ARTICLE IN PRESS
Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
www.elsevier.com/locate/nima
Absolute activity measurement of radon gas at IRA-METAS
Philippe Springa,, Youcef Nedjadia, Claude Bailata, Gilles Trisconeb, Franc- ois Bochuda
a
Institut Universiataire de Radiophysique Appliquée, Grand Pré 1, 1007 Lausanne, Switzerland
b
École d’ingénieurs de Genève, rue de la Prairie 4, 1202 Geneva, Switzerland
Received 12 June 2006; received in revised form 5 July 2006; accepted 25 July 2006
Available online 23 August 2006
Abstract
This paper describes the system of the Swiss national metrological institute (IRA-METAS) for the absolute standardisation of radon
gas. This method relies on condensing radon under vacuum conditions within a specified cold area using a cryogenerator, and detecting
its alpha particles with an ion-implanted silicon detector, through a very accurately defined solid angle. The accuracy of this defined solid
angle standardisation technique was corroborated by another primary measurement method involving 4pg NaI(Tl) integral counting and
Monte Carlo efficiency calculations. The 222Rn standard submitted by IRA-METAS to the Système International de Référence (SIR) at
the Bureau International des Poids et Mesures (BIPM) has also been found to be consistent with an analogous standard submitted by the
German national metrological institute (PTB). IRA-METAS is able to deliver radon standards, with activities ranging from a few kBq to
350 kBq, in NIST-Type ampoules, and glass or steel containers usable for calibrating radon-measuring instruments.
r 2006 Elsevier B.V. All rights reserved.
PACS: 29.40.Wk; 29.30.Ep
Keywords: Radon; Solid-angle particle counting; Primary standard; Traceability; 4pg counting
1. Introduction
Radon is known to be the major component of natural
irradiation of the population. Over 20 occupational
epidemiological studies of radon-exposed underground
miners unequivocally demonstrated that prolonged exposure to radon increases the risk of lung cancer [1]. Pooled
studies of 65 000 miners found a linear relationship
between radon exposure and lung cancer deaths [2,3]. This
relationship was maintained even among a subgroup of
miners that had lower exposures extending into the range
for some homeowners [4]. These findings therefore suggest
that residential radon also carries a risk of cancer.
Improving radon metrology is then necessary for the
traceability of secondary measurements of radon in air
concentration. In spite of its important contribution to
irradiation, only few intercomparisons of either radon
standards or radon in air concentration have been carried
Corresponding author.
E-mail address:
[email protected] (P. Spring).
0168-9002/$ - see front matter r 2006 Elsevier B.V. All rights reserved.
doi:10.1016/j.nima.2006.07.055
out, and furthermore these have revealed a spread in the
results of about 5% [5] to 40% [6].
Traceability chains for radon measurement reference
systems in the national metrology institutes (NMI) are
realised in two ways. In some cases, laboratories measure
the gamma emitting daughters of 222Rn gas traceable to
226
Ra Standard Reference Materials (SRMs) [6–9]. Alternatively, measurements performed in some laboratories are
traceable to radon primary standards in which radon itself
is measured with an absolute method. To our knowledge,
only two absolute measurement methods have been
implemented. The first one, introduced by Piccolo [10], is
based on the detection of alpha particles under an
accurately measured solid angle. This system has also been
implemented at the PTB [11] and at IRA-METAS. A
second absolute method described in Ref. [12] involves a
proportional counter of calculable 222Rn efficiency and
quantifiable active volume.
This paper describes the system developed at IRAMETAS, the procedures to validate it and the first primary
standardisation of 222Rn submitted [13,14] to the Système
ARTICLE IN PRESS
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
International de Référence (SIR) for activity comparison at
the Bureau Internationnal des Poids et Mesures (BIPM).
753
2.2. Measurement chamber
2.1. Radium source
2.2.1. Alpha-detector
The measurement chamber consists of a closed cylindrical volume. An ion-implanted silicon detector (ORTEC
ULTRA)—with 450 mm2 active area, 100 mm minimum
depletion depth and 17 keV guaranteed maximum resolution for alpha particles—is located at the top of this
volume. There is a copper diaphragm right underneath this
alpha detector. A 75 mm stainless 316L steel foil closes the
bottom part of the chamber. In its centre, a laser-welded
nickel rod (5.8 mm of diameter) is linked to the cold head
of a cryogenic system (EBARA 531-120). The external part
of the stainless-steel foil is kept at room temperature. This
volume can be evacuated down to 104 hPa with a dryrunning-type pumping system Picodry (membrane and
turbo-molecular system manufactured by Edwards).
In addition to the detector, the spectrometry acquisition
chain includes an Ortec 478 high-voltage supply operated
at 50 V, an Ortec 142C preamplifier, an Ortec 472 amplifier
and an IBM PC with an Ortec 8K MCA Trump card. The
spectrum is acquired using Interwinner 5 (Ortec).
Radon is generated by a 500 kBq commercial source of
Ra (type RN-1025-500 Pylon Electronics Inc.). Radium
is in a dry powder form and the emanation rate is assumed
to be close to 100% and unaffected by normal variations of
pressure and temperature.
A special loop allows the rinsing of the source with
gaseous helium to remove possible water condensation
before growing the radon. The source is flushed then
evacuated (to 102 hPa) before each operation of the
system.
Radon growth within the sealed 226Ra source reaches
equilibrium after about 65 days. In our routine procedures
radon is grown for about 10 days yielding 420 kBq of 222Rn
(84% of the nominal activity), which allows measurement
with a sufficient statistic for our experimental conditions
and leaves enough time for quality controls before sending
the standard to the end user.
2.2.2. Radon condensation and geometry
The top of the nickel rod, shown in Fig. 2, is used as the
condensing surface for the 222Rn. Its base is rigidly linked
to the cold head of the cryogenic system. Temperature is
regulated with a Lakeshore 331 temperature controller
supplemented with a heating resistor circuit. Operating
temperature can be adjusted from about 15–400 K.
Measurements are made at low temperature, typically
40 K, in order to have a negligible saturation vapour
pressure [10]. Because of the strong temperature gradient
between the nickel rod and the external part of the
stainless-steel foil, the surface where the radon condenses
is very well defined. The diameter of the condensed radon
source can be measured by autoradiography with a pinhole
camera system, as described by Dersch [11]. A device with a
pinhole collimator was mounted at the place of the
detector. Its dimensions were calculated to give an image
2. IRA-METAS radon measurement system
A 226Ra source is connected through vacuum circuitry to
the measurement system. Gaseous radon diffuses into the
evacuated measuring chamber where it condensates gradually onto a pre-cooled defined surface (cold-finger). The
activity is measured by counting the alpha particles emitted
by 222Rn. The detection efficiency only depends on a
geometrical efficiency factor G ¼ O/4p which is the fraction
of the total solid angle O under which alpha particles can
be detected. After measurement, radon can be transferred
into a liquid nitrogen-cooled container by heating the cold
finger. The radon source thus obtained can be used as a
calibration standard. Fig. 1 shows a general scheme of the
method.
226
alpha
detector
radium
Source
collimator
solid radon
gaseous radon
(1)
(3)
cold finger
(2)
liquid
nitrogen
(4)
measuring
assembly
solid
radon
Fig. 1. General scheme of the measuring method: (1) radon gas is introduced into the system, (2) radon is frozen on the cold finger and measured by alpha
spectrometry, (3) radon is heated and then frozen into an ampoule bathing in liquid nitrogen and (4) the ampoule is sealed while radon is still solid.
ARTICLE IN PRESS
754
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
Alpha detector
Diaphragm
Diaphragm axis
a
z
e
Fig. 2. The nickel rod soldered using laser technique to the 75 mm
stainless 316L steel foil.
b
Cold finger
Fig. 4. The distances involved in the calculation of the geometric
efficiency G are: the diaphragm radius a ¼ 5.997(2) mm, the source radius
b ¼ 3.2(2) mm, the off-axis eo0.3 mm, and the source–diaphragm distance
z ¼ 100.18(9) mm.
Fig. 3. Autoradiography of around 200 kBq of condensed radon.
Exposure time is 14 h. The measured source radius is 3.2(2) mm.
enlargement factor of 1. Fig. 3 shows the autoradiography
obtained using a Kodak MIN RH film.
Fig. 5. Top flange incorporating the calibrated micrometre.
2.2.3. Source–detector geometry
At the operating pressure, which is below 101 hPa in all
cases, all the alpha particles reaching the detector are
assumed to be counted. The intrinsic detection efficiency is
1 for the 5.5 MeV alpha particles emitted by radon since
they loose only about 0.7 keV along the 10 cm distance
between the source and the detector—their stopping power
is 71.3 eV/cm in air at this pressure.
The efficiency then depends solely on the solid angle O.
As indicated in Fig. 4, the dimensions involved in the
calculation of O are the diaphragm radius (a), the source
radius (b), the distance between the source and diaphragm
(z) and the off-axis of the source and detector distance (e).
Several algorithms are available for calculating the solid
angle [15,16]. For our configuration, these differ by less
than 0.05%.
The first critical parameter from the uncertainty point of
view is the diaphragm radius a. It was therefore measured
by an accredited laboratory (Brown & Sharpe technology,
Crissier, Switzerland) traceable to international standards.
Its value is 5.997(2) mm. Because the diaphragm stays at
room temperature during radon measurements, contraction effects can be disregarded.
The source–diaphragm distance z is the next critical
parameter. This length was measured with a purpose-built
special device (Fig. 5) incorporating a calibrated micrometre installed at the place of the flange containing the
alpha detector. A z value of 100.18(9) mm was found.
ARTICLE IN PRESS
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
755
3. Primary measurements
3.1. Measurement conditions
Fig. 6. Absolute variation (DZ) of the source–diaphragm distance (z) as a
function of temperature. Full symbols are measurements from 2005 and
open symbols are measurements from 2002
Measurements performed at different temperature and
pressure conditions resulted in an increase of 0.3 mm
between normal conditions (room temperature and atmospheric pressure) and typical measuring conditions (40 K,
103 mbar). Measurements of the source–diaphragm distance z at operating conditions are repeatable within 0.1%
(standard deviation). The absolute variation of z between
300 and 40 K has been very stable over the years as shown
in Fig. 6.
The eccentricity e was well controlled during the set-up
of the system. The nickel rod was placed in front of the
diaphragm with a centring device. This ensures that e is
lower than 0.3 mm.
The geometry factor G ¼ O/4p of the system, assuming a
uniform distribution of the activity over the cold finger,
thus amounts to 8.927(18) 104.
2.3. Source–chamber radon transfer
Radon is transferred from the 226Ra source to the
measurement vacuum chamber. During the growth of
222
Rn, pressure in the radium source raises due to its
imperfect sealing allowing constituents of atmospheric air
to enter the source volume. During the transfer, these
constituents condense with radon on the cold surface and
degrade the alpha spectra. In order to remove these
residual elements, especially water vapour, which is the
most critical, a commercial gas clean filter (Varian
Chrompack) was integrated into the vacuum circuit
between the radium source and the measuring chamber.
Typical measurements are realised with about
250–350 kBq of condensed 222Rn at the cold point at
40 K. Piccolo showed that at 50 K, uncondensed radon
represents only a few part per million of the total activity
[10]. To ensure that all alpha particle emitted in the
measured solid angle are detected, measurements are made
at sufficiently low pressures, typically between 103 and
102 hPa. Our measurement system maintains this vacuum
level for about 6 h without pumping, which is more than
enough to carry out the measurements.
A typical radon measurement sequence consists in
background measurements, then a transfer of 222Rn. Six
or seven successive 10-min alpha spectra are acquired so as
to reach a counting precision of 0.1% or less. Radon is
then transferred into an appropriate container, maintained
at low temperature in liquid nitrogen, by warming the
nickel rod (the cold finger) up to 340 K. Once the transfer is
complete, the measurement chamber is sealed again, the
cold finger is cooled once more at 40 K, and the activity of
possible radon residues is measured after about 12 h.
3.2. Measurements results
The activity of radon is measured by detecting the 5.49
and 4.99 MeV alpha particles emitted by 222Rn. The sharp
resolution of the semiconductor detector discriminates easily
between these alpha emissions and those emitted by the
218
Po and 214Po offspring of radon, at 6.00 and 7.68 MeV,
respectively. A typical spectrum is represented on Fig. 7.
For each 10-min measurement, the counts under the two
radon peaks are integrated. The lower energy limit—
denoted ‘—is set to include the 4.99 MeV peak. The exact
placement of ‘ is not critical because the count rate in this
energy region is about that of the background. By contrast,
the position of the upper energy limit—referred to as h—of
the radon peak leads to a significant uncertainty over the
count rate because of its partial overlap with the 218Po
peak.
The method used to deal with this problem is as follows.
h is set midway between the 5.49 MeV peak of radon and
the 6.00 MeV peak of 218Po assuming both are symmetrical. One may compute the following set of integrated
count rates Rn between ‘ and 720 channels about h:
Z h20
R20 ¼
SðEÞ dE;
R19 ¼
R20 ¼
Z
Z
‘
h19
SðEÞ dE; . . . ;
‘
hþ20
SðEÞ dE,
‘
where S(E) refers to the energy spectrum. For any
given measurement, we take the arithmetic mean of the
ARTICLE IN PRESS
756
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
Fig. 7. Typical spectrum acquired in 10 min. Alpha peaks of daughters of 222Rn are easily discriminated from those of radon. The counts at low energy are
attributed to high-energy beta particles of 214Bi.
Table 1
Typical uncertainty budget for a primary measurement using the solid
defined angle method
Type A
Upper limit of integration for
222
Rn peak (%)
Geometry factor G (%)
Background (%)
222
Rn half-life (%)
Counting statistics (%)
Fig. 8. Seven successive 10-min measurements performed on April 5,
2006. Error bars includes uncertainty (k ¼ 1) for counting statistics and
for radon peak position (see Section 3.2). The solid line shows the average
value while the dotted one marks one standard deviation.
{R20, y, R20} count rate distribution to be the actual
count rate of radon.
The results of a typical measurement, performed on
April 5, 2006 at 10:00 UT, are shown on Fig. 8. The count
rates are corrected for background, and decay during
measurement. In the present example, activity was found to
be 352.60(99) kBq.
3.3. Uncertainty budget
Table 1 presents the uncertainty budget for the radon
measurement mentioned above. The main contribution to
the uncertainty comes from the geometrical efficiency
factor G discussed in Section 2.2.3. Calculation of the
uncertainty of this factor gives a relative standard
uncertainty of 0.2%.
The next main contribution is the statistical uncertainty,
which depends on the activity and the measurement
Total (%)
Combined uncertainty
(quadratic sum of type A and B) (%)
Type B
0.138
0.201
0.01
0.002
0.09
0.09
0.264
0.28
duration. For a typical 1-h measurement of about
250 kBq of 222Rn, the relative standard uncertainty is
about 0.1%. The propagation of the uncertainty on the
upper limit of the integration window is taken to be the
spread of the Rn count rate set assuming a rectangular
distribution. Collected data over 20 spectra ground this
assumption. The propagated uncertainty was found to be
in this case 0.14%. This, of course, presumes that no
uncertainty is entailed by the counting method used by the
multi-channel analyser, in our case an Ortec 8 K MCA
Trump card, a hypothesis we are currently probing and
shall report about in a subsequent publication.
Finally, the typical global standard uncertainty amounts
to 0.28%.
3.4. Comparison with the 4pg NaI(Tl) integral counting
method
Comparing the results of this defined solid angle
standardisation technique with those of another primary
ARTICLE IN PRESS
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
measurement method is useful to probe its accuracy. At
IRA-METAS, radon standards are therefore also measured using 4pg NaI(Tl) integral counting with Monte
Carlo efficiency calculations.
Once a radon condensate has been measured, it can be
transferred through a vacuum circuit into a NIST-type
ampoule whose bottom bathes in liquid nitrogen (see
Fig. 1) and then flame-sealed. After secular equilibrium
between radon and its daughters this ampoule is measured
in a well-type 500 500 NaI(Tl) monocrystal. The radonloaded NIST ampoule is held at the bottom of the 43.5 cm3
volume well with a closely fitting Plexiglas tube, and
integral countings above a 22.6 keV energy cut-off are
repeatedly made using an adequate electronic chain.
The detection efficiencies for radon and its descendents
are computed numerically using Monte Carlo simulations
with the GEANT code [17]. At equilibrium, from the point
of view of emissions able to reach the sensitive volume of
the NaI(Tl) detector, only 222Rn, 214Pb and 214Bi matter.
The rest of the daughters of radon do not contribute
significantly because either their branching ratios or the
intensities of their radiative emissions are negligibly small.
If, after the onset of secular equilibrium between radon
and its daughters, N counts are registered during a
measurement initiated at tini. and lasting Dt, the activity
of radon at tini can be shown to be
ARn ðtini Þ ¼
N
1
lRn Dt
,
Dt Rn þ aPb Pb þ aBi Bi 1 elRn Dt
where ei (i ¼ Rn, Pb, Bi) are the Monte Carlo computed
total detection efficiencies for 222Rn, 214Pb and 214Bi,
respectively, while aPb (aBi) is the ratio of the activity of
214
Pb (214Bi) over that of 222Rn. The fraction on the right,
where lRn stands for the radioactive constant of 222Rn,
accounts for the decay of radon during measurement.
As a typical example, the radon condensate discussed in
Section 3.2 was transferred into a NIST ampoule and
measured over a 3-week period in the 4pNaI(Tl) detector.
The measurements [18] displayed a robust internal
coherence and yielded an activity of 353.62(40) kBq, to be
contrasted to the defined solid angle value of
352.60(99) kBq at the same date of reference. The 4pg
NaI(Tl) measurement is therefore about 0.3% higher but
compatible with the defined solid angle primary measurement within the uncertainties, thus lending confirmation to
it. This conclusion is supported by three other similar
standardisation comparisons we made in recent years.
Note that the uncertainty of the 4pg NaI(Tl) result above
does not take into account the propagation of the
uncertainties of the dimensions of the geometries or those
of the decay schemes of 222Rn, 214Pb and 214Bi on the
detection efficiencies. It only includes the statistical
uncertainty associated with the counting, the Monte Carlo
statistical uncertainty, and the propagation of the uncertainties of the half-lives of radon and its daughters on
aPb and aBi at secular equilibrium. The latter were
estimated by varying stochastically 10 000 times the periods
757
of radon and all its daughters within their distributions—
assumed to be Gaussian—in the Bateman equations, and
then taking the standard deviations of the distributions of
aPb and aBi values as the relevant uncertainties.
4. Delivery of radon gas standards
4.1. Transfer to a container
Condensed radon measured with the absolute method
has to be transferred into a sealed container in order to be
used as an activity standard. Radon is transferred by
heating the cold finger onto which radon solidified while, at
the opposite end of the vacuum circuit, bathing the
container for delivering radon in liquid nitrogen (see
Fig. 1). The transfer can be monitored by measuring
intermittently the alpha spectra generated by the radon left
in the chamber.
Fig. 9 displays the evolution of the 5.49 MeV peak of
222
Rn during heating and transfer. During heating, the
sublimation of radon can be observed on the alpha spectra
since the radon peak shifts towards higher energy and
narrows as self-absorption in the evanescent solid radon
disk decreases. Counting under the 222Rn peak increases
due to the spread of radon over the whole volume of the
measurement chamber. Radon progressively condensates
into the cold container. After several minutes, the radon
peak begins to decrease and the pressure in the measurement chamber falls down to a value close to the pressure at
the beginning of the transfer.
Finally, the container is sealed and measurements
are performed to check whether the transfer was total
and quantify possible radon residues as described in
Section 3.1.
4.2. Available geometries
IRA-METAS is able to produce radon gas standards in
three different geometries. The NIST type 5 ml gas
ampoule is a flame sealed container. It is the SIR reference
geometry for the measurement of gaseous samples. Two
different two-valves-containers allowing the flushing of the
gas content are also available. The first one is a classical
spherical glass bulb of about 30 cm3. The second is a more
robust cylindrical stainless-steel container with valves of
better quality. Fig. 10 shows a picture of the different
geometries.
4.3. Secondary measurements
Sealed NIST-type ampoules were used both for the SIR
contributions and calibrating IRA-METAS secondary
standard ionisation chamber of the IG11 type. This reliable
instrument [19] has been under quality control surveillance
since 1983, through regular measurements of 137Cs and
226
Ra reference sources, and consistently demonstrated a
strong stability. For this geometry, each prepared radon
ARTICLE IN PRESS
758
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
Fig. 9. Evolution of the alpha spectra (radon peak) during radon transfer. The increase in counting is due to an increase in the solid angle caused by the
spread of the radon molecules throughout the chamber.
Fig. 10. IRA-METAS deliverable geometries for radon standards. From left to right: (a) NIST-type sealed ampoule, (b) glass ampoule with flushing
manifolds and (c) stainless-steel container.
standard is measured after a 24-h period of latency
necessary for the gamma-emitting daughters of radon to
reach equilibrium—in the ionisation chamber in order to
check whether the condensed radon got transferred entirely
into the ampoule, by an alternative to solid angle counting.
In the case of the primary measurement reported in Section
3.2, the radon was subsequently transferred into a NISTtype ampoule and then measured in the ionisation
chamber; a calibration factor (Ae) was inferred.
For geometrical reasons, both types of valve-ampoules
cannot be measured in the IG11 ionisation chamber.
Nevertheless, secondary measurements, with a precision
typically better than 1%, can be performed with a well-type
NaI(Tl) counter in order to ascertain the entirety of
the radon transfer from the measurement chamber into the
ampoule and rule out radon leakage after sealing. The
container is placed on top of a NaI(Tl) scintillator in a
reproducible position, and integral countings above a
121.78 keV threshold are performed in well-defined electronic conditions. The NaI(Tl) detector is periodically checked
with a 152Eu source and has shown a stable efficiency over
the past years. The detection efficiency is calculated for
each radon standard produced. The linearity of the
response for this system was tested by measuring a
stainless-steel radon standard over a 23-day period covering the range of typical activities requested by customers.
The observed standard deviation of the calibration factor
was found to be 0.2% at most.
5. Conclusions
Radon activity was measured at IRA-METAS with an
absolute standardisation technique based on the condensation of radon and the detection of its alpha particles
ARTICLE IN PRESS
P. Spring et al. / Nuclear Instruments and Methods in Physics Research A 568 (2006) 752–759
through an accurately specified solid angle in vacuum
conditions. This method yields a typical standard uncertainty of about 0.3%.
The accuracy of this way of making absolute activity
measurement of radon was corroborated by comparing the
results of the defined solid angle standardisation technique
with those of another primary method, 4pg NaI(Tl)
integral counting with Monte Carlo efficiency calculations.
The latter technique yields measurements that about 0.3%
higher but compatible with the former.
IRA-METAS was the first NMI to submit a 222Rn
sample of known activity to the SIR at the BIPM. Since
then, the Physikalisch Technische Bundesanstalt (PTB) also
submitted a radon sample to the SIR. The degree of
equivalence computed by the BIPM indicates that our
activity determination and that of the PTB are consistent
with each other [20].
IRA-METAS can now produce radon standards, with
activities ranging from several kBq to 350 kBq, in
ampoules, and glass or steel containers usable for calibrating radon-measuring instruments.
Acknowledgements
We are grateful to Pr. Jean-Franc- ois Valley for
supporting this project and for his critical reading of the
manuscript. Manuel Santos has our special thanks for
designing the measurement chamber and carrying out all
the mechanical work with great effectiveness. Jean-Jacques
Gostely’s useful suggestions for improving the manuscript
are warmly acknowledged. This work is dedicated to the
memory of Marc Décombaz.
759
References
[1] R. Wiliam Field, J. Duport Philippe, Med. Phys. 30 (2003) 485.
[2] National Research Council, Health Effects of Exposure to Radon,
BEIR VI, Committee on Health Risks of Exposure to Radon (BEIR
VI), Board on Radiation Effects Research, Commission on Life
Sciences; National Academy, Washington, DC, 1998.
[3] J.H. Lubin, et al., J. Natl. Cancer Inst. 87 (1995) 817.
[4] J.H. Lubin, et al., Radiat. Res. 147 (1997) 126.
[5] J.M.R. Hutchinson, J. Cessna, R. Collé, P.A. Hodge, Appl. Radiat.
Isot. 43 (1992) 175.
[6] J.C.J. Dean, M. Burke, Nucl. Instr. and Meth. A 339 (1994) 264.
[7] R. Collé, J.M.R. Hutchinson, M.P. Unterweger, J. Res. Natl. Inst.
Stand. Technol. 95 (1990) 155.
[8] P. De Felice, Xh. Myteberi, Nucl. Instr. and Meth. A 369 (1996) 445.
[9] T.H. Gan, S.B. Salomon, J.R. Peggie, J. Res. Natl. Inst. Stand.
Technol. 95 (1990) 171.
[10] J.L. Picolo, Nucl. Instr. and Meth. A 369 (1996) 452.
[11] R. Dersch, Appl. Radiat. Isot. 60 (2004) 387.
[12] I. Busch, H. Greupner, U. Keyser, Nucl. Instr. and Meth. A 481
(2002) 330.
[13] G Ratel, C Michotte and F O Bochud, BIPM comparison
BIPM.RI(II)-K1.Rn-222 of activity measurements of the radionuclide 222Rn, Metrologia 41 (Technical Supplement 06002) (2004).
[14] G. Triscone, M. Santos, J.-J. Gostely, J.-F. Valley, F. Bochud,
metINFO 10 (2003) 4.
[15] M.L. Curtis, J.W. Heyd, R.G. Olt, J.F. Eichelberger, Nucleus 13
(1955) 38.
[16] S. Pommé, L. Johansson, G. Sibbens, B. Denecke, Nucl. Instr. and
Meth. A 505 (2003) 286.
[17] M. Décombaz, J.P. Laedermann, Nucl. Instr. and Meth. A 369 (1996)
375.
[18] Y. Nedjadi, P. Spring, C. Bailat, F. Bochud, M. Decombaz,
G. Triscone, J.-J. Gostely, J.-P. Laedermann, Primary activity
measurements with 4pg NaI(Tl) counting and Monte Carlo calculated
efficiencies, Appl. Radiat. Isot. A, submitted for publication.
[19] J.-J. Gostely, J.-P. Laedermann, Appl. Radiat. Isot. 52 (2000) 447.
[20] G. Ratel, C. Michotte, K. Kossert, H. Janben, Update of
BIPM.RI(II)-K1.Rn-222 comparison of activity measurements for
the radionuclide 222Rn to include the PTB, BIPM 2006.