J-PET: A Novel TOF-PET Detector based on Plastic Scintillators

J-PET: A Novel TOF-PET Detector based on Plastic Scintillators

2016
Abstract
The purpose of the reported research is the elaboration of the method for construction of the cost-effective whole-body single-bed positron emission tomography scanner enabling simultaneous PET/CT and PET/MR imaging. The Jagiellonian Positron Emission Tomograph (J-PET) is built out of 192 scintillator strips arranged axially in three layers forming a cylindrical diagnostic chamber with the diameter of 85 cm and axial field-of-view of 50 cm. The novelty of the concept lies in employing long strips of plastic scintillators instead of crystals as detectors of the annihilation quanta, and in using the timing of signals instead of their amplitudes for the reconstruction of Lines-of-Response. To take advantage of the superior timing properties of plastic scintillators a novel multi-voltage-threshold front-end electronics was developed allowing for sampling of signals in a voltage domain. An axial arrangement of long strips of plastic scintillators, and their small light attenuation allows us to make a TOF-PET scanner with a long axial field-of-view. The presented solution opens unique possibilities of combining PET with CT and PET with MRI for scanning the same part of a patient at the same time with both methods. The relative ease of the cost effective increase of the axial field-of-view makes the J-PET tomograph competitive with respect to the current commercial PET scanners as regards sensitivity and time resolution.