in the
center of the chamber due to a high transport resistance in
the direction of the flow (e.g., due to incorporated membranes). This hydrodynamic flow perpendicular to the
separation axis often results in crescent-shaped sample
deformation comparable to that caused by hydrodynamic
band broadening. This effect is generally referred to as
electrodynamic band broadening. As discussed later, not all
m-FFE systems are subject to this type of band broadening,
therefore we will not discuss the theory in more detail here
and one is referred to ref. [3].
As the electrical current heats the separation medium
(Joule heating) inside the FFE device a temperature gradient
is established between the two cover plates with the maximum in the center. This temperature change leads to a viscosity decrease in the center locally affecting the analyte mobility, and thus leading to band broadening. In contrast to
actively cooled conventional FFE systems the shallower
separation region of m-FFE systems favors fast heat dissipation reducing this type of sample distortion and usually also
much lower electrical currents are involved. Experiments
have confirmed that m-FFE separations could be performed
without noticeable influence of Joule heating applying electrical field strengths of up to 60 V/mm [22, 23] with an
exception given by ref. [24].
Zones of different electrical conductivity in sample and
carrier electrolyte will lead to electromigration dispersion [4,
19]. This type of distortion results from different migration
velocities of the analytes in zones having different electrical
field strengths. This effect can be minimized by choosing
appropriate buffer systems with similar conductivities for
sample and carrier electrolyte.
An interesting approach was shown by Fonslow and
Bowser [22] in which the authors derived an analog of the van
Deemter equation describing the separation in FFZE. They
showed that linear velocity, electric field, and migration distance must all be considered to optimize bandwidth and
resolution.
www.electrophoresis-journal.com
980
D. Kohlheyer et al.
Electrophoresis 2008, 29, 977–993
The separation resolution Rs of two adjacent peaks is
defined by
Rs ¼
d1 d2
2ðsT1 þ sT2 Þ
(6)
Assuming that the band broadening is dominated by diffusion and inserting Eqs. (1) and (4) one can derive (assuming
equal diffusion coefficients)
Rs ¼
ðmt1 mt2 ÞEt
pffiffiffiffiffiffiffi
4 2Dt
(7)
Substituting the residence time t with L/v, where L is the
length of the separation chamber in the flow direction and v
is the linear flow velocity, and substituting E with Veff/W,
where Veff is the effective separation voltage (the effective
voltage utilized for separation) and W is the separation
chamber width, one obtains
Rs ¼
rffiffiffiffiffiffiffi
ðmt1 mt2 Þ
L
pffiffiffiffiffiffi
2
4 2D |fflffl{W
zfflffl}
|fflfflfflfflfflffl{zfflfflfflfflfflffl}
Analytes
V eff
pffiffi
v
|{z}
(8)
Device Tunable
geometry parameters
This equation shows that when scaling down FFZE devices,
the separation resolution is independent of the device size
(assuming a constant device aspect ratio L/W) and is dependent only on the applied separation voltage and the linear
flow velocity. In the derivation of this equation many
assumptions were made and it is just intended to guide the
reader and to demonstrate that resolution eventually is independent of the actual device size and does not have to suffer
from downscaling. Of course at the same time the separation
time L/v decreases proportionally with downscaling.
2.2 FFIEF
In FFIEF, the used carrier electrolyte is composed of a mixture
of ampholytes, which lead in the presence of the applied voltage
and a natural pH gradient (mostly achieved by low and high pH
anodic and cathodic electrode electrolytes), to the formation of a
linear pH-gradient perpendicular to the flow direction. Sample
components migrate within this pH gradient due to the electrical field until they reach the point where their pI is equal to
the local pH value of the buffer, where they become neutrally
charged and focus (Fig. 2b). Unlike the linear separation technique FFZE where the SD of the band increases as the separation continues, the SD of a sample fraction bandwidth in FFIEF
stays constant once equilibrium is reached. Band broadening,
e.g. due to diffusion, is continuously counteracted since species
leaving the equilibrium zone become charged again and
migrate back to the location of zero charge [25]. Therefore, the
sample injection width and band broadening factors as discussed for FFZE play a minor role in FFIEF. However, the low
solubility of analytes at their pI and therefore precipitation
often leads to nonideal and distorted focusing.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
The following differential equation, valid under steady
state conditions, describes the equilibrium conditions between simultaneous electrophoretic and diffusional mass
transport during IEF,
dðCmEÞ
d dC
¼
D
dx
dx dx
(9)
where C is the analyte concentration at position x in the
separation channel, m is its mobility at that point, E is the
electrical field strength, and D is the diffusion coefficient
[26].
Useful parameter to express the quality of an equilibrium
gradient separation system such as an IEF system, include
the SD of the peak width s (Eq. 10), minimum pI value that
can be resolved D(pI)min (Eq. 12), and peak capacity n (Eq.
14). The solution to the differential Eq. (9) at final steady state
gives a Gaussian concentration distribution, with an SD
expressed by
sffiffiffiffiffi
D
(10)
s¼
pE
where D is the diffusion coefficient, E is the electrical field
strength, and
p¼
dm dðpHÞ
dðpHÞ dx
(11)
where d(pH)/dx is the pH gradient and dm/d(pH) the mobility slope of the analyte [26]. A way to express the resolving
power of IEF is given by Eq. (12) which expresses the minimum difference in pI of two species that still can be separated (peak distance 3s) [26].
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
DðdðpHÞ=dxÞ
DðpIÞmin ¼ 3
(12)
E ðdm=dðpHÞÞ
By replacing d(pH)/dx with DpH/L, where L is the length
of the pH gradient and DpH is the total difference of the
applied pH gradient, and replacing E with Veff/L, where Veff
is the voltage drop utilized for separation, one can rewrite
Eq. (12) to
sffiffiffiffiffiffiffiffiffiffi sffiffiffiffiffiffiffiffiffiffiffi
DpH
D
(13)
DðpIÞmin ¼ 3
V eff
dm
|fflfflffl{zfflfflffl} |fflfflfflffl{zffldpH
fflfflffl}
Device
Analyte
This approach has been utilized for CIEF by Das and Fan
[11]. One can see from Eq. (13) that only DpH and Veff
are device-related parameters while the others depend on
the analyte. It thus becomes clear that the resolution
D(pI)min is independent of the device dimensions but
increases with total applied voltage, a result similar to
that for FFZE. Assuming ideal separation conditions,
www.electrophoresis-journal.com
Microfluidics and Miniaturization
Electrophoresis 2008, 29, 977–993
981
such as no Joule heating one can therefore conclude that
microfluidic FFIEF systems can achieve comparable resolution to their larger counterparts. Obviously, a good
resolution is favored by a high separation voltage and a
narrow pH gradient. The separation time, however, benefits from downscaling, since it decreases linearly with
the chamber length.
A common method to express the number of peaks that
can be resolved is the peak capacity n, defined by
n¼
L
4
s
(14)
where L is the total length of the pH gradient [25]. Also the
peak capacity is theoretically independent of the device
dimensions as it can be rewritten to
n¼
rffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi dm=dpH
DpHV eff
|fflfflfflfflfflfflffl{zfflfflfflfflfflfflffl} |fflfflfflfflfflfflfflffl {16D
zfflfflfflfflfflfflfflffl }
Device
(15)
Analytes
2.3 FFITP
In FFITP, the sample is introduced between leading and
terminating buffers respectively having the highest and the
lowest mobility of their ions with respect to the mobility of
the analytes. During FFITP separation the analytes form
adjacent regions according to their descending electrophoretic mobility (Fig. 2c). For more theory about ITP the reader
is referred to refs. [27–29].
2.4 FFFSE
A fourth mode is FFFSE in which an electrical field step
gradient is built up by introducing a less conductive buffer in
the center of the separation chamber, enclosed by more conductive buffers on the sides. The analytes to be separated
move relatively fast through the center zone with high electrical field strength until they reach the boundary with the
high ionic buffer concentration and lower electrical field
strength. This field step results in a drastic reduction of the
analyte electrophoretic velocity and the components therefore become concentrated and focus [3, 30].
3
Device technology
A conventional large-scale FFE separation device, as illustrated in Fig. 3, consists of two insulating plates (often glass
or acrylic glass (polymethyl methacrylate, PMMA)) which are
separated by a membrane spacer (e.g., cellulose nitrate [7]) of
up to several hundred micrometers in thickness. This spacer
defines the separation chamber height and is conductive for
an electrical current but due to its internal structure acts as a
confinement for the pressure-driven fluid inside the separation chamber. In this configuration, it spatially defines three
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Figure 3. Illustration of a disassembled conventional FFE
separation unit: 1, top plate; 2, bottom plate; 3, cooling plate; 4,
electrode; 5, membrane spacer; 6, electrode electrolyte inlet; 7,
electrode electrolyte outlet; 8, electrode connection; 9, sample 1 carrier electrolyte inlet; and 10, fraction collector.
regions: the separation region, the anode bed, and the cathode
bed. Electrodes are placed outside the separation region
avoiding the disturbance of generated gas bubbles and minimizing the interference of chemical side products migrating
into the separation region. Continuous electrolyte flow inside
the electrode beds ensure stable electrical properties by
flushing away generated oxygen, hydrogen bubbles, and
chemical side products. Usually the device is placed on an
actively cooled plate in order to efficiently remove heat generated by the electrical current. To some extent this principle is
applicable for smaller FFE systems as well, using traditional
micromachining techniques such as injection molding and
milling [6, 7, 31]. However, when the lateral feature size
decreases to millimeters and the separation chamber heights
decrease to tens of micrometers the fabrication of the necessary thin cellulose membranes sheets and their manual
handling including device assembly is not practicable anymore. Applying modern cleanroom fabrication technologies
new manufacturing principles for FFE microdevices have
therefore been developed whereby the developments mainly
focused on alternatives to conventional membranes. Generally, three aspects are thereby of main importance: (i) efficient removal of gas and chemical side products formed during electrolysis with no crosscontamination between the
separation chamber and the side beds; (ii) a stable electrical
field over the separation chamber, and (iii) good voltage efficiency. The voltage efficiency ZV is defined by
ZV ¼
V eff
V total
(16)
where Vtotal is the total applied voltage and Veff is the voltage
effectively utilized for separation. In conventional large-scale
FFE devices the width of the membranes is much smaller
www.electrophoresis-journal.com
982
D. Kohlheyer et al.
than the actual width of the separation chamber, resulting in a
negligible voltage loss across the membranes when the
separation voltage is applied. In contrast to this, in m-FFE
chips the membrane-replacing structures often have widths
of the same size as the separation chamber causing a large
extent of voltage loss, resulting in low voltage efficiency.
Unfortunately, reducing the width of the membranes or
similar structures to achieve low electrical resistance normally leads to structures with low hydrodynamic resistance as
well, possibly causing fluid leakage towards the electrode side
beds or vice versa. In general, as shown in Fig. 4, four different
technological approaches were demonstrated in m-FFE systems trying to fulfill the mentioned criteria: (i) Electrodes
were placed in open side reservoirs with membrane-like
structures of high hydrodynamic resistance isolating the
separation chamber. (ii) Electrodes were placed in closed side
beds with a continuous electrode flow, with a membrane
equivalent or different structures to shield the separation
chamber. (iii) Electrodes were integrated into the separation
chamber with no additional structures. (iv) The electrodes
were electrically and mechanically isolated from the separation chamber. Table 1 briefly compares the papers using these
different approaches which are discussed in the following
sections. Comparisons should be taken with care, since many
aspects have to be considered that are not always simple to
compare in the different approaches.
3.1 Open electrode side beds with membrane
equivalent
The implementation of open electrode side beds to place the
electrodes allows for an easy ventilation of gas products
formed during electrolysis. This design, however, results in a
Figure 4. Illustration of various design categories applied for mFFE systems. (a) Open electrode beds with membrane-like structure or structures of a similar function; (b) closed electrode beds
with membrane-like structures; (c) electrodes inside the separation chamber; and (b) mechanically and electrically insulated
separation chamber.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Electrophoresis 2008, 29, 977–993
pressure gradient from the separation chamber towards the
electrode side beds which can cause fluid leakage. This fluid
leakage has to be compensated by proper shielding of the
separation chamber with membrane equivalent structures
with high hydrodynamic resistance. Since there is no need
for a precise control of electrolyte fluid streams the working
setup can be simplified. However, for longer separation
times a continuous refreshment of the electrode solutions
should be implemented, to avoid electrical conductivity
changes.
Zhang and Manz [32] developed an FFE chip using
PDMS and Pyrex glass. In their design, they used an array of
narrow side channels connecting the separation chamber
with the electrode side beds. These side channels were of
high hydrodynamic resistance acting as a membrane
equivalent structure. Gas bubbles were efficiently hindered
from entering the separation chamber and eventually left
through the open electrode side beds. The design allowed for
relatively fast flow rates and furthermore, the usage of glass
and PDMS made the application of high voltages possible.
This resulted in an FFZE separation of fluorescent components and labeled amino acids with residence times below
2 s. The same chip design was used later to demonstrate
rapid FFIEF by Xu et al. [33]. They showed the 400-fold
focusing of Angiotensin II at its pI within 430 ms. Despite
the very short separation times, more than 90% of the
applied voltage was lost across the side channels. The high
hydrodynamic resistance of the side channels, necessary for
rapid operation, in a trade off therefore caused a tremendous
increase in the electrical resistance of the side channels,
making the device less efficient in terms of applied voltage.
Applying FFZE in this device, the low voltage efficiency
turned out not to be a limiting factor, especially when considering it as a proof-of-principle device. However, in FFIEF a
large part of the pH gradient was established inside the side
channel membranes and not in the separation chamber due
to carrier ampholytes migrating into the side channels.
Therefore, only the part of the pH gradient which formed
inside the separation chamber was available for IEF. This
effect significantly limited the separation capacity. Janasek et
al. [34] applied a comparable PDMS chip to demonstrate for
the first time FFITP of fluorescein and Eosin G.
In conventional gel electrophoresis, a dense swollen
hydrogel made from cross-linked acrylamide monomer is
used as a separation matrix. Similar gels were used in
microfluidic systems as so called ion-bridges or salt-bridges
(e.g., used in electroosmotic pumps [35]) making it also a
suitable membrane material for m-FFE. This principle was
applied by several groups in miniaturized FFE devices to
form conductive membranes.
Kohlheyer et al. [21] presented a m-FFE glass chip with
photopolymerized acrylamide membranes. This device was
fabricated by using two wafers of Borofloat glass, one containing the 15 mm high separation chamber as well as inlet
and outlets, as shown in Fig. 5. The glass wafers were bonded
by direct glass wafer bonding with no need for an additional
www.electrophoresis-journal.com
Microfluidics and Miniaturization
Electrophoresis 2008, 29, 977–993
983
Table 1. Rough comparisons of different FFE design approaches
Category
(related
to Fig. 4)
Author
Year of
publication
Principle
a
Xu, Zhang and
Janasek [32–34]
Kohlheyer et al.
[21, 23]
Albrecht and
Jensen [24]
de Jesus et al. [37]
2003, 2006
Side channels
4.07
2006, 2007
1994, 1996
Acrylamide
membranes
Acrylamide
membranes
Acrylamide
membranes
Side channels
2005
2006
a
a
a
b
2006
2006
c
Raymond et al.
[18, 20]
Fonslow and
Bowser [36]
Fonslow et al.
[22, 38]
Kobayashi et al.
[39]
Lu et al. [40]
c
Song et al. [41]
2006
d
Janasek et al. [42]
2006
b
b
b
2003
2004
Chamber
width
(mm)
Removal of
bubbles and
side products
Stable electrical field,
separation
Comments
4.5
o
o
3.5
63a)
1
1
1
15a)
1
1
Reduced pH gradient
in FFIEF
Low mechanical
stability
Joule heating
18
69
1
1
10
60a)
o
2
Side channels
10
50
2
2
Deeper electrode
beds
Shallow side
banks
Integrated
electrodes
10
91
1
1
56.5
95a)
1
1
n.a.
1
Diffusion
potential
Electrostatic
induction
1
Voltage
efficiency
(%)
100
0.05
n.a.
n.a.
1
4.4
50
n.a.
1
Laborious electrode
connections
Low break-down
voltage
Bubbles distorted
separation
Precise flow control
required
Relatively large, ø
100 mm wafer sized
Only low diffusive
analytes, long
residence times
Ampholyte less IEF
Still under
investigation
a) Estimated by the author; (2, inefficient; o, fair; 1, efficient).
Figure 5. Illustration of the m-FFE device during FFZE: 1, glass
plates; 2, separation chamber; 3, sample inlet; 4, sheath flow
inlet; 5, electrode wire; 6, fraction outlet; 7, acrylamide membrane. Reproduced with permission of The Royal Society of
Chemistry from [21].
intermediate silicon layer as used for example by Fonslow
and Bowser [36]. The electrodes were placed in open reservoirs to allow the ventilation of gas. Depending on the used
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
separation chamber width, between 60 and 40% of the
applied voltage was utilized across the separation region. The
stability of the membranes turned out to be the limiting factor. At elevated fluid velocities and especially during FFIEF
the membranes eventually broke. An improved version of
this FFE chip was developed and used to achieve an
increased separation resolution in FFIEF [23].
Albrecht and Jensen [24] demonstrated a FFIEF PDMS
device with integrated functionalized photopolymerized
acrylamide membranes, as shown in Fig. 6. These membranes of low and high pH respectively provided buffering
capacity during IEF to confine a pH gradient between
pH 3 and 9 with no need for additional sheath flow
streams as used, e.g., by Kohlheyer et al. [21]. Furthermore,
an integrated cooling system was used to reduce the negative effect of Joule heating at high electrical field strengths.
Considering the width of the membranes, the utilized
voltage drop across the separation channel is estimated by
us to be around 15% of the total applied voltage. Albrecht
and Jensen [24] did not report on low mechanical stability
of their membranes.
de Jesus et al. [37] fabricated a low-cost m-FFE chip
using laser printing toner as a structural material on a
glass substrate. Furthermore, they used a polymerized
hydrogel to fill the electrode reservoirs shielding the
www.electrophoresis-journal.com
984
D. Kohlheyer et al.
Electrophoresis 2008, 29, 977–993
Figure 7. Principal FFE layout used by Raymond et al. [18]. A
highly dense side channel array acted as membrane between the
separation chamber and electrode beds. Reproduced with permission from ref. [18], copyright American Chemical Society.
Figure 6. Top view (a) shows the PDMS device with the sample
channel bordered by left and right porous material regions
(crosshatched areas) and anode and cathode, respectively. Silicone sealant (solid array) is used to form the reservoirs for the
anolyte and catholyte buffers, as well as to hold the platinum
electrodes in place. Reproduced with permission from ref. [24],
copyright Wiley-VCH Verlag GmbH & Co. KGaA.
separation chamber. The electrodes were placed inside two
electrolyte filled syringes connected to the side openings via
polyethylene tubing. Gas formation inside the syringes did
not enter the side openings nor caused disturbance of the
separation. de Jesus et al. achieved a similar voltage efficiency
as Kohlheyer et al. [21].
3.2 Closed electrode side channels
The usage of closed electrode channels requires additional
flow streams ensuring an effective removal of produced gas
bubbles and chemical side products, and to ensure stable
electrical properties. Structures or obstacles also have to be
implemented to avoid gas bubbles from entering the separation chamber. In addition, an exact flow balancing of the
fluid flow streams in the electrode and separation channels
becomes important to avoid cross-contamination of fluids
between the electrode channels and the separation chamber.
Raymond et al. [18] were the first to develop a m-FFE chip,
and it operated with closed side channels. This FFE microchip incorporated a separation region, inlet and outlet channels, electrode beds, and a dense array of microchannels
which acted as a membrane, separating the electrodes from
the actual separation chamber, as shown in Fig. 7. This
liquid-filled channel array formed a high hydrodynamic
resistance for the pressure-driven fluid but could conduct the
electrical current. This was a very elegant solution, since all
channel features could be fabricated in one standard siliconetching step. The chip was used to separate fluorescent
markers as well as amino acids during FFZE.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
The same device was used two years later to separate
FITC-labeled proteins and amino acids with FFZE and was
further characterized [20]. The major drawback of this system was the low breakdown voltage of 100–200 V due to the
use of silicon as chip material, limiting the device in its
separation power. Due to low electrical field strength, long
residence times of up to 1 min were required to achieve good
separations. Although the side channels formed a high
hydrodynamic resistance, fluid flow from the side beds into
the separation chamber, negatively influencing the separation, was still observed. Considering the width of the separation chamber (10 mm) and of the side channels arrays
(261 mm), the estimated actual voltage efficiency was
around 80%.
Fonslow and Bowser [36] presented an FFE chip fabricated from two glass plates with an intermediate layer of
amorphous silicon (a-Si) for anodic bonding. This device
contained closed electrode channels, in which gas bubbles
were flushed out by a pressure-driven flow passing the integrated microfabricated gold electrodes. These electrode side
channels were separated by connecting side channel arrays
comparable to those shown in Fig. 7. Due to the side channel
arrays and its electrical resistance, 50% of the applied voltage
was utilized across the separation region in this device. Although the device could withstand higher voltages and fluid
pressure, the successful separation of several fluorescent
dyes was eventually distorted by gas bubbles inside the side
channels, since they could not be removed efficiently anymore.
Fonslow et al. [38] subsequently improved their FFE
concept as mentioned above and presented a device with a
shallow separation region (20 mm deep) and roughly four
times deeper closed electrode beds, completely avoiding side
channel arrays or membrane equivalents. The flow rate
through the electrode channels was now significantly higher,
effectively removing electrolysis products without disrupting
the flow pattern in the shallow separation channel as shown
in Fig. 8 [38]. As a complicating consequence, this approach
www.electrophoresis-journal.com
Electrophoresis 2008, 29, 977–993
Figure 8. (a) A top view of the m-FFE mask of Fonslow et al. [38]
with the following features: 1, separation buffer inlet; 2, sample
inlet; 3, electrode buffer inlets; 4, Au electrodes in electrode
channels; 5, separation channel; 6, electrode buffer outlets; and
7, separation buffer outlet. (b) A side view of the electrode and
separation channels etched into the bottom glass wafer and
bonded to the top glass wafer. Reproduced with permission from
ref. [38], copyright American Chemical Society.
required a precise control over the flow rates and more
equipment. In the new design, 91% of the applied voltage
was utilized across the separation channel where it actually
impacts the separation. Although this multiple-depth m-FFE
device required more fabrication steps, it could be operated
continuously at electric fields (in the separation channel) as
high as 589 V/cm, a four-fold improvement over their previous design. They now found that neither Joule heating, nor
electrolysis product formation, was a limiting factor when
applying high separation potentials. The device was applied
to separate several fluorescent standards with FFZE. The
chip was additionally used for a more detailed study of FFZE
investigating several parameters affecting separation resolution in order to find optimum separation conditions [22].
A third approach to confine the separation region was
achieved by Kobayashi et al. [39] who implemented parallel
shallow side banks (20 mm) in-between the separation channel (30 mm) and the electrode beds (30 mm). Similar to the
side channels discussed before, these shallower regions have
a higher hydrodynamic resistance than the separation
chamber, thus leading the carrier electrolyte flow through
the separation region. Kobayashi et al. [39] presented a
100 mm diameter wafer-sized Pyrex FFE device where they
implemented this so-called bank shape design. The efficiency of the banks shape of the m-FFE was enough to prevent dispersion to the separation chamber of bubbles generated at the electrodes.
3.3 Integrated electrodes inside the separation
chamber
The direct integration of electrodes into the separation
chamber allows for easy flow control and chip layout. However, in order to avoid the disturbance of gas bubbles one has
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Microfluidics and Miniaturization
985
to operate the device with voltages below the electrolysis
potential. Even then the occurrence of electrode reactions
cannot be excluded.
As shown in Fig. 9, Lu et al. [40] used a microfabricated
glass chip with integrated vertical gold electrodes directly
inside the separation channel. They demonstrated FFIEF
of subcellular organelles with voltages applied below the
potential required for electrolysis, therefore completely
avoiding gas formation inside the channel. However, due
to the low electrical field strength available for separation,
long residence times of up to several minutes were
required. Due to this, most likely only components of low
diffusivity, such as cell fractions and organelles could be
separated.
Song et al. [41] presented a microfabricated FFE chip for
IEF without the usage of carrier ampholytes. Instead of
applying an external electrical field required to run the pIbased sorting the field is generated by the diffusion of buffer ions in situ, at the liquid junction between two laminar
flows within the microfluidic channel. However, the
separation could be enhanced by applying an additional low
voltage of 1.3 V using integrated electrodes as demonstrated.
Figure 9. (A) Schematic of field flow IEF of organelles. (B) Photograph of the microfabricated device before final assembly with
(C) enlarged view of the fractionation end of the device. The device consists of electroplated gold electrodes and microfluidic
channels formed in a photopatternable epoxy. Reproduced with
permission from ref. [40], copyright American Chemical Society.
www.electrophoresis-journal.com
986
D. Kohlheyer et al.
3.4 Mechanically and electrically insulated
separation chamber
A different approach to isolate the electrodes from the
separation chamber was investigated by Janasek et al. [42].
Janasek and co-workers faced the question, if it is possible
to achieve a stable electrical field across the FFE separation
chamber by electrostatic induction, like in an ideal DC capacitor. A glass FFE chip was fabricated containing integrated aluminum electrodes shielded from the separation
chamber by 146 mm wide glass walls. If an electrical field
is applied over an insulator, a rearrangement of the mobile
charges in the liquid close to the wall of the insulator will
take place. For example, if a negative potential is applied at
the electrode, the amount of positive charge in the double
layer adjacent to the wall will increase due to the decrease
in the local electrical potential as illustrated, e.g., in the
paper of Schasfoort et al. [43]. After a certain charging time
it would be expected that the entire electrical field drop will
occur over the capacitances formed by the glass walls and
the electrical double layer, and that no electrical field would
be present in chamber any more (see Fig. 10). As a result
separations would be impossible. This might be the explanation of the phenomenon, which was observed in the
presented FFE device under stagnant conditions, when no
flow was applied. However, under flow conditions a stable
FFE separation was observed, with which FFITP of fluorescein was demonstrated while no electrical current was
measured [42]. The authors concluded that the flow is
counteracting the accumulation of charges at the glass
walls, but a clear explanation of the phenomenon was not
given. More detailed experiments are currently performed,
to understand this phenomenon in more detail. Comparable to earlier devices, the voltage efficiency was approximately 50%.
Figure 10. Principle of electrostatic induction by charge displacement caused by dipole orientation. Lower panel: equivalent
circuit diagram. The capacitor C1 and the resistor R1 for the
dielectric barrier are in the range of pF and TV, respectively; C2
and R2 for the liquid compartment are in the range of tens of fF
and a few V, respectively. Reproduced with permission of The
Royal Society of Chemistry from [42].
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Electrophoresis 2008, 29, 977–993
3.5 Device technology – concluding remarks
Several technological approaches for m-FFE devices have
been demonstrated in literature over the past few years, in
which the main challenges were to avoid separation interference due to gas bubbles, to establish a stable electrical
field, and to optimize voltage efficiency. It is difficult to
choose a favorite solution, since many aspects have to be
compared. A brief comparison of the different approaches
can be found in Table 1.
An FFE chip design with open electrode reservoirs
(Fig. 4, category a) allows indeed for easy ventilation of gas
products. However, such an approach puts higher demands
on a proper shielding of the separation chamber. This
shielding structure, usually a kind of membrane substitution has to be of low electrical resistance in order to achieve
good voltage efficiency but also of high hydrodynamic
resistance to prevent for example fluid leakage from the
separation chamber towards the open electrodes reservoirs.
Therefore, several groups implemented so-called side channel arrays which acted as a membrane allowing electrical
current flow but preventing fluid flow. Using integrated side
channel arrays was an elegant solution since they can be
fabricated usually in one step. However, the main drawbacks are a poor voltage efficiency and especially in FFIEF
the formation of a pH gradient inside the side channels. As
a positive side effect, it has been reported, that with the
open characteristics of side channels rather than closed
membranes electrodynamic band broadening could be
avoided. Another technological solution to isolate the
separation chamber from the open electrode reservoirs was
the implementation of gel membranes, usually of acrylamide. This more laborious implementation of conductive
acrylamide membranes has proven to be an efficient method with good voltage efficiency. Acrylamide gel membranes
enable high hydrodynamic resistance with low electrical
resistance and a relatively small width which seems very
promising for further miniaturization. However, the
mechanical stability of the membranes can be the limiting
factor in some designs. Generally, it should be noted that a
continuous replacement of the electrolyte solutions is of
importance for longer separation times, in order to avoid
changes of the solutions electrical properties.
A continuous refreshment of the solutions surrounding
the electrodes seems straightforward when implementing
electrodes into closed channels parallel to the separation
chamber (see also Fig. 4, category b). Additional pumping
equipment is then required to achieve flows in order to
remove gas bubbles efficiently. To avoid crosscontamination between the main flow inside the separation chamber and the electrode electrolyte flows, a proper
flow balancing in this technological approach becomes
more important. In these devices also side channel arrays
have been used, but of lower hydrodynamic resistance.
However, disturbance of the electrical field was reported
due to gas bubbles moving along the electrodes and evenwww.electrophoresis-journal.com
Electrophoresis 2008, 29, 977–993
tually entering and blocking the interconnecting side
channels. Furthermore, without proper fluid and pressure
balancing fluid flow through the side channels occurred.
Not many results have been published on the usage of
shallower side banks, although this design seemed to work
well. The device we reviewed that used this method was
relatively large with 100 mm in diameter. Instead of using
any membrane substitution the usage of closed deep electrode channels was found to be an efficient method to
remove gas bubbles without scarifying voltage efficiency.
This method seems very promising, although it requires
more and precise control of flow rates. However, it has to
be investigated, if the design is also applicable for other
FFE methods such as FFIEF.
Electrodes can also be placed directly inside the separation chamber as shown by some groups (see Fig. 4, category
c). As a result however, only voltages below the electrolysis
potential can be applied. This usually leads to long separation times of up to several minutes making this approach
only suitable for low diffusivity substances. The usage of a
diffusion potential rather than applying an external electrical
field forms an interesting method for pI-based sorting, although the resolution seemed to be low.
The technological approach where a mechanical and
electrical isolation of electrodes from the separation chamber
(here electrostatic induction, see also Fig. 4, category d) is
used, has to be studied more intensively and a clear explanation and understanding of the phenomenon should be
found, but might offer an interesting alternative to other
methods.
4
Separation results
As mentioned, the four standard modes of FFE include zone
electrophoresis, IEF, ITP and, field step electrophoresis. All
modes have been applied in m-FFE devices, generally as
proof-of-principle. In this section, we review the separation
results obtained.
4.1 FFZE
Raymond et al. [18] were the first to publish results on FFZE
applying their microfabricated FFE device. The potential of
the system was demonstrated by the separation of three rhodamine-B isothiocyanate-labeled amino acids, namely lysine,
glutamine, and glutamic acid as shown in Fig. 11. A residence time of 73 s was needed to separated the components
when 50 V were applied, which resulted in an electrical field
strength (estimated by the authors) of 3 V/mm (I = 7.5 mA).
Raymond et al. [18] found that the separation resolution was
controllable by varying the side bed electrical conductivity,
which was mainly due to the large contribution of the side
channels to the total electrical resistance. Higher side bed
conductivity resulted in an improved resolution, as the voltage drop across the separation chamber increased. EOF did
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Microfluidics and Miniaturization
987
Figure 11. Intensity plot of three labeled amino acids continuously separated by FFZE with a residence time of 73 s when
50 V were applied. Reproduced with permission from ref. [18],
copyright American Chemical Society.
not significantly affect the experimental baseline widths.
This was probably a consequence of using side channels
rather than closed membranes for the isolation of the
separation chamber. However, this more open characteristic
of the side channels resulted in flow from the side beds into
the separation chamber, affecting separation. Eventually the
bubbles created during electrolysis filled up the side beds
resulting in a loss of the electrical field. It was reported, that
no apparent Joule heating was observed.
For a more detailed separation study, Raymond et al. [20]
used the same chip design for the continuous separation of
high molecular weight compounds with FFZE. In fact, they
separated a mixture of FITC-labeled HSA, bradykinin, and
ribonuclease A. An electrical field strength of 100 V/cm and a
residence time of 62 s was needed to fully separate these
components. Furthermore, the authors investigated different
sources of band broadening and found that, initial bandwidth, diffusion, and hydrodynamic band broadening were
the main contributors to the band broadening. Despite some
problems, Raymond et al. [20] reported a peak capacity of
8 bands/cm. This was in the same range as reported for
conventional systems (10 bands/cm), indicating that miniaturized FFE devices could achieve similar separation results.
They also demonstrated that the chip is capable of separating
more complex samples such as diluted rat plasma and trypic
digests of bovine cytochrome C and melittin.
Kobayashi et al. [39] demonstrated FFZE of two native
proteins, cytochrome C and myoglobin. The residence time
for a complete separation was 10 s when 2 kV was applied
(estimated by the authors: E = 33 V/mm, I = 0.7 mA). The
separated analytes were collected at different outlets and analyzed offline by RP HPLC. The authors applied a hydroxymethylcellulose (HPMC) coating in order to reduce the glass
www.electrophoresis-journal.com
988
D. Kohlheyer et al.
surface charge, trying to minimize EOF. Although the EOF
was suppressed, the coating efficiency was not stable during
longer separations. The device was also used for a theoretical
study on the temperature distribution and Joule heating [44].
Zhang and Manz [32] presented a m-FFE chip for highspeed separations of analytes. With their improved chip
design, they achieved a full separation of the fluorescent dyes
rhodamine 110 and fluorescein within 75 ms. The total residence time (sample flow rate was 2 nL/s) in the separation
chamber was 2 s when 1.75 kV was applied (E = 13.5 V/mm,
I = 0.14 mA). As shown in Fig. 12, the negatively charged
fluorescein (left stream) is slightly deflected towards the
negative electrode, which was caused by EOF, as Zhang and
Manz [32] concluded. The chip has also been used for the
separation of FITC-labeled amino acids in both aqueous and
binary media.
FFZE of a mixture of fluorescent dyes, namely fluorescein, rhodamine 110, a rhodamine 110 impurity, and rhodamine 123 was presented by Fonslow and Bowser [36]. With
the application of 515 V (I = 310 mA), resulting in an electrical field strength inside the separation chamber of 25.9 V/
mm, the three dyes as well as the impurity clearly separated
with a residence time of 9.6 s before detection. In agreement
with previous publications, Joule heating was found not to be
significant at the studied field strengths.
Fonslow et al. [38] further developed their original FFE
chip design in order to apply higher voltages and improve
voltage efficiency. As before, they separated fluorescein, rhodamine 110, rhodamine 110 impurity, and rhodamine 123 by
FFZE. With a linear flow velocity of 5 mm/s and electrical
field strength up to 58.6 V/mm an improved and more stable
separation compared to their previous device was observed.
With separation voltages of over 645 V Joule heating became
Figure 12. FFZE of rhodamine 110 and fluorescein with a flow
velocity of 6 mm/s. The resulting electrical field strength was
135 V/cm with an induced current of 140 mA. The right separated
stream was rhodamine-110 and the left fluorescein. Reproduced
with permission from ref. [32], copyright American Chemical
Society.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Electrophoresis 2008, 29, 977–993
significant. The same chip was used for a more theoretical
study to optimize bandwidth and separation resolution in
FFZE [22].
Kohlheyer et al. [21] presented FFZE of fluorescein and
rhodamine B. Both analytes were separated within a total
residence time of 3.3 s when 180 V was applied (E = 25 V/
mm, I = 50 mA). The linear flow velocity was 3 mm/s. The
precise control of two sheath flow streams enabled positioning of the separated components within the separation
chamber. Separated components could be steered towards
different outlets by varying the sheath flow rates without
altering the separation voltage, as shown in Fig. 13. This
steering technique can be used in future devices to purify a
sample mixture in such a way that only the separated component of interest is steered to a specific outlet. Steering of
fluorescently labeled, easy visible components seems
straightforward. However, it is questionable how to control
the precise position in more realistic label-free applications.
In order to demonstrate the feasibility of their FFE chip,
de Jesus et al. [37] separated a mixture of the ionic dyes bromophenol, brilliant blue and crystal violet. The separation
was carried out at a linear flow velocity of 0.54 mm/s with a
total residence time of 18.2 s. 300 V were applied to achieve
an electrical field strength inside the separation chamber of
app. 11.5 V/mm. During 3 h of evaluation under a maximum
current of 706 mA a stable separation was observed.
4.2 FFIEF
The first to publish results on microfluidic FFIEF were Xu et
al. [33]. The authors demonstrated the IEF of fluorescently
labeled angiotensin by a factor of 400 with 430 s focusing
Figure 13. Steering of rhodamine B and fluorescein bands in
FFZE. The sample stream is hydrodynamically focused by two
parallel sheath flow streams. By adjusting the flow rates F1 and
F2 the sample injection position is shifted. Reproduced by permission of The Royal Society of Chemistry from [21].
www.electrophoresis-journal.com
Electrophoresis 2008, 29, 977–993
time. They applied 1750 V to realize an electrical field
strength of 13.5 V/mm, as shown in Fig. 14. To generate the
required pH gradient an ampholyte solution was applied.
However, due the technical layout, as discussed before, only a
part of the pH gradient was available for separation since the
low-pH and high-pH ampholytes migrated into the side
channels. The chip could be operated between 7.7 and 30 V/
mm before Joule-heating became significant and bubbles
appeared.
Continuous IEF of large molecules such as sub-cellular
organelles was presented by Lu at al. [40]. In order to avoid
electrolysis low voltages were applied. However, due to this
procedure, long residence times were necessary in order to
reach stable focusing. The method is therefore suited to
separate components with low diffusion. Among other
results, FFIEF of stained mitochondria from a cell lysate was
shown. For this purpose an ampholyte-based pH gradient
(pH 3–6) was utilized and 2 V were applied with a residence
time of 6 min. Especially with its mild voltage conditions this
method is suitable for more sensitive biological components.
Using the same device as used for FFZE, Kohlheyer et al.
[21] applied FFIEF for the separation of several fluorescent
IEF markers. IEF markers with pI0 s at pH 4.5, 5.5, 7.6, and
8.7 were separated and focused. With a residence time at the
point of measurement of 3 s (linear flow velocity 2 mm/s)
and the application of 20 V (approximately 10 V/mm inside
the separation chamber) all components fully focused at
their pI0 s within a 500 mm wide pH gradient (pH 3–10). The
full pH gradient was available for separation since outer
sheath flows of high pH and low pH, respectively, formed a
border for arranging ampholytes.
Kohlheyer et al. [23] reported later an improved FFIEF
chip based on their previous publication. Here, a preseparated ampholyte buffer was used which reduced the
electrical current and the focusing time. FFIEF of various
Figure 14. IEF of fluorescently labeled angiotensin II: Ang
II = 10 mM, U = 1750 V; E = 135 V/cm. Reproduced with permission of The Royal Society of Chemistry from [33].
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Microfluidics and Miniaturization
989
IEF markers ranging from pI 3 to 10 and the protein HSA
was demonstrated, as shown in Fig. 15. The authors
determined, that the device is capable of separating analytes with a minimum difference in pI of D(pH) = 0.23.
This resulted in a theoretical peak capacity of 29 peaks
within 1.8 mm for a pH gradient pH 2.5–11.5. With optimized pI markers, commonly used for calibration, the
peak capacity of conventional FFE systems is between 50
and 70 peaks within 70 mm separation width (personal
communication, Dr. Gerhard Weber, BD Diagnostics). In
terms of resolution, this would mean for a pH gradient
(pH 2.5–11.5) a D(pI)min between 0.13 and 0.18 pH units.
To achieve this, in the conventional FFE system usually
voltages of up to 2 kV have to be applied, while Kohlheyer
et al. applied only 150 V, clearly indicating the potential
advantages of microfluidic FFIEF devices. In practice, the
resolution is reduced when separating proteins and peptides [45]. This comparison shows that with m-FFIEF one
can achieve comparable and even better resolution with
applying much lower voltages.
Instead of using sheath flow streams to confine the pH
gradient, Albrecht and Jensen [24] used functionalized gel
membranes with low and high pH buffering capacity,
respectively. FFIEF of several fluorescent IEF markers
(pI 3.5, 5.1, 7.2, and 7.6), with 200 V separation voltage and a
residence time of 14 s, was shown. Although a lower resolution than in the publication of Kohlheyer et al. [21] was
achieved, the device involved less control over fluid streams
Figure 15. FFIEF of seven fluorescent IEF markers: When 150 V
voltage (I = 50 mA) was applied, the markers (pI 4, 5.1, 6.2, 7.2, 8.1,
9, and 10.3) fully separated within less than 2 s. The sample flow
rate was 0.4 mL/min (v = 2 mm/s). The apparent kinks in the fluorescent tracer paths are caused by merging multiple photographs. Reproduced with permission from ref. [23], copyright
American Chemical Society.
www.electrophoresis-journal.com
990
D. Kohlheyer et al.
Electrophoresis 2008, 29, 977–993
and pumps. In contrast to previous results Albrecht and
Jensen [24] were the first to report the need for a thermal
cooling system to optimize their separation resolution.
Therefore, a thermoelectric element was used to cool the
top of the FFE device. A possible explanation is the lower
heat dissipation of the chosen material (PDMS) compared
with glass devices [46]. Recently, Albrecht et al. [47] also
reported a cascaded FFIEF device based on their earlier
published results. The streams of separated components
from a first FFIEF separation chamber were guided into a
second FFIEF stage resulting in an improved separation
resolution. Their FFIEF device was used to focus native
model proteins, denatured proteins, as well as protein
complexes.
Song et al. [41] demonstrated successful FFIEF of two
different pI markers. This device operated differently than
other FFIEF systems, since neither ampholytes nor an external voltage has to be applied. The electrical potential gradient
was achieved by diffusion potential in a separation channel
of around 100 mm in width.
4.3 FFITP
Janasek et al. [34] were the first to report on microfluidic
FFITP, and demonstrated the method by separating fluorescein, acetylsalicylic acid (ASS), and Eosin G, as shown in
Fig. 16. Since in ITP the bands flow adjacent to each other,
for a clear experimental validation ASS was chosen as a
nonfluorescent spacer analyte between the two fluorescent
components, fluorescein and Eosin G. For the experiment
the electrical field strength was 21 V/mm. At higher field
strength movement of the analytes into the side channels
was observed.
As a proof-of-principle, Janasek et al. [42] applied FFITP
of fluorescein to investigate their principle of electrostatic
induction for FFE. They successfully focused fluorescein
with the application of 150 V, which corresponds to an electrical field strength of 18 V/mm inside the separation chamber, and a total flow rate of 20 mL/min.
Figure 16. FFITP separation and concentration of fluorescein,
ASS, and eosin G displayed as an assembly of single micrographs over the whole chamber. Reproduced with permission
from ref. [34], copyright American Chemical Society.
4.4 FFFSE
To the best of our knowledge, no results on microfluidic
FFFSE have been published until now. To demonstrate the
principle of FFFES, the authors separated a mixture of rhodamine B and fluorescein using their m-FFE device. For this
purpose, the sample stream was hydrodynamically focused
in-between two sheath flow streams of higher electrical conductivity. As shown in Fig. 17, this resulted in an electrical
field step when 150 V were applied. The negatively charged
fluorescein migrated towards the anode and eventually concentrated (approximately four-fold) at the field step due to a
decrease in its mobility. The neutrally charged, nonmigrating Rhodamine B showed only a slight band broadening
maintaining its centered position.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Figure 17. FFFSE demonstrated in a microfluidic device of the
authors. (a) Mixture of rhodamine B and fluorescein continuously
flowing with no voltage applied. (b) When 150 V were applied the
negatively charged fluorescein migrated and focused due to the
sudden decrease in electrical field strength (Kohlheyer et al.,
unpublished result).
www.electrophoresis-journal.com
Electrophoresis 2008, 29, 977–993
4.5 Separation results – concluding remarks
As shown by various publications, the four know separation
modes, FFZE, FFIEF, FFITP, and FFFSE, have successfully
been demonstrated in microfluidic devices. A wide spectrum
of analytes has been separated including cell fragments,
organelles, proteins, amino acids, fluorescent dyes, and
markers. The results shown until now, which can be considered as proof-of-principle experiments, are difficult to
compare in terms of resolution, bandwidth, and peak capacity, since too many parameters differ from experiment to
experiment. Obviously, also predicted by scaling laws,
separation times were drastically reduced by miniaturization, making it more interesting for online monitoring tools
rather than conventional systems meant for sample preparation. Samples were fully separated within several seconds
and below. Furthermore, sample flow rates of several microliters per minute and below were applied. Generally one can
say that Joule heating, a major concern in conventional FFE
systems seems to be less relevant on a microfluidic scale.
Due to the lack of lateral separation length higher electrical
field strengths have to be applied in order to achieve comparable separation resolution and peak capacity as on the
large scale. This is possible, because due to the lower electrical currents involved (typically in the microampere range)
less heat is generated, and because of the higher surface-tovolume ratios that favor fast heat dissipation. It would be
beneficial for the field to investigate in more detail scaling
effects by matching experiments performed on the microand macroscale. Optical detection methods, usually involving large fluorescent microscopes and optical equipment,
until now have been sufficient for first characterizations.
However, as proposed already by Raymond et al. [20], the
major advantage of a m-FFE device is the possibility to integrate it more easily in a complex analysis system. Such an
integration for online monitoring applications of separated
analytes is not a straightforward task, as will be discussed in
the following section.
5
Detection and hyphenation
Size-reduced conventional mini-FFE devices have already
been implemented into hyphenated analytical systems. Such
systems can guide developers of integrated m-FFE systems,
and we, therefore, mention a few examples.
Chartogne et al. [7] coupled an FFE device to CIEF as a
cleaning step to remove interfering ampholyte substances
prior to ESI-MS to improve the MS sensitivity. This CIEFFFE-ESI-MS system was successfully realized for three
model proteins (myoglobin, carbonic anhydrase I, and b-lactoglobulin B). Furthermore, Mazereeuw et al. [6] realized a
system in which FFE was used as an online separation step
of fluorescent compounds in a biochemical detection system.
This system was based on the interaction of the affinity protein streptavidin with biotin, in which biotin acted as an
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Microfluidics and Miniaturization
991
analyte. This FFE detection system was coupled to HPLC for
the bioanalysis of biotin in human urine.
In the field of m-FFE systems not much progress has
been shown on the implementation of FFE with other
separation or detection systems into so-called micro total
analysis systems (m-TAS) or LOC devices [8, 9, 48, 49]. In fact,
until now only propositions for more integrated systems
have been published.
Zhang and Manz [32] concluded from their investigations, that m-FFE presents a potentially powerful tool in proteomics studies. They proposed coupling m-FFE with electrospray MS as an online automated analytical tool in proteomics studies. By moving the chip perpendicular to the
outlets separated components spray from respective channels and analyte streams are then received by the MS in
scanning mode. Although they found that only little modifications are necessary for this detection approach, to our
knowledge no results have been published until now. Instead
of scanning the entire chip, the capability to steer the analyte
by hydrodynamic focusing during FFZE separation, as
shown by Kohlheyer et al. [21], might offer an interesting
alternative to scan separated components through a single
chip outlet connected to an ESI-MS system. A step towards
this approach was the coupling of a glass microchip for CE to
ESI-MS as, e.g., successfully demonstrated by Hoffmann et
al. [50] and others [51]. Furthermore, Zhang and Manz [32]
proposed the use of m-FFE combined with a microreactor, for
on-chip synthesis and separation of products. Using microchip CE, on-chip synthesis and separation of products was
shown by Belder et al. [52]. In this initial study, the device was
successfully applied for testing biocatalysts created by directed evolution of enzymes. The use of m-FFE instead of
microchip CE could enable high throughput and continuous
screening of this reaction rather than the analysis of discrete
plugs achieved by microchip CE.
An integrated system approach combining m-FFE and
surface plasmon resonance (SPR) detection was proposed by
Kohlheyer et al. [53]. As illustrated in Fig. 18, a thin gold
detection area is integrated into the FFE separation chip,
located after the separation chamber. A monochromatic ppolarized light beam is coupled into the chip device and
reflected under total internal reflection at the gold surface.
Refractive index changes, near the gold surface, e.g., caused
by immobilization of molecules on the gold surface lead to
local changes in the reflected light intensity. These light
intensity changes can be monitored for the whole detection
region simultaneously by a CCD camera. For more information about SPR and its theoretical background one is referred
to the review of Homola [54]. For experiments the proposed
chip is placed inside an SPR imaging instrument (IBIS
Technologies, The Netherlands) to monitor immobilization
of separated fractions. Eventually this generated fraction
pattern is used in a second step to study biomolecular interactions in the search for specific biomarkers present in, e.g.,
patients sample. Collected fractions of immobilized lanes,
which positively react with specific antibodies, will be anawww.electrophoresis-journal.com
992
D. Kohlheyer et al.
Electrophoresis 2008, 29, 977–993
as reviewed in this paper will further be developed for new
applications in life sciences. Broadening the options of parallelization and assay implementation, including sample treatment
on-a-chip will certainly contribute to an increased number of
publications in the near future. Definitely a total analysis system
approach instead of a device-based approach will lead to a
growth in the field of m-FFE.
We would like to acknowledge the funding of this research by
the Dutch Technology Foundation STW. Furthermore, Dr. Gerhard Weber and Professor Petr Boček are acknowledged for help
with theoretical aspects.
The authors have declared no conflict of interest.
7
References
[1] Hannig, K. Z., Anal. Chem. 1961, 181, 244–254.
[2] Bauer, J. (Ed.), Special Issue on Free-Flow Electrophoresis,
Electrophoresis 1998, 19, 1057–1235.
Figure 18. Illustration of the FFE microchip with an integrated
gold area for SPR detection. (a) The chip is placed inside the SPR
imaging instrument with a monochromatic light beam coupled
into the chip device. (b) Cross-section of the chip detection
region: biomolecular interaction and molecule immobilization
cause refractive index changes which can be monitored real time
by local light intensity changes [53].
lyzed offline by MS for identification purpose. However,
until now only preliminary results were presented and more
work is currently performed.
[3] Hannig, K., Heidrich, H. G., Free-Flow Electrophoresis, GIT
Verlag, Darmstadt 1990.
[4] Weber, G., Boček, P., Electrophoresis 1998, 19, 1649–1653.
[5] Weber, P. J. A., Weber, G., Eckerskorn, C., in: Simpson, R. J.
(Ed.), Purifying Proteins for Proteomics: A Laboratory Manual, Cold Spring Harbor Laboratory Press, New York 2003.
[6] Mazereeuw, M., de Best, C. M., Tjaden, U. R., Irth, H., van der
Greef, J., Anal. Chem. 2000, 72, 3881–3886.
[7] Chartogne, A., Tjaden, U. R., van der Greef, J., Rapid Commun. Mass Spectrom. 2000, 14, 1269–1274.
[8] Tudos, A. J., Besselink, G. A. J., Schasfoort, R. B. M., Lab
Chip 2001, 1, 83–95.
[9] Dittrich, P. S., Tachikawa, K., Manz, A., Anal. Chem. 2006, 78,
3887–3908.
6
Conclusions
[10] Manz, A., Eijkel, J. C. T., Pure Appl. Chem. 2001, 73, 1555–
1561.
[11] Das, C., Fan, Z. H., Electrophoresis 2006, 27, 3619–3626.
In FFE, an electrical field is applied perpendicular to the flow
direction and charged molecules are deflected from the carrier
electrolyte flow direction, in a way that is controlled by the electrophoretic mobility, the flow velocity, and the electrical field
strength. The research field of m-FFE methods is rather small, as
evident from the number of papers reviewed. However, it carries
great promise due to the continuous flow of separated sample
enabling integration and continuous monitoring in miniaturized analysis systems. In this paper, we showed that steady
innovation is occurring in this research field. The results shown
so far indicated that with m-FFE systems one can achieve high
separation resolutions which are comparable to or even better
than those reached in their larger counterparts. The microfluidic devices furthermore accomplish this separation in a
fraction of the time the larger devices need, and consuming a
fraction of the sample volume. A trend can be foreseen in which
the integration of LOC devices using one of the FFE principles
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
[12] Pamme, N., Lab Chip 2007, 7, 1644–1659.
[13] Baker, D. R., Capillary Electrophoresis, John Wiley and Sons
Inc., New York 1995.
[14] Křivánková, L., Boček, P., Electrophoresis 1998, 19, 1064–
1074.
[15] Roman, M. C., Brown, P. R., Anal. Chem. 1994, 66, 86A–94A.
[16] Kašička, V., Prusík, Z., Pospísek, J., J. Chromatogr. 1992, 608,
13–22.
[17] Kašička, V., Prusík, Z., Sazelova, P., Jiracek, J., Barth, T., J.
Chromatogr. A 1998, 796, 211–220.
[18] Raymond, D. E., Manz, A., Widmer, H. M., Anal. Chem. 1994,
66, 2858–2865.
[19] Gebauer, P., Boček, P., Electrophoresis 2005, 26, 453–462.
[20] Raymond, D. E., Manz, A., Widmer, H. M., Anal. Chem. 1996,
68, 2515–2522.
[21] Kohlheyer, D., Besselink, G. A. J., Schlautmann, S., Schasfoort, R. B. M., Lab Chip 2006, 6, 374–380.
www.electrophoresis-journal.com
Electrophoresis 2008, 29, 977–993
Microfluidics and Miniaturization
993
[22] Fonslow, B. R., Bowser, M. T., Anal. Chem. 2006, 78, 8236–
8244.
[38] Fonslow, B. R., Barocas, V. H., Bowser, M. T., Anal. Chem.
2006, 78, 5369–5374.
[23] Kohlheyer, D., Eijkel, J. C. T., Schlautmann, S., van den Berg,
A., Schasfoort, R. B. M., Anal. Chem. 2007, 79, 8190–8198.
[39] Kobayashi, H., Shimamura, K., Akaida, T., Sakano, K. et al., J.
Chromatogr. A 2003, 990, 169.
[24] Albrecht, J. W., Jensen, K. F., Electrophoresis 2006, 27, 4960–
4969.
[40] Lu, H., Gaudet, S., Schmidt, M. A., Jensen, K. F., Anal. Chem.
2004, 76, 5705–5712.
[25] Wang, Q., Tolley, H. D., LaFebre, D. A., Lee, M. L., Anal.
Bioanal. Chem. 2002, 373, 125–135.
[41] Song, Y. A., Hsu, S., Stevens, A. L., Han, J., Anal. Chem.
2006, 78, 3528–3536.
[26] Righetti, P. G., Isoelectric Focusing: Theory, Methodology
and Applications, Elsevier Biomedical, Amsterdam 1983.
[42] Janasek, D., Schilling, M., Manz, A., Franzke, J., Lab Chip
2006, 6, 710–713.
[27] Boček, P., Deml, M., Gebauer, P., Dolník, V., Analytical Isotachophoresis, Verlag Chemie, Weinheim 1988.
[43] Schasfoort, R. B. M., Schlautmann, S., Hendrikse, J., van den
Berg, A., Science 1999, 286, 942–945.
[28] Everaerts, F. M., Beckers, K. L., Verheggen, T. P. E. M., Isotachophoresis – Theory, Instrumentation and Applications,
Elsevier, Amsterdam 1976.
[44] Matsumoto, H., Komatsubara, N., Kuroda, C., Tajima, N. et
al., Chem. Eng. J. 2004, 101, 347–356.
[29] Gebauer, P., Boček, P., Electrophoresis 1995, 16, 1999–2007.
[45] Malmstrom, J., Lee, H., Nesvizhskii, A. I., Shteynberg, D. et
al., J. Proteome Res. 2006, 5, 2241–2249.
[30] Kuhn, R., Hoffstetter, S., Wagner, K. H., Electrophoresis 1990,
11, 942–947.
[46] Fu, R., Xu, B., Li, D., Int. J. Therm. Sci. 2006, 45, 841–847.
[31] Stone, V. N., Baldock, S. J., Croasdell, L. A., Dillon, L. A. et al.,
J. Chromatogr. A 2007, 1155, 199–205.
[32] Zhang, C.-X., Manz, A., Anal. Chem. 2003, 75, 5759–5766.
[33] Xu, Y., Zhang, C. X., Janasek, D., Manz, A., Lab Chip 2003, 3,
224–227.
[34] Janasek, D., Schilling, M., Franzke, J., Manz, A., Anal. Chem.
2006, 78, 3815–3819.
[35] Takamura, Y., Onoda, H., Indokuchi, H., Oki, A., Horiike, Y.,
Electrophoresis 2003, 24, 185–192.
[36] Fonslow, B. R., Bowser, M. T., Anal. Chem. 2005, 77, 5706–
5710.
[37] de Jesus, D. P., Blanes, L., do Lago, C. L., Electrophoresis
2006, 27, 4935–4942.
2008 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
[47] Albrecht, J. W., El-Ali, J., Jensen, K. F., Anal. Chem. 2007, 79,
9364–9371.
[48] Schasfoort, R. B. M., Expert Rev. Proteomics 2004, 1, 123–
132.
[49] Janasek, D., Franzke, J., Manz, A., Nature 2006, 442, 374–380.
[50] Hoffmann, P., Häusig, U., Schulze, P., Belder, D., Angew.
Chem. Int. Ed. 2007, 46, 4913–4916.
[51] Koster, S., Verpoorte, E., Lab Chip 2007, 7, 1394–1412.
[52] Belder, D., Ludwig, M., Wang, L.-W., Reetz, M. T., Angew.
Chem. Int. Ed. 2006, 45, 2463–2466.
[53] Kohlheyer, D., Besselink, G. A. J., Schlautmann, S., Schasfoort, R. B. M., Nanotech 2006, Montreux, Switzerland 2006.
[54] Homola, J., Anal. Bioanal. Chem. 2003, 377, 528–539.
www.electrophoresis-journal.com