Academia.edu no longer supports Internet Explorer.
To browse Academia.edu and the wider internet faster and more securely, please take a few seconds to upgrade your browser.
2000, IEEE Engineering in Medicine and Biology Magazine
…
7 pages
1 file
Describes an automatic optical system that is attached to a slit lamp in order to provide automatic keratometry at low cost. The system consists of projecting a light ring as a target onto the patient's cornea as well as analysis of the deformation of the target in order to obtain the radius of curvature and the axis of the associated astigmatism. The reflected image of the target is displayed on a PC monitor, while dedicated software performs the analysis of the image, which provides the corneal keratometry
British Journal of Ophthalmology, 1998
Aims-To evaluate intra-and interobserver variability in measurements on normal and astigmatic corneas with keratometry and computerised videokeratography. Methods-Keratometric readings with the 10 SL/O Zeiss keratometer and topographic maps with the TMS-1 were obtained by two independent examiners on 32 normal and 33 postkeratoplasty corneas. Inter-and intraobserver coeYcients of variability (COR) for measurements of steep and flat meridian power and location, in addition to the magnitude of astigmatism, were assessed. Results-Compared with TMS-1, the 10 SL/Okeratometershowedasuperiorrepeatability in measuring normal corneas (intraobserver COR for keratometry and TMS-1 respectively: 0.22 and 0.30 D for steep meridian power; 0.18 and 0.44 D for flat meridian power; 0.26 and 0.40 D for astigmatism; 5°and 26°for steep meridian location; 5°and 13°for flat meridian location). Astigmatism intraobserver COR (0.20 D and 0.26 D for the two observers) and interobserver COR (0.28 D) of the keratometer for normal corneas was very good and not aVected by observers' experience. Repeatability of the TMS-1 on normal corneas was found to be: (a) observer related, and (b) astigmatism related. A novice observer showed a much greater COR (1.62 D for astigmatism, 30°f or flat meridian location) compared with the experienced examiner (0.40 D for astigmatism, 13°for flat meridian location). Higher deviation scores were observed for corneas with higher astigmatism. For the postkeratoplasty corneas, again the keratometer achieved superior reproducibility (astigmatism interobserver COR 1.12 D for keratometry, 4.06 D for TMS-1; steep meridian location interobserver COR 10°f or keratometry, 34°for TMS-1). Conclusion-Keratometric readings are more reproducible than topographic data both for normal and postkeratoplasty corneas. The two instruments should not be used interchangeably especially on highly astigmatic corneas. For the TMS-1, users with the same level of experience should be employed in clinical or experimental studies.
Journal of Cataract & Refractive Surgery, 2012
To compare the corneal astigmatism (magnitude and axis location) derived by total corneal power (TCP), automated keratometry, and simulated keratometry. Siriraj Hospital, Mahidol University, Bangkok, Thailand. Prospective comparative study. Eyes with previous ocular surgery or abnormalities were excluded. All patients were examined with the ARK 730A autokeratometer and the Galilei analyzer. The steepest and flattest corneal power along with the steepest axis of the TCP, automated keratometry, and simulated keratometry were recorded. Vector analysis (J0 and J45) was calculated. Analysis of variance with Bonferroni correction was performed for multiple comparisons. Outcome measures were the magnitude and axis location of astigmatism. One hundred eyes of 100 cataract patients were randomly selected. There was no statistically significant difference in the mean steepest axis between TCP (93.31 ± 68.75 [SD]), automated keratometry (94.24 ± 64.78), and simulated keratometry (92.42 ± 64.30). However, the mean magnitude of astigmatism measured by TCP (1.23 ± 0.75) was significantly higher than that measured by automated keratometry (0.93 ± 0.68) (P=.01) but not than that measured by simulated keratometry (1.08 ± 0.68) (P=.43); there was no statistically significant difference in J0 or J45. Twenty two (40%) of 54 eyes with more than 1.00 diopter of TCP astigmatism had more than 10 degrees of axis difference from automated keratometry. The magnitude of TCP astigmatism was higher than that of automated keratometry. The axis location was similar. However, there was more than 10 degrees of axis difference between automated keratometry and TCP in patients with high astigmatism. No author has a financial or proprietary interest in any material or method mentioned.
Acta Ophthalmologica, 2009
Keratoscope photograph negatives of one normal cornea and calibration steel spheres were digitized by a CCD video camera connected to a frame grabber. The centroid of the innermost reflected mire was determined and used for conversion of the image to polar coordinates. The distance to each reflected ring was computed. Calculations of the radius of curvature describing the reflecting surface locally were performed by an iterative procedure based on differential equations. The calculated mean radius for steel spheres in the range of 5.5 to 11 mm in diameter conformed with the known values. The average difference between observed and expected values was 0.2% and the average coefficient of variation was 0.1%. The standard deviations on observed ring means decreased inversely with ring number with an average coefficient of variation of 0.2%. Likewise, on a corneal surface SD also decreased inversely with ring number from 0.080 to 0.032 mm. In terms of refractive power this meant, that outwards from the third ring 95% of observations were within f 0.4 diopters from the average value.
Journal of Cataract & Refractive Surgery, 2000
To evaluate the effectiveness of indices derived from the EyeSys System 2000 in detecting keratoconic corneas.
BioMedical Engineering OnLine, 2009
Background: High astigmatisms are usually induced during corneal suturing subsequent to tissue transplantation or any other surgery which involves corneal suturing. One of the reasons is that the procedure is intimately dependent on the surgeon's skill for suturing identical stitches. In order to evaluate the influence of the irregularity on suturing for the residual astigmatism, a prototype for ophthalmic surgical support has been developed. The final intention of this prototype is to be an evaluation tool for guided suture and as an outcome diminish the postoperative astigmatism. Methods: The system consists of hand held ring with 36 infrared LEDs, that is to be projected onto the lachrymal film of the cornea. The image is reflected back through the optics of the ocular microscope and its distortion from the original circular shape is evaluated by developed software. It provides keratometric and circularity measurements during surgery in order to guide the surgeon for uniformity in suturing. Results: The system is able to provide up to 23D of astigmatism (32D-55D range) and is ± 0.25D accurate. It has been tested in 14 volunteer patients intraoperative and has been compared to a commercial keratometer Nidek Oculus Hand-held corneal topographer. The correlation factors are 0.92 for the astigmatism and 0.97 for the associated axis.
Journal of Cataract & Refractive Surgery, 1992
The EyeSys Corneal Analysis System Model I measures corneal topography using digital image analysis of placido rings reflected off the cornea. With three observers, we compared the accuracy and reproducibility (precision) of the Marco Keratometer Model 1 and EyeSys Corneal Analysis System Model I using four poly(methyl methacrylate) spheres (37.50, 42.51, 47.54, and 55.06 diopters [D]), three steel spheres (40.50, 42.50, and 44.75 D), and 20 normal human eyes (41.50 to 46.00 D). For the spheres, the standard deviations of intra-observer and overall reproducibility for both devices were less than 0.12 D; the absolute mean differences between the measurements of the seven spheres and the known values were 0.25 D or more for two spheres as measured by the keratometer and none as measured by the EyeSys. For the normal corneas, the standard deviations of intra-observer and overall reproducibility for dioptric measurements were 0.07 D and 0.14 D for the keratometer and 0.13 D and 0.19 D for the EyeSys. The EyeSys Corneal Analysis System Model I exceeds keratometer accuracy in reading calibrated spheres and approaches keratometer reproducibility in measuring the 3-mm zone of normal human corneas.
PubMed, 1994
Purpose: Although visual inspection of corneal topography maps by trained experts can be powerful, this method is inherently subjective. Quantitative classification methods that can detect and classify abnormal topographic patterns would be useful. An automated system was developed to differentiate keratoconus patterns from other conditions using computer-assisted videokeratoscopy. Methods: This system combined a classification tree with a linear discriminant function derived from discriminant analysis of eight indices obtained from TMS-1 videokeratoscope data. One hundred corneas with a variety of diagnoses (keratoconus, normal, keratoplasty, epikeratophakia, excimer laser photorefractive keratectomy, radical keratotomy, contact lens-induced warpage, and others) were used for training, and a validation set of 100 additional corneas was used to evaluate the results. Results: In the training set, all 22 cases of clinically diagnosed keratoconus were detected with three-false-positive cases (sensitivity 100%, specificity 96%, and accuracy 97%). With the validation set, 25 out of 28 keratoconus cases were detected with one false-positive case, which was a transplanted cornea (sensitivity 89%, specificity 99%, and accuracy 96%). Conclusions: This system can be used as a screening procedure to distinguish clinical keratoconus from other corneal topographies. This quantitative classification method may also aid in refining the clinical interpretation of topographic maps.
Journal of Cataract & Refractive Surgery, 1996
Purpose: To compare the morphologic appearance and measurements of in situ keratomileusis performed with the UniversaiKeratome™ (UK) with those done with the Automated Corneal Shaper (ACS). Setting: Surgical suite within private practice. Methods: Procedures were performed the same day on mate eye-bank eyes. In situ keratomileusis was done using existing nomograms for each instrument to resect a cap thickness of 160 11-m and a myopic resection of 1 00 11-m. Intraocular pressures were increased by inflating the globes with balanced salt solution and were measured with the suction fixation rings in place. The excised caps and stromal resections were measured twice independently after surgery, again after tissue fixation, and then evaluated with light and scanning electron microscopy. Results: No complications were encountered. Compared with the ACS, the UK was easy to set up, use, clean, and take down. Its excised tissue dimensions were greater and more predictable, it resected a concave shaped lenticule (edges imperceptibly blending with the host stroma), and it created a smoother power resection surface and primary resection base. Conclusions: Smoother, predictable tissue resection, and simple assembly/disassembly and use give the UK an apparent advantage over the ACS. The UK corrects astigmatism and hyperopia by changing the shape of the poly(methyl methacrylate) optical insert.
Journal of Cataract & Refractive Surgery, 1998
A ccurate optical modeling of the human cornea is crucial for diagnostic purposes and refractive surgery planning. Sophisticated videokeratographic instruments have been developed that can be used to build bidimensional and tridimensional maps of anterior
Scientific reports, 2018
To investigate a method for precision analysis to discriminate true corneal change from measurement imprecision in keratoconus (KC). Thirty patients with KC and 30 healthy controls were included. Coefficients of repeatability and limits of agreement (LOA) were compared using multiple measurements for inter-observer and inter-device agreement with the Pentacam HR, Orbscan IIz, and Tomey Casia SS-1000. Correlation of repeated measurements was evaluated using a linear mixed effect model (also called random effect model). A formula was derived for the theoretical expected change in precision and compared with measured change. Correlation between measurements from the same eye was small (R = 0.13). The 99.73% LOA (3 SD) of the mean of three measurements, provided better precision than 95% LOA (2 SD) of single cut-off values as expected from statistical theory for uncorrelated measurements for evidence of a significant change in corneal shape in patients with keratoconus. This enabled the...
Revista de Políticas Públicas
The Malta Independent, 2022
Anuario 2009, Centro de Estudios Superiores de México y Centroamérica, 2010
European Proceedings of Social and Behavioural Sciences EpSBS , 2021
The Indonesian Journal of Electrical Engineering and Computer Science (IJEECS), 2023
Environmental Research, 1997
Dialectica, 2015
Nature Chemistry, 2022
Cancer Cell, 2007
Nuclear Technology and Radiation Protection, 2014