A SU-8 liquid cell for surface acoustic wave biosensors
L. A. Francisa,b , J.-M. Friedtc , C. Barticb and A. Campitellib
a
PCPM, Université catholique de Louvain, Croix du Sud 1, B-1348 Louvain-la-Neuve,
Belgium;
b Biosensors Group, IMEC, Kapeldreef 75, B-3001 Leuven, Belgium;
c LMN, Université de Franche-Comté, La Bouloie, 25030 Besançon, France
ABSTRACT
One significant challenge facing biosensor development is packaging. For surface acoustic wave based biosensors,
packaging influences the general sensing performance. The acoustic wave is generated and received thanks to
interdigital transducers and the separation between the transducers defines the sensing area. Liquids used in
biosensing experiments lead to an attenuation of the acoustic signal while in contact with the transducers. We
have developed a liquid cell based on photodefinable epoxy SU-8 that prevents the presence of liquid on the
transducers, has a small disturbance effect on the propagation of the acoustic wave, does not interfere with the
biochemical sensing event, and leads to an integrated sensor system with reproducible properties. The liquid
cell is achieved in two steps. In a first step, the SU-8 is precisely patterned around the transducers to define
120 µm thick walls. In a second step and after the dicing of the sensors, a glass capping is placed manually
and glued on top of the SU-8 walls. This design approach is an improvement compared to the more classical
solution consisting of a pre-molded cell that must be pressed against the device in order to avoid leaks, with
negative consequences on the reproducibility of the experimental results. We demonstrate the effectiveness of
our approach by protein adsorption monitoring. The packaging materials do not interfere with the biomolecules
and have a high chemical resistance. For future developments, wafer level bonding of the quartz capping onto
the SU-8 walls is envisioned.
Keywords: surface acoustic wave, biosensor, liquid cell, SU-8, packaging, protein, fibrinogen
1. INTRODUCTION
Initially developed for filters and delay lines in the telecommunication field, surface acoustic wave (SAW) devices
have also attracted a growing interest in the sensor field.1 The sensing principle is based on the perturbation
of an elastic field propagating along the surface of a planar substrate. When the SAW is shear polarized, like
SH-SAW and Love-mode sensors, the acoustic energy loss is low if a liquid is present on the surface.2, 3 These
sensors are adequate as biosensors where the perturbation is caused by the adsorption of biomolecules present in
solution.4, 5 The biochemical recognition results in a SAW velocity change that is monitored in real-time as an
electrical delay phase through the transducers. The transduction is obtained on piezoelectric substrates thanks
to interdigital transducers6 (IDTs) that consist in a set of interdigitated electrodes with alternative electrical
polarity. The separation between the IDTs defines the sensing area as sketched in Figure 1.
The transfer function of the device, which relates the amplitude of an AC electrical signal at the output
transducer to its value at the input transducer, is affected when a liquid with a high dielectric permittivity value
is partly or fully contacting the transducers.7 As example, Figure 2 shows the effect of deionized water in contact
with any of the transducers on a quartz substrate. The attenuation observed in the amplitude and the phase of
the transfer function is mainly due to an extra capacitance loading of the transducers; more details related to this
effect are given in Appendix A. The intensity of the attenuation is function of the value of the electromechanical
coupling constant K 2 of the piezoelectric substrate. It is more intense for low K 2 substrates such as quartz than
for high K 2 substrates such as lithium niobate or lithium tantalate. For Love-mode sensors, it also depends of
the thickness of the guiding dielectric overlayer that acts as shielding and reduces the interaction with liquids.
Further author information:
L. A. Francis: E-mail:
[email protected]
SPIE USE, V. 1 5455-43 (p.1 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
INPUT IDT
OUTPUT IDT
SENSING
AREA
W
SUBSTRATE
D
L
Figure 1. Schematic representation of the surface acoustic wave (SAW) biosensor seen from top. The acoustic signal
is generated via interdigital transducers with a mid-distance separation L and with an aperture W corresponding to the
overlap length of the interdigitated electodes. The space left between the edges of the transducers is the sensing part with
an area D × W .
Amplitude [dB]
−20
−30
−40
−50
−60
123
123.5
124
124.5
125
125.5
Frequency [MHz]
126
126.5
127
123.5
124
124.5
126
126.5
127
200
Phase [ ° ]
100
0
−100
−200
123
125
125.5
Figure 2. Experimental record of the transfer function of a quartz substrate SAW device showing the effect of water
on the transducers. The top graph is the amplitude and the bottom graph is the phase of the electrical signal passing
through the acoustic device as a function of the frequency. The solid line is the transfer function in air while the dashed
line is the device loaded with water on one or on the two transducers, either partly or totally covering the electrodes of
the transducers.
SPIE USE, V. 1 5455-43 (p.2 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
In any cases, the signal attenuation affects the general performance of the sensor from the instrumentation
and the sensing aspects. It results in a lack of correlation between surface changes caused by the biomolecules
adsorption and the SAW velocity changes measured by phase shifts in the transfer function. Therefore, one
desires a method for the efficient shielding of the IDTs against liquids that serves in the same time as liquid
holder, namely a liquid cell. Examples of liquid cell are reported in the literature.8–11 Typically, they consist
in a pre-molded polydimethylsiloxane (PDMS) cell or in a mechanically drilled Teflon cell that comes in close
contact with the device. The cell is mechanically pressed against the surface of the sensor in order to avoid
leaks. Despite the flexibility allowed by external liquid cells, these methods miss an integration step at the
device processing scale. A liquid cell integrated during the fabrication of SAW biosensors will certainly open and
enhance their application. A well designed liquid cell maximizes the sensing surface and minimize interactions
with the acoustic wave and with the biochemical species.
In this paper, we address the design and the fabrication of a liquid cell that allows the packaging of SAW
biosensors. The liquid cell aims at three characteristics:
• it prevents the presence of liquids on the transducers;
• it has a small disturbance effect on the propagation of the surface acoustic wave;
• it does not interfere with the biochemical sensing event.
2. DESIGN OF THE LIQUID CELL
Our approach is based on the protection of the transducers and of every area out of the acoustic sensing area
by a vertical structure above the IDTs. The protecting structure is made of a polymer spacer defined between
the sensor surface and a rigid capping layer atop of the spacer. Microfabrication techniques are required for
the fabrication of the liquid cell. They allow to reduce the lateral dimensions of polymers that cause acoustic
attenuation while in contact with the surface. Therefore, these techniques help to minimize interactions between
polymers and the acoustic wave. In addition, they also allow the integration of the packaging at the wafer
processing level with reproducible characteristics that address the micron-range scale of the sensor.
Four aspects have been taken in consideration in the design of the liquid cell: (1) the analyte solution must be
correctly displayed on the sensing area; (2) the rigid capping layer above the IDTs must not come in contact with
the electrodes to avoid a damping of the acoustic wave and their capacitive loading; (3) polymers as spacer cause
an acoustic attenuation, consequently these materials must cover the smallest length of the acoustic path on the
sensing area; and (4) the sensing area must be kept open for its cleaning prior to any biochemical recognition
experiment, for instance with an UV-O3 cleaning treatment.
Additionally, the packaging is required to withstand the cleaning procedure without contaminating the biochemical solutions. This involves the use of materials presenting good chemical resistance to organic solvents and to
UV-O3 surface treatment.
The SU-8 is a photodefinable epoxy selected to perform the spacer function.12
adequate properties:
Its selection is based on
• photosensitive that allows for a precise patterning of its structure around the IDTs by using standard
photolithography techniques;
• high thicknesses attainable (up to 500 µm in one step, much thicker than other photosensitive encapsulant
materials like BCB and PMMA) with a high aspect ratio about 1:25;
• stiff epoxy leading to smaller absorption of the acoustic signal than soft polymers;
• good chemical resistance to a wide range of organic solvents and of surface treatments;
• low thermal budget and good adhesion on various materials.
SPIE USE, V. 1 5455-43 (p.3 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
CROSS SECTION AA
(A)
A
A
SUBSTRATE
(B)
(C)
GLASS
(D )
Figure 3. Schematic of the fabrication process. On each drawing, the left side is a top view of the device and the right
side is the cross section AA. (A) Initial SAW device is a part of the wafer. (B) SU-8 walls are patterned partly around the
transducers and the wafer is diced to release each sensor. (C) The sensor is attached to the printed circuit board and the
contact pads are wire bonded to it (step not shown), the top part of the SU-8 walls is manually covered with an epoxy
glue, immediately followed by the placement of the quartz glass capping and the curing of the epoxy. (D) The left parts
are filled with the same epoxy glue to complete the protection and the packaging of the device. The liquid cell defines a
well of precise dimensions above the sensing area of the packaged sensor.
The SU-8 spacer shaped as walls around the IDTs performs a continuous physical separation between the
transducers and the sensing area. Because of its position above the acoustic path, this separating wall must be
thin to minimize the acoustic damping and wide enough to assure a continuous and rigid physical contact with
the sensor. In the present case, a compromise has been found with a 150 µm wide wall. Outside the acoustic
path, the walls are larger to increase the adhesion surface between the SU-8 and the sensor on one side, and with
the capping layer on the other side.
As capping material, quartz is selected because of its mechanical stiffness, its optical transparency that allows
to visually check for the unwanted presence of liquids above the capped transducers, its electrical insulation to
avoid interaction with the electromagnetic field emitted by the IDTs, and finally its resistance to a numerous
amount of chemical products and surface treatments.
3. FABRICATION OF THE LIQUID CELL
We have used a two steps method to fabricate the liquid cell. The first step is achieved at the wafer level. It
consists in the patterning of the SU-8 walls. The second step is achieved at the device level after the dicing of the
wafer. It consists in the manual placement of the quartz capping. The fabrication methodology is schematically
depicted in Figure 3 and described for a particular example here after. The fabrication method can be extended
to various types of substrates and transducers designs without requiring extra developments.
The liquid cell was fabricated on a Love-mode SAW devices made on a 3 inches diameter and 500 µm thick
ST-cut (42.5◦ Y-cut) quartz substrate (China National Scientific Instruments and Materials Corp., Hangzhou,
P. R. of China) covered with a 1.2 µm thick silicon dioxide layer obtained by plasma-enhanced chemical vapor
deposition (Plasmalab 100 from Oxford Plasma Technology, England). The IDTs have 100 finger pairs each with
L=9 mm, W =3.2 mm and D=5 mm. The finger electrodes were made of 200 nm thick sputtered aluminum with
SPIE USE, V. 1 5455-43 (p.4 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
split-finger geometry to minimize ripples in the passband of the device; each finger is 5 µm wide, and equally
spaced such that the finger pairs periodicity corresponds to an acoustic wavelength λ of 40 µm.
For the first step, SU-8 walls with a thickness about 120 µm were obtained by using the following recipe.
Photosensitive SU-8 type 2075 (Microchem Corp., MA) was spin coated with an acceleration of 100 RPM/s to
reach 500 RPM that was held for 10 seconds, and followed by an acceleration of 300 RPM/s to reach 1500 RPM
that was held during 30 seconds. After the spinning, the resist was baked for 3 hours at 95◦ C on a hotplate
after a slow temperature ramp up in order to avoid stresses cracks in the epoxy layer. The SU-8 patterns
were obtained by soft contact lithography on a Karl-Süss MA6 mask aligner (SUSS MicroTech Inc., VT) for 5
cycles of 10 seconds and a rest time of 60 seconds between each exposure cycle. The exposure was followed by
a baking of 15 minutes at 95◦ C on a hotplate after a slow temperature ramp up to finish the reticulation of
the SU-8. The unexposed epoxy was then dissolved in propylene glycol monomethyl ether acetate (PGMEA)
with ultrasonication for 10 minutes. Figure 4(a) shows the resulting SU-8 structure patterned around the IDT.
Following that step, the wafer was diced to release each sensors with the patterned SU-8 walls. The devices were
subsequently wire bonded to a printed circuit board for electrical characterization.
For the second step, quartz glasses of 5 mm by 5 mm and 500 µm thick were placed and glued above the
SU-8 structures. An epoxy glue (Epotek H54 from Epoxy Technology, MA, USA) was manually dispensed above
the walls and followed by the placement of the glass. The final bonding between the SU-8 walls and the glass
was obtained by a 15 minutes curing of the epoxy at 100◦ C. The flatness of the top surface of SU-8 helped for
the perfect sealing between the walls and the capping. The same epoxy glue was used to make a 1 mm wide and
1-2 mm high well around the rest of the device and above the contact pads. This procedure defined an open liquid
cell above the acoustic sensing surface with dimensions 4.7 mm (= D) by 3.5 mm (> W ). Figure 4(b) shows a
scanning electron microscope (SEM) image of the cross section of the packaged device. The fully packaged SAW
biosensor itself is pictured in Figure 4(c).
4. PERFORMANCE AND DISCUSSION
In this section, the performance of the SU-8 liquid cell is evaluated and discussed. For that evaluation, we analyze
the acoustic damping due to the liquid cell in a first place and we report a biochemical recognition experiment
performed on the packaged SAW biosensor in a second place.
The liquid cell prevents the presence of liquids on the transducers and has a small disturbance effect on the
propagation of the acoustic wave. The influence of the SU-8 walls on the acoustic performance of the device
is shown in Figure 5 where the amplitude of the transfer function was measured in different situations on an
identical device before the capping: with and without the SU-8 walls and with and without water loading. A
loss of 7.7 dB is measured at the center frequency f0 =123.2 MHz, caused by the SU-8 walls on the acoustic path
between the two transducers. Although designed with a 150 µm width, each wall covers an effective length of
160 µm caused by the aspect ratio of SU-8. Therefore, we estimate the attenuation of the shear acoustic wave by
a factor of ∼1 dB per acoustic wavelength (∼1 dB/λ). This figure helps to evaluate the influence of the packaging
on the acoustic characteristics of the sensor and serves as guideline for the miniaturization of the device. It can
also be seen on Figure 5 that the SU-8 provides an efficient shielding against the negative influence of the water
loading when it spreads over the entire sensor, including the transducers.
In terms of the critical dimensions to attribute to the walls and to the acoustic wavelength, one is interested
in smaller wavelengths to achieve SAW biosensors with higher operating frequencies. The attributed dimensions
of the liquid cell is determined by the figure of 1 dB/λ. Different factors are involved in the design of the SU-8
liquid cell for the miniaturization of the device. Because of the precise patterning allowed with the photosensitive
epoxy, minimum dimensions are mainly limited by the aspect ratio of the SU-8 and in our case by the manual
alignment step of the glass capping that requires a relatively wide top surface. The height of the SU-8 walls must
exceed one acoustic wavelength to avoid interactions between the electromagnetic field of the transducers and
the capping material. In the present case the SU-8 was three λ high; a carefully addressed investigation of the
dimensions of the SU-8 walls would help to determine the impact of the liquid cell geometry on the performance
of the SAW biosensor.
SPIE USE, V. 1 5455-43 (p.5 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
(a)
(b)
(c)
Figure 4. (a) Optical top view picture of the SU-8 walls patterned around one interdigital transducer. The sensing area
is located on the left and is separated from the aluminum electrodes of the transducer by a 150 µm wide on acoustic path,
120 µm high SU-8 wall. (b) A scanning electron microscope (SEM) image of the cross section of the SU-8 walls above
the transducers and the quartz capping. (c) The fully packaged SAW biosensor on the printed circuit board.
SPIE USE, V. 1 5455-43 (p.6 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
−20
In air, no SU−8 walls
In air, with SU−8 walls
Water on sensing area only
Water on sensing area and IDTs
−25
−30
Amplitude [dB]
−35
−40
−45
−50
−55
−60
−65
−70
120
121
122
123
124
125
126
127
128
129
130
Frequency [MHz]
Figure 5. Effect of the SU-8 walls of the liquid cell on the amplitude of the transfer function. The figure displays the
amplitude of the electrical signal as a function of the frequency in different situations indicated by the legend (the water
loading is in addition to the SU-8 walls).
To illustrate the effectiveness of the SU8 liquid cell, we report the monitoring of protein adsorption by the
fully packaged Love-mode SAW biosensor described in section 3. The sensing area was covered by a 50 nm gold
layer on a 10 nm titanium adhesion layer to interface the sensor with an hydrophobic self-assembled monolayer
(SAM). The device was cleaned for 15 minutes in a UV-O3 chamber, immediately followed by the coating of the
sensing area with a drop of a 1mM stearyl mercaptan (Sigma-Aldrich) solution in ethanol. The deposition of
the hydrophobic SAM took place during 3 hours after which the device was rinsed with ethanol and dried with
a stream of nitrogen. As protein, type I fibrinogen from human plasma (Sigma-Aldrich) was diluted in PBS
buffer (137 mM N aCl, 6.44 mM KH2 P O4 , 2.7 mM KCl and 8 mM N a2 HP O4 ) to a concentration of 46 µg/ml.
The adsorption of the protein on the sensing surface was monitored by the delay phase of the SAW biosensor at
the central frequency f0 =123.2 MHz on a HP4396A Network Analyzer (Hewlett-Packard Company, CA). The
fibrinogen adsorption was preceded and followed by a rinsing step with PBS buffer and water. Figure 6 reports
the monitoring of the sensor as a function of time.
The packaging is not preventing the correct operation of the device and is apparently not interfering with
the biochemical experiment. The adsorption of the protein on the sensing area causes a SAW velocity shift that
appears as a delay phase shift through the IDTs. The adsorbed protein mass per surface unit ∆m is related to
the phase difference ∆φ (in degrees) between starting and stopping points of the experience by3
∆m =
λ × ∆φ
360 × D × S
(1)
where D=4.7 mm is the effective sensing length, λ=40 µm and S the mass sensitivity of this specific Love-mode
SAW biosensor. The mass sensitivity has been calibrated to a value of 280 cm2 /g via copper electrodeposition.13
SPIE USE, V. 1 5455-43 (p.7 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
142
141
140
PBS
o
SAW Delay Phase [ ]
139
138
137
136
H2O
46 µg/ml
fibrinogen in PBS
135
134
133
132
PBS
131
1000
2000
3000
4000
time [s]
5000
PBS
6000
7000
Figure 6. Biorecognition fibrinogen experiment that shows no interference because of the presence of the liquid cell. The
delay phase of a Love-mode SAW biosensor is monitored as a function of time. The fibrinogen adsorption on the sensing
surface modifies the SAW velocity. This change appears as a delay phase shift because of the electrical transduction.
In the present case, the observed phase shift of 8◦ in Figure 6 corresponds to a surface density of 675 ng/cm2
that is a likely value for fibrinogen surface coverage. Based on this observation, we conclude to the absence of
interaction between the liquid cell and the sensing aspect.
A detailed comparison of our approach with other types of liquid cell is difficult to give because of a lack of
data and studies with regards to its influence on the general performance of the SAW biosensor and partly because
of the innovative aspect of the SU-8 liquid cell. One of the most difficult aspect related to a pre-molded liquid
cell is the necessity to manually adjust its position above the sensing surface and to mechanically press the cell
against the SAW device till a perfect sealing is reached so that no leaks can occur during the whole biochemical
experiment. The position of the cell and the pressure are inevitably varying from experiments to experiments
and may result in a lack of experimental reproducibility. In comparison, we claim that the precise patterning
of the SU-8 associated to the rigidity and the stability of our whole packaging solution is an improvement that
assures the experimental reproducibility of SAW biosensors.
5. CONCLUSION AND FUTURE DEVELOPMENTS
The packaging of surface acoustic wave biosensors is solved by the realization of a SU-8 based liquid cell. We
have shown that the SU-8 assures the sealing of a capping layer above the transducers and prevents liquids to
contact the electrodes of the transducers. The design of the liquid cell results in an attenuation factor about 1 dB
per acoustic wavelength of SU-8 on the acoustic sensing path. The materials selected for the fabrication of the
liquid cell have a good chemical resistance and are apparently not interfering with the biochemical sensing event.
The presented SU-8 liquid cell is a basic step in the integration of the packaging of SAW biosensors at the wafer
processing level but the influence of design parameters on the instrumentation and the sensing characteristics of
the SAW biosensor is part of future developments.
SPIE USE, V. 1 5455-43 (p.8 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
For future developments, improvements are required to increase the integration of IC fabrication techniques in
the realization of the packaging solution. A first improvement to the presented solution resides in a microfluidic
system for liquid flow above the sensing surface, an aspect that is missing in the developed liquid cell and although
a static fluid display does not prevent the correct operation of the biosensor. An hybrid approach is based on
the advantage provided by the SU-8 walls in term of integration, low acoustic attenuation and reproducibility,
and by the flexibility of a pre-molded external cell. The SU-8 plays the role of a microfabricated O-ring precisely
defining the sensing area; above this O-ring, the glass capping could be advantageously replaced by a removable
PDMS molded microfluidic. This solution has the disadvantage of requiring a mechanical set-up to apply a
pressure for the sealing of the SU-8/PDMS interface.
A second improvement at the wafer level is the wafer bonding of the capping. The identification of two steps
in the fabrication methodology let appear a clear break between batch and manual processes in the packaging
of the SAW biosensor. The wafer bonding of a full glass wafer requires a glass with holed structures above
the sensing area, for instance with sand blasted openings. The advantage is that SU-8 can be used as bonding
material. This solution is an opportunity to have a full batch processing of the packaging, unless other packaging
methods are developed.
APPENDIX A. ELECTRICAL MODEL OF THE SAW DEVICE
Surface acoustic wave devices and interdigital transducers have been described abundantly in the literature.14, 15
We use the electrical model to evaluate the effect a liquid loading on the electrical instrumentation of the device.
The ideal transducer is modeled by a radiation conductance Ga (f ) that is function of the frequency f connected
in parallel with a static capacitance CT . The electrical admittance of the transducer is written as
YIDT (f ) = Ga (f ) + j2πf CT
(2)
and an expression for CT is N W C0 , where N is the number of finger pairs with an acoustic aperture W
corresponding to the overlap length of the fingers and C0 the capacitance per finger pair and per unit length.
The synchronous frequency f0 is the ratio between the acoustic velocity V and the finger pairs periodicity λ
(i.e. f0 = V /λ). For frequencies near the synchronous frequency, the radiation conductance is approximated by
2
2
Ga (f ) ≃ 8K f0 N W C0
sin x
x
2
,
(3)
where x = N π(f − f0 )/f0 and K 2 is the electromechanical coupling constant of the piezoelectric substrate. The
transfer function H(f ) of a lossless SAW delay line relates the output (or load) voltage VL to the source voltage
VS and is given by
YDL R
VL
=
(4)
H(f ) =
VS
(1 + YIDT R)2 − (YDL R)2
where R is the source and load impedances supposed identical and equal to 50 Ohms, and YDL is a transfer
admittance given by
(5)
YDL = Ga (f ) exp {j (π − 2x − 2πf L/V )} .
The electrical model of the SAW device allows us to evaluate the effect of water loading on the transducers.
In a rough estimation, the capacitance per finger pair and per unit length is given by C0 = (εrS + εrL )× ε0 , where
ε0 is the vacuum permittivity (8.854 pF/m), εr is the relative permittivity and the subscripts S and L refer to
the substrate and the loading, respectively. We compare the effect of water loading (εr = 80) of the transducers
for two classical substrates with low and high K 2 : quartz and lithium niobate (LNO). The maximum amplitude
of the transfer function, that is 20 log H(f ), is simulated as a function of the value of C0 . The parameters of
the simulation are given in Table 1 and with W = 3.2 mm, L = 9 mm and λ = 40 µm.
Figure 7 displays the net change of the amplitude with air and with water above the electrodes of the
transducers. The attenuation is more pronounced on the quartz with a loss about 12 dB, while for LNO the loss
is lower with a value of 6.9 dB. This difference is mainly due to the high value of both the dielectric permittivity
and the electromechanical coupling constant of LNO.
SPIE USE, V. 1 5455-43 (p.9 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
Quartz
LNO
K2
0.0011
0.046
f0
126.5 MHz
87.2 MHz
V
5060 m/s
3488 m/s
N
100
40
Table 1. Simulation parameters
0
−5
LNO/AIR
Maximum amplitude [dB]
−10
LNO/WATER
−15
−6.9 dB
−20
QUARTZ/AIR
−25
−30
−12 dB
−35
−40
0
200
QUARTZ/WATER
400
600
800
Capacitance C0 [pF/m]
1000
1200
Figure 7. Theoretical evolution of the maximum electrical amplitude of the SAW device as a function of the capacitance
per finger pair and per unit length (C0 ) for lithium niobate (LNO, upper curve) and quartz (lower curve). The dots are
the values corresponding to air and water above the finger electrodes.
SPIE USE, V. 1 5455-43 (p.10 of 11) / Color: No / Format: A4/ AF: A4 / Date: 2004-03-26 10:23:36
Please verify that (1) all pages are present, (2) all figures are acceptable, (3) all fonts and special characters are correct, and (4) all text and figures fit within the
margin lines shown on this review document. Return to your MySPIE ToDo list and approve or disapprove this submission.
ACKNOWLEDGMENTS
L. A. Francis thanks the FRIA (Fonds pour la Formation à la Recherche dans l’Industrie et dans l’Agriculture)
for financial support.
REFERENCES
1. D. S. Ballantine, R. M. White, S. J. Martin, M. J. Ricco, E. T. Zellers, G. C. Frye, and H. Wohltjen, Acoustic
wave sensors. Theory, design, and physico-chemical applications, Academic Press, San Diego, 1997.
2. A. Leidl, I. Oberlack, U. Schaber, B. Mader, and S. Drost, “Surface acoustic wave device and applications
in liquid sensing,” Smart Mater. Struct. 6, pp. 680–688, 1997.
3. B. Jakoby and M. J. Vellekoop, “Properties of Love waves: applications in sensors,” Smart Matr. Struct. 6,
pp. 668–679, 1997.
4. R. L. Baer, C. A. Flory, M. Tom-Moy, and D. S. Solomon, “STW chemical sensors,” in IEEE Trans.
Ultrasonics Symp., pp. 293–298, 1992.
5. G. L. Harding, J. Du, P. R. Pencher, D. Barnett, and E. Howe, “Love wave acoustic immunosensor operating
in liquid,” Sensors and Actuators A 61, pp. 279–286, 1997.
6. R. M. White and F. W. Voltmer, “Direct piezoelectric coupling to surface elastic waves,” App. Phys. Lett.
7, pp. 314–316, 1965.
7. B. Jakoby and M. J. Vellekoop, “Analysis and optimization of Love wave liquid sensors,” in IEEE Trans.
on UFFC, pp. 1293–1302, 1998.
8. E. Gizeli, A. C. Stevenson, N. J. Goddard, and C. R. Lowe, “A novel love-plate acoustic sensor utilizing
polymer overlayers,” in IEEE Trans. on UFFC, pp. 657–659, 1992.
9. O. Tamarin, C. Déjous, D. Rebière, J. Pistré, S. Comeau, D. Moynet, and J. Bezian, “Study of acoustic
Love wave devices for real time bacteriophage detection,” Sensors and Actuators B 91, pp. 275–284, 2003.
10. E. Gizeli, F.Bender, A. Rasmusson, K. Saha, F. Josse, and R. Cernosek, “Sensitivity of the acoustic waveguide biosensor to protein binding as a function of the waveguide properties,” Biosensors and Bioelectronics
18, pp. 1399–1406, 2003.
11. D. W. Branch and S. M. Brozik, “Low-level detection of a bacillus anthracis simulant using Love-wave
biosensors on 36◦ YX LiTaO3 ,” Biosensors and Bioelectronics 19, pp. 849–859, 2004.
12. R. J. Jackman, T. M. Floyd, R. Ghodssi, M. A. Schmidt, and K. F. Jensen, “Microfluidic systems with
on-line UV detection fabricated in photodefinable epoxy,” J. Micromech. and Microeng. 11, pp. 263–269,
2001.
13. J.-M. Friedt, L. Francis, K.-H. Choi, F. Frederix, and A. Campitelli, “Combined atomic force microscope
and acoustic wave devices: Application to electrodeposition,” J. Vac. Sci. Technol. A 21, pp. 1500–1505,
2003.