Clear Active Contact Lens
(CACL)
Senor Project Report
Advisor:
Dr. Smilkstein
Multidisciplinary Preparees:
BMED BS - Paul Hecker (
[email protected])
BMED BS - Phillip Azar (
[email protected])
EE BS - Alexander Do (
[email protected])
EE 4+1 – Benny Ng (
[email protected])
EE 4+1- Errol Leon (
[email protected])
0
Table of Contents
Section
I.
II.
III.
IV.
V.
VI.
VII.
VIII.
IX.
Pages
Abstract
Introduction
Background
Customer Requirements and Design Requirements
Material Properties
Clearness
Flexibility
Biocompatibility
Electrical Requirements
Size
Power and Heat
Functionality
Design Development
Concept Generation
Top Concepts
1.25 cm Contact Lens - PET - RF Scavenging
2.00 cm Contact Lens - PDMS - Wireless Charging
1.50 cm Contact Lens - PDMS/PET - Wireless Charging
Preliminary Analysis - Design Selection Process
Functional
Power Dissipation and Bioheat Transfer
Radiofrequency Engineering
Battery Design
Integrated Circuitry
Safety Concerns
Manufacturing
Cost
Final Design
Description Final Design
Analysis Results
Cost Breakdown
Safety Considerations
Maintenance and Repair
Differences from Planned Design
Project Realization
Manufacturing Process
Difficulties and Recommendations for Future Designs
Design Verification and Results
Clear Vision
Refractive Index Testing
Circuitry and Antenna Printing Verification
Contact Angle Testing
Antenna Testing and Characterization
Gain
Impedance and Bandwidth
Polarization and Beam width
Radiation pattern
Efficiency
Comfortability
Biocompatibility Assessment
Thickness Testing
Substrate Breathability Testing
Bioheat Analysis
Impact, Conclusion, and Recommendations
Appendix A – Senor Project Analysis
Appendix B – References
Appendix C - Final Drawings
Appendix D - List of Vendors, Contact Information, Pricing
Appendix E - Vendor Supplied Component Data Sheets
2
2-3
3-4
4-8
9-23
24-26
27-34
35-58
58
59-62
63-64
65-70
70-71
72-75
1
I Abstract
The clear active contact lens project aims to address safety and hazard awareness with
an unexplored field of eye wear technology. With advancements in nanotechnology and the
advent of circuits on contact lens, this project is one of the first research and development into
this new field, following University of Washington and Google. The team focuses on the safety
and biocompatibility of the contact lens for a comfortable ease of use. The designs push the
limits of thin film printed technology with its pursuit of fine designs of 250μm antennas. The
project streamlines the manufacturing process for a combination substrate of PET and PDMS
and mounting of antenna, IC, and battery. To produce a product that operates at simulated
specifications, the team tests and characterize the substrate, antenna, IC, and battery
separately, while ensuring their designs function effectively together. The designs and
processes provide a large stepping stone to the realization of a marketable active contact lens.
II Introduction
In today’s working world, a significant amount of attention is placed upon occupational
health and safety. In the United States, a section of the U.S. Department of Labor known as the
Occupational Health and Safety Administration (OHSA) proposes, manages, and enforces laws
detailing safety requirements in the workplace. These efforts have contributed to an overall
decrease of 7% in occupational fatality rates from 2011 to 2012 [1]. The clear active contact
lens project is an effort to continue to improve the safety and quality of life of hazardous
environmental workers by providing them an internal means of awareness in unsafe situations.
The project also seeks to build on already available technology in designing active contact
lenses and improve on them with thinner integrated circuitry, more robust battery design, and
wireless inductive charging.
In a hazardous environment, an individual's awareness of risk factors is critically
important to their safety. In many situations, these risk factors are invisible to our senses, such
as radiation or odorless fumes. The individual must rely on external equipment to provide them
with the information needed to complete their task and keep themselves safe. This equipment
can be large, cumbersome, and in some cases (such as Geiger counters) outdated. The project
aims to facilitate this provision of information by changing the means of delivery. Rather than an
individual having to process the information provided to them by external equipment, the
individual will be able to react to information provided immediately through a small, unobtrusive
device that can be integrated internally. The device will rely on sensory circuitry to process the
information for the user, and when harmful conditions are detected, deliver a sensory stimulus,
such as a flashing light or audio message. Through proper training, this quick reflex can help the
user remove themselves from a dangerous situation much more rapidly and effectively, and
thus lower incidence of injury.
The Clear Active Contact Lens project began as an interdisciplinary effort, and after
2
being awarded the CP connect interdisciplinary project grant in winter of 2014, was realized in
the spring of 2014. Final prototypes were also delivered as a fulfillment of the arrangement set
in biomedical engineering design. Motivation for the project lies within each individual group
members commitment to building devices that contribute to the wellbeing of others, and
financial incentive in the project are a result of a relatively premature market ripe for research
and development.
III Background
The foundation for this project was set by research conducted at University of
Washington by Otis et al. [2]. In their research, a proof-of-concept active contact lens powered
by radio frequency waves was developed and tested. This research demonstrated the feasibility
of a small-form factor application of an integrated device within a contact lens, and can be seen
below in Figure 1.
Figure III-1: Radio frequency powered contact lens. Left side shows unlit LED, right side shows LED when lit. [1]
Contacts are one of the few disposable devices that come into direct contact with the
interior of the body for extended periods of time [4]. A lens equipped with the proper detection
facilities could continuously measure biomarkers present on the surface of the eye. Many of
these biomarkers are likely the same ones analyzed by doctors following a blood test, meaning
that this device could provide a less invasive form of measurement and provide patients with a
more continuous stream of data. Early implementations of this concept can be seen in devices
that have been designed to measure glucose concentrations in tear solution, and were built
using micro fabrication and silk-screen deposition processes [2]–[4].
The device does not need to be complicated to provide functionality [2]. By designing the
device so that it responds to external stimulus, much of the hardware associated with
processing and detection can be offloaded, while maintaining the benefits of an ocular indication
system. To our knowledge, no one has shown a fully integrated device that is capable of
powering an onboard LED in response to external stimulus.
3
Up until this point as well, biocompatibility has largely been neglected in the quest for a
working device. Babak et al. partially acknowledged this concern by coating their device in a
biocompatible substance known as parylene. However, parylene has a significant downside for
ocular implementation, as it is impermeable to oxygen, which is necessary for proper functioning
of the cornea. [8] In addition to biochemical concerns, effects of heat dissipated by the battery,
circuitry, and antennae will have to be taken into account when determining the safety of the
device. Since the device is electromagnetic in nature, it will have to comply with a wide variety
of standards governed by different regulatory bodies including the FCC and FDA [9].
IV Customer Requirements and Design Requirements
The goal of this project is to create a prototype that can prove the efficacy of an active
contact lens. In our design process this year, we identified a number of customer requirements
that are most important in the usage of the device, and translated them into engineering
specifications. The first requirement we could gather was that the lens needed to be clear. Our
initial target customer base is largely situated in hazardous waste control. In this occupation,
unclouded vision is absolutely necessary. The individual must be completely aware of what is in
front of them and around them through the use of visual recognition, since environmental suits
will obstruct other senses, such as smell, hearing, and touch. Another requirement is
comfortability, since the individual will more than likely not be able to perform any field
adjustments on the lens itself. This is by far one of the most difficult requirements to incorporate
into device design. As such, any materials used will need to be inspected for minimal adverse
reactions, with the most important being inflammation, when in contact with the body. Finally,
the terminal requirement is isolated functionality. To remain competitive in the young market
place of active wearable devices, the clear active contact lens must not require any input or
command from the customer. The device must maintain autonomy from the user so that he or
she can focus on the task at hand. When the device has meaningful data to provide, it will do so
with a designated stimulus, in our case an LED. Otherwise, it will remain off and out of the way.
In developing accurate engineering specifications, the clear active contact lens project
group opted to split the device into separate categories of requirements that are interlinked:
material properties and electrical properties. An updated design table of design requirements
has also been included in the appendix for reference.
Material Properties
Clearness
The clearness of the lens is largely governed by the material used to fabricate the lens,
as well as any surface coating applied to the lens surface. The refractive index of a material, nD,
is a dimensionless number that describes how light and other radiation propagates through a
material. It is obtained by the following equation, which divides the speed of light in a vacuum by
the speed of light in a substance:
4
(EQ 1)
Since the prototype will not be a prescription lens, a best fit material would have a
refractive index close to 1, meaning that light entering through the lens would be refracted a
very small amount, and the individuals vision would not be blurred significantly. This, along with
biocompatibility and flexibility, will govern the materials selection process in our project. It is
important to note that achieving a refractive index of 1 with commercial polymers is unlikely.
Commercially available polymers have refractive indices that range between 1.3 and 1.7. [10]
As such, we have determined that a refractive index between 1 and 1.5 will be acceptable for
our application. This range of indices is acceptable because the eye itself will refract light based
on a desired level of acuity. If the refractive index is within the range we have specified, the eye
should be able to distinguish objects without any loss of acuity.
Also in the way of clearness is the placement of any circuitry on the lens. In the current
design, the pupil area of the lens must be free of any and all obstruction by circuitry. This
includes up to 9 mm in diameter of obstruction-free material. Any circuitry on the lens must be
clear circuitry, or be such that the eye will not be able to resolve the obstructing items at any
time.
Flexibility
The next requirement that will largely impact the design is flexibility. The defining aspect
of this project is the development of a contact lens. As such, any material chosen to support that
design must be able to flex to the curvature of the surface of the eye, and in doing so not incur
significant stress as to damage the lens itself. The method of quantifying flexibility relies on the
computation of the stiffness, k, of the material. This is given in relationship to the young’s
modulus, E, by the following equation:
(EQ 2)
Where A is the cross sectional area of the sample and L is the length of the sample. In
general, the higher the value of k, the stiffer the material becomes. In materials selection, this
means that the higher the young’s modulus of the material, the stiffer it becomes per element.
However, the lower k is, i.e. the more flexible the material is per element, and the lower the
young’s modulus becomes. A material with too low a young’s modulus can be at a higher risk of
plastic deformation, and in the worst case, failure. Failure in this category is defined by
exceeding the ultimate tensile strength of the material and resulting fast fracture. It is important
to note that the above equation represents axial stiffness. In the current problem,
generalizations regarding the bending of the lens as a two dimensional stiffness problem can be
5
made by assuming young’s modulus is isotropic for the contact lens.
The target for flexibility is measured directly by Young’s modulus, E, of the material
chosen. A large range of these values are listed in table III in the appendix. This variance is a
result of the two top material candidates having opposite flexibility, with PDMS being relatively
compliable compared to PET.
Another factor which is directly impacted by flexibility is the integrity of any onboard
circuitry. The circuitry must be able to bend without incurring significant damage or wear. The
device will be tensile tested before and after circuitry is laid on, so as to characterize the
mechanical durability of the device.
Biocompatibility
A staple of biomedical design is the process of determining the suitability of a material in
a medical device application. This means that the material must not do harm to the body within
the intended function, and the body must not do harm to the material. This is by far the most
difficult requirement to contend with from an engineering perspective because it does not have a
quantifiable nature. Instead, biocompatibility analysis is performed during acute and chronic
studies on cells, living systems, and even live patients. The data from these studies is then
compiled and analyzed by a team of biomedical engineers, who look at notable examples of
bioactivity such as inflammation, systemic toxicity, and rejection. Since the product will be in
direct contact with the user’s eye for the duration of their task, any material selected as the base
for the lens must not be sufficiently biocompatible.
Many commercial polymers used in the biomedical industry have already been
characterized for biocompatibility. This, however, can be a double edged sword. This data may
not be for the region of interest, i.e. the surface of the eye. Also, this data can limit the materials
selection process significantly, to the point where a suitable material may not currently exist. As
such, determining a material from the perspective of biocompatibility is a peripheral effort: it is
impossible perform studies with resources currently available. A reliance on currently available
data will be the primary means of analysis of biocompatibility.
Currently, both PET and PDMS are characterized as relatively biocompatible, with the
former being a known irritant due to its relatively low oxygen permeability when compared to
PDMS. PDMS stands currently as the best material for biocompatibility when in contact with the
eye, as many modern contact lenses utilize PDMS as a substrate material. What this means for
the clear active contact lens project is substantial: the usage of a PDMS substrate would
significantly strengthen the pursuit of a 510(k) clearance with the FDA.
Electrical Requirements
6
Size
Size is an important parameter for the design of any electrical components on the
electronic side of the lens. The contact lens itself is very small, with a maximum surface area of
.8 cm2. No circuitry will be allowed in the area bound by the pupil, thus further decreasing the
area available for electronic components. The resulting size budget changes the method in
which the project will approach manufacturability of the lens itself. The small size also directly
impacts the functionality, which will be limited by the physical constraints associated with small
circuits [11]. Micro fabrication is a likely route to remedy this issue, since printed circuitry may
result in critical conductor discontinuities. This, in turn, would create a non-functional device.
Determining what is too big and what is too small will come directly from the cost and
relative need of both. Since the electrical components of our design include an integrated circuit,
an antenna, and a battery, the budgeting will be dependent on the importance of the component
and the cost required to manufacture at that size. As of early March, the clear active contact
lens electrical size budget will include a narrow strip revolving around the eye equivalent to the
difference of the total area and the maximum pupil dilation area. Further technical information
regarding constraints specified by the client can be seen in the Appendix under IC Design and
Antennae.
Power and Heat
Modern electronics run largely on AC dynamics, and as such, power generated and
used by a circuit will consist of two components: real power and imaginary power. From a
systems perspective, the power that has the potential to be dissipated as heat comes in the
form of real power, given by the following equation:
(EQ 3)
Where Vrms represent root mean square voltage, Irms represents root mean square
current, and θ represents the phase angle. This power is in units of Watts, and is consumed by
the system in question. Calculations can be done after obtaining real power to determine the
heat dissipated within a specific timeframe of use, by multiplying the real power by the total
active time. In our design, the active portion of the contact lens must generate and use a little
power as possible, due to the danger of excess heating in the eye and electrical safety.
Comfortability of the user will directly depend on how hot the circuitry becomes.
In the project, it will be assumed that the system will not become hot enough to damage
the onboard circuitry (i.e. temperatures above 60 degrees Celsius.) COMSOL simulations
performed indicated that given the maximum power output of 10 mW. In this regard, maximum
allowed change in temperature is given as 1.5 degrees Celsius, and will be tested with PET,
PDMS, and combination lens models.
7
Functionality
The device’s functionality will be the primary determinant of the efficacy of the overall
product. If the device cannot provide a degree of functionality that will alert the user to
hazardous environmental conditions based on a preset threshold, then the device will fail to
meet FDA efficacy requirements. The device also has to improve in some way on past iterations
of active contact lenses. This requirement is largely accounted for by the concept of isolated
functionality, in which the device requires no interaction from the user besides putting the
contact lens on properly to perform within specified limits.
This requirement necessitates the creation of an external component that communicates
with the contact lens itself. Sensory circuitry is too large to fit in the budgeted area of the contact
lens, and must be placed off board. This off board circuitry will be contained within a small
device on the user in order to process any information collected by the onboard integrated
circuitry. The antenna will provide a means of communication between the two circuits, and
effectively “link” them. Antenna design and verification is largely based on the mode of
communication used, in this case, the IEEE 2.4 GHz band. The radiative resistance must be
high enough to send and receive signals, and the inductance must be such that wireless
inductive charging is permissible. A spectrum analyzer will be sufficient to test the
radiofrequency properties of the antenna, whereas simulations must be used to determine the
inductance of the proposed antenna design prior to manufacture.
8
V Design Development
Concept Generation
Indicating danger to a user in a low foot-print setting allows for a wide variety of
solutions. Any of the five senses could be stimulated to indicate danger. To narrow down the
options we limited it to sight and touch, as these sense could be easily stimulated with
technologies we had experience with. From here, we decided to focus on the eye. Our sponsor
had a strong desire to make the project medically oriented, and since the eye is one of the most
unique sources of biometric information that can be obtained noninvasively, we decided to focus
our efforts there.
Once a focus had been established on the eye, the question became whether to use a
external framework or an on-eye solution such as a contact lens. Products such as a google
glass already existed, so we looked into how we could capitalize upon this to provide active
functionality. It was soon apparent however, that while google glass was an incredible
technology, it was more obtrusive and was designed for much more complex interaction than
we required. The focus then shifted towards contact lenses.
Our group had limited experience with active contact lenses, but healthy imaginations.
We drew on fantastical ideas such as x-ray vision to more realistic lenses that changed the color
of the eye. With ideas such as this in mind, we went forth and did a comprehensive literature
review. Here we found a small but blossoming field of active contact lenses. These lenses
changed color in response to tear glucose levels, measured the pressure of the eye, and could
receive wireless signals that activated LED’s on the lens [5]. While many of these lenses were
impressive in concept, a major challenge that had not been overcome was transporting data
from lenses externally without the use of wires.
This challenge seemed beyond the scope of the project, so we instead focused our
efforts on making a lens that had an on-board power source and could actively respond to
indicated danger to users. This would allow us to create a fully functional device that could
operate in conditions, such as inside a helmet or during sports, that a more external device such
as google glass could not. It also allowed us to take one step closer towards a fully functioning
biosensor by acting as a proof of concept for a fully-functioning embedded lens system,
something which has never been done before and also met our sponsors desire to keep the
project medically oriented. With this path in mind, we began translating our general concept into
more concrete designs that would fulfill our design requirements.
9
Top Concepts
Since the clear active contact lens project was a heavily interdisciplinary, the three top
concepts listed below largely represent the physical parameters of the lens. Each individual
component, including the antenna, battery, integrated circuit, substrate material, and indicator,
will fit into any of these designs. Rather than discuss the heavily iterative and rapidly changing
design process for each individual component, we will focus more on discussing the implications
of each component and the analysis performed to assist in the decision making process.
Figure 1: Map illustrating component connection to physical design.
1.25 cm Contact Lens - PET Substrate - RF Scavenging
The original design of the active contact lens was that of a typical contact lens, with a
1.25 cm diameter. The RF antenna would be laid onto the substrates around the rim until the
proper inductance was reached for wireless charging. The differentiating factor of this lens,
however, is that it would have to primarily rely on RF scavenging to recharge the battery, as the
current tests indicate an inductance too small to support wireless charging at 2.4 GHz. Then,
the integrated circuitry and thin-film battery would be placed parallel to the LED at the center of
the lens, connected by conductive tracing ink. A layout sketch of this design is shown in figure
V-2 on the next page.
10
Figure 2: Layout sketch for 1.25 cm diameter lens design.
What is distinct about this design is that it was originally intended to be an all PET lens,
following suit with work done at the University of Washington. PET is not only very easy to
obtain in very thin sheets (~25 microns per sheet,) but it is also readily compliable when heated
to 180 degrees Celsius, allowing for an easy curving process. Another benefit of PET is that the
antenna can be printed directly on using a screen printer, simplifying the manufacturing process
immensely. At first glance PET seems like the right choice for hosting a functional platform on a
contact lens. PET, however, is not a comfortable material. It is traditionally considered a “hard”
lens material and has a rough surface similar to paper. Also, after it has been curved, it has a
tendency to crumple like paper around the edges, damaging the edge of the device and
potentially preventing proper usage.
2 cm Contact Lens - PDMS Substrate - Wireless Charging
A newer design that takes cues from scleral lenses was conceived to produce better
inductance for the wireless charging capabilities of the lens. The differences in this design and
the previous include the type of substrate, the size and number of turns of the antenna (which
correlate directly to inductance,) and the placement of the battery and integrated circuitry. A
layout sketch is shown in figure 3 on the next page.
11
Figure 3: Layout sketch for 2 cm diameter lens design.
While this lens is clearly much larger, its conception was largely a result of a large
quantity of simulation performed on both electromagnetic and thermal properties. The larger
lens allows for the implementation of a much more powerful wireless platform that can both
charge the embedded battery and receive incoming RF signals. This lens also is completely
constructed of PDMS, a material well characterized for application where in vitro and in vivo
biocompatibility is a major factor.
When our team evaluated this idea, we came to the consensus that while it solves many
of the problems of the 1.25 cm design, it comes with a host of manufacturing challenges. The
first is that in order to make a full PDMS lens, the entire process must be done in a micro
fabrication space. This means that actually generating a prototype will expensive, arduous, and
relatively temperamental. Additionally, a number of the resources needed to make an all PDMS
design work are restricted to micro fabrication trained individuals. This means that much of the
development of a proposed prototype would have to be supervised, which could potentially
further slow the development process.
12
1.5 cm Contact Lens - PDMS Exterior, PET Interior - Wireless Charging
The final design concept our team generated was a fusion of the previous two designs,
and an appropriate middle ground. With a proposed 1.5 cm diameter, wireless charging, and
dual composition, the 1.5 cm lens represented the strongest and most feasible candidate for a
prototype in our design process. The caveats, however, were apparent from the start. The lens
would have to be manufactured in many more steps than previous iterations. Instead of a two or
three step process, we now had a five step process over many hours. Also, since this lens
design had effectively three layers, the risk of air gaps and material separation was much
higher. A sketch of this design can be seen in figure 4.
Figure V-4: Layout sketch for 2 cm diameter lens design.
13
Preliminary Analysis - Design Selection Process
The selection process was broken into three stages: functional requirements,
manufacturing requirements, and cost limitations.
Functional
The first stage focused on meeting the functional requirements. This required that the device
meet the material and electrical properties specified in the design requirements, in addition to
achieving its overall goal of operating as a warning indicator embedded in an active contact
lens. Functional requirements were extensively tested to determine the overall direction of the
project. The five major areas of concern included power dissipation and bioheat transfer,
antennae design, battery design, integrated circuit design, and safety concerns.
Power Dissipation and Bioheat Transfer - COMSOL Simulations
The goal of simulations performed this quarter was to characterize the thermal effects of
the active contact lens on the human eye for prolonged periods of time. Simulations were
performed with both candidate substrates, PET and PDMS, and circuitry was simulated as a
constant power source. Bioheat transfer analysis based upon the Pennes bioheat equation was
performed in COMSOL with a close approximation of the finite element model proposed by
Gokul K.C. et al. [12]
Figure 5: Proposed finite element sketch of Human eye.
Where T0 through T6 represent nodal temperatures up to the nth element of the model, and L1
14
through L6 represent distances from the surface of the cornea to the nth element of the model.
The above model was used primarily as a reference in which to approximate our simplified
model, seen below:
Figure 6: 2D Axisymmetric COMSOL sketch of the human eye.
The above model was simplified significantly primarily to reduce computational time
required in COMSOL and to reduce complexity of the resulting temperature distribution. Also, it
can be seen that at the top of the model, around r = 12 mm, we have included an approximation
of the contact lens, with a varying thickness of 250 to 400 micrometers. The thickness is such
that at the center, it is smallest, and at the edges, it is largest. This variance was designed in
order to remove the need for a complex Bezier polygon geometry when simulating the contact
lens in COMSOL.
In our model, we assumed that the temperature of the surface of the cornea was 35
degrees Celsius, 2 degrees Celsius lower than the rest of the eye. There are two primary
reasons for this assumption. First, convective heat flux at the surface of the eye associated with
tear flow and gas flow naturally reduce the surface temperature of the cornea. Second, surface
to ambient radiation will further reduce the temperature of the surface of the eye as temperature
increases. Blood perfusion rates within the ciliary body and retina and various thermal
properties of the eye were extracted from works by Gokul K.C. et al. and Cvetkovic et al. [13]
Using a 10 mW total power dissipated by the contact lens, the following temperature
distributions were obtained from a time dependent simulation with 180 second end time and 0.1
second step size for both PET and PDMS:
15
Figure 7: 3D surface temperature distribution after 30 second exposure - PDMS
Figure 8: 2D surface temperature distribution after 30 second exposure - PDMS
16
Figure 9: 3D temperature distribution after 30 second exposure - PET
Figure 10: 2D surface temperature distribution after 30 second exposure - PET
For the simulations shown above, PET has a thermal conductivity of 0.24 W/m*K, a
density of 1380 kg/m3, and a specific heat capacity at constant pressure of 1000 J/kg*K. PDMS
has a thermal conductivity of 0.15 W/m*K, a density of 970 kg/m3, and a specific heat capacity
17
at constant pressure of 1460 J/kg*K. What can be determined from the above results is that
PDMS is more thermally insulative than PET, and when exposed to the same conditions,
causes a higher average temperature throughout the cornea. The gravity of this difference,
however, is quite small. The values are within 0.005 degrees Celsius of each other. More
importantly, however, is that the above bioheat simulations indicate that even at our maximum
allowable power dissipation, the temperature change throughout the cornea is very small,
ranging from a 0.5 to 0.7 degrees Celsius increase in both cases. From this it was concluded
that from a bioheat transfer perspective, both substrates were equivalent and therefore the size
is irrelevant, and the power budget for the onboard circuitry was appropriate.
Radiofrequency Engineering - Antenna design and Inductance
The primary goal of radio frequency engineering performed this quarter was to
characterize the efficacy of the antenna as not only a receiver component of the active contact
lens, but also as an inductive coil capable of charging the onboard battery. Since all
components are largely immovable following manufacturing, without proper inductance from the
antenna, sufficient power transfer will not be achieved, and the active contact lens will be a one
use product. Since a nearly five thousand dollar disposable contact lens has no real market, the
antenna is a crucial aspect of the project.
Three different antenna designs were investigated when selecting an overall design for
the project. The first design, shown in figure 11, consisted of a continuous loop antenna of
uniform trace width. This antenna was conceived under the assumption that the antenna could
double as both an inductive coil and signal receiver. The second design, shown in figure 12 on
the next page, attempts to improve efficiency at 2.4 GHz by adding an inner coupling loop. The
third design, shown in figure 13 on the next page, is an adaptation of an antenna designed by
Steve Yates that uses a small inner loop to improve efficiency at 2.4 GHz.
N (turns)
Di (inner diameter)
Do (outer diameter)
W (trace width)
S (trace spacing)
8
7885um
11630um
117um
117um
Figure 11: First antenna design, consisting of one solid loop of uniform trace width.
This was the first design with the assumption that we can use the inductive coil as the
antenna. In order to fit the outer diameter size constraint, we were not able to use a λ/3 design,
18
so we used a 7λ/24 circumference, which has an 11.63mm outer diameter. This gives better
efficiency than λ/4 while remaining under 12mm diameter. This coil has 8 turns with an inner
diameter of 7.885mm, which is only 1.5% less than our inner diameter requirement. Out of our 3
designs, this had the highest inductance, but severely lacked power efficiency at 2.4GHz. Our
simulation gave us an antenna gain of -22.3dB at 2.4GHz, which was the lowest of our designs.
Using Harold A. Wheeler approximations for planar coils operating under 30MHz (𝐿(𝜇𝐻) =
𝐷𝑜2 ×𝐴2
,𝐴
30×(𝐴−11)×𝐷𝑖
=
𝐷𝑖 +𝑁(𝑊+𝑆)
),
2
we calculated the inductance to be 1.005μH. This design is by far
the best for inductive charging.
Antenna Ring
Antenna Ring
Antenna Ring
Inner coil
Inner coil
Inner coil
Inner coil
Inner coil
diameter
trace width
trace spacing
N (turns)
Di (inner diameter)
Do (outer diameter)
W (trace width)
S (trace spacing)
11900um
500um
103.5um
1.605
8000um
9262.7um
250um
250um
Figure 12: Second antenna design, consisting of two separate loops of uniform trace width.
This second design makes use of a large trace antenna outer loop and a 1.6 turn inner
coupling coil to achieve a gain of -14.5dB at 2.4GHz. This greatly improved gain sacrifices
inductance of 20nH, which is based on the minimal windings of the inner coil. This was able to
almost double its power efficiency by doubling the trace width of the antenna coil.
Figure 13: Third antenna design, consisting of two separate loops of non-uniform trace width.
19
This dual loop design uses an inside coil, even though they are not connected, couples
into the actual antenna and has a direct effect on the resonant frequency of the outside loop.
The loop antenna has a small coupling loop (roughly 1/5 of the size of the antenna), which is fed
the signal instead of the larger loop. The thicker loop is the antenna. The outer diameter is
about 1.3cm and the inner diameter is about 7.2mm. Using a thicker trace for the antenna,
Benny was able to improve efficiency compared to sticking to a uniform trace width.
After all three antenna were simulated, it was determined that regardless of substrate
and size, only the dual loop antennas would be appropriate for achieving the desired
functionality. Figures 14, 15, and 16 below present the results from these simulation.
Figure 14: Antenna design 1 efficiency and power results.
Immediately from figure 15 above we see that antenna design 1 will not be a sufficient
platform in providing wireless charging. The efficiency, which sits a 0.537%, is too low to
sufficiently facilitate power transfer.
Figure 15: Antenna design 2 efficiency and power results
20
Antenna design 2, unlike design 1, displays much more promise as a platform for
wireless charging. With an efficiency at 20.715%, the design is more than viable for facilitating
wireless power transfer.
Figure 16: Antenna design 3 efficiency and power results.
Antenna design 3 follows closely in line with design 2, albeit with a lower efficiency.
Ideally, all the antenna designs will be able to receive signals regardless of efficiency. To meet
our design requirements, however, it was imperative to choose a design that best suited the
wireless charging platform. While antenna design 2 represents the highest efficiency, it has a
much lower effective area, making it less suitable for wider range of communication distances.
Geometrically, antenna design 2 also interferes with our clear vision requirement, just barely
clearing the maximal pupil dilation diameter requirement. As such, we decided that antenna
design 3 would be the best option to pursue towards the final design.
Battery Design - Charging considerations
During the research phase, different types of portable power supplies (batteries and
capacitors) were compared and contrasted. Not only was power, size, and flexibility a concern,
but also health hazards from the materials being used. Graphene supercapacitors represent a
potential answer to these requirements. Supercapacitors have unique properties that make
them an attractive choice as a power source. They are essentially maintenance-free, possess a
longer cycle-life, require a very simple charging circuit, experience no memory effect, and are
generally much safer [14]. Also, one of the key characteristics of a supercapacitor is that it can
be charged and discharged at high rates.
Graphene is a two dimensional crystalline allotrope of carbon where the atoms are
densely packed in a sp2-bonded hexagonal pattern [15]. Graphene can be described as the top
skin layer of graphite. High quality graphene (pristine graphene) is not only an excellent
conductor, but it is also strong, flexible and nearly transparent. The conductivity of a single-layer
graphene sheet has been measured up to 649 S/cm [16]. The most important property of
21
graphene with respect to supercapacitors is the high theoretical specific surface area of 2675
m2/g which corresponds to a theoretical specific capacitance of 550 F/g [14]. The target
capacitance value for the fabricated supercapacitor that was calculated is 32mF.
Unfortunately, graphene supercapacitors are highly experimental. Even after conducting
large amounts of research into integrating the technology into our lens, our group was unable to
fabricate the component in a reasonable time frame. Thus, we decided to use the current
paradigm of thin film lithium ion batteries, which are readily mass produced by manufacturers
such as Cymbet. While not as flexible and thin as a graphene supercapacitor, thin film lithium
ion batteries are affordable and easy to use, making them a good choice to demonstrate our
embedded power platform. One such battery is shown in figure 18 below. Unfortunately, this
battery is not appropriate for the final implementation of the contact lens due to biocompatibility
issues with lithium.
Figure 17: Thin film lithium ion battery, with penny for scale.
Integrated Circuitry - Protection considerations
The majority of the design selection process in this project stemmed from requirement
considerations associated with the battery and the antenna. The integrated circuitry will work on
all of the proposed designs, regardless of substrate, size, and charging platform. The circuitry,
however, must operate such that it is not damaged by the discharging of the thin film battery
and the radiofrequency transmission from the antenna. Therefore, much of the design work on
the integrated circuitry revolved around designing protection circuitry ensuring both regulated
charging voltage and low battery voltage cut off to preserve battery longevity. The new complete
circuit is shown in the figure 18 on the next page, with accompanying charging circuitry:
22
Figure 18: Integrated Circuit schematic, including input and output terminals.
Figure 19: Enhanced image of charging circuitry.
The two red boxes above indicate the new protection circuitry developed this quarter. In
box one, a three stage charge pump rectifies an input sinusoidal wave to ensure it matches
specific requirements listed in the appendix. In box two, a voltage regulator clamps the charging
input voltage to 4.1 volts to ensure that the batter receives optimal voltage and is not damaged
during charging.
23
VI Final Design
Our final design consisted of a circuit platform embedded in PDMS. The base of the lens
consisted of a printed antenna upon a PET substrate. The printed and insulated electro chromic
ink lays on the same layer as the PET. There are additional printed insulation layers on top of
the antenna upon which the battery and IC are bonded using a thermo-epoxy. The battery, IC,
and electro chromic are connected via wire-bonding, which can be seen in Figure 2 on the next
page. The entire system is embedded in PDMS.
Our modular focus for the analysis and testing of device components resulted in a welldefined set of geometries and component characteristics that integrated to form our final device.
IC simulations using Cadence Virtuoso showed a total current draw of 5.41uA. Inductive
charging required 4.5Vp at the input of the chip. The display will activate when the IC sees 1
dBm from the wireless sensor. Based upon the desired material characteristics and size
constraints of the Antenna and IC, the geometry of the lens was designed to match the
curvature of the eye as closely as possible. The edge-to-edge diameter of the lens was 1.5 cm,
with a thickness of 950 ums. This allowed for a 50 um layer of PDMS on the top and bottom of
the lens with 200um of edge coverage.
Figure 1: Exploded view of lens highlighting individual layers and components. Wiring has been removed for clarity.
24
Figure 2: The wiring schematic above shows interconnections between device components.
Analysis Results
Preliminary analysis of various aspects of the initial three designs indicated the chosen
final design was not only the most feasible option, but also the most robust. Bioheat analysis of
PET and PDMS indicated that both polymers would be suitable for the construction of a lens
exposed to a constant 10 mW power dissipation. By encapsulating a PET frame in PDMS,
however, an extra layer of thermal insulation is created, and it is possible to further isolate the
eye from the power dissipated by the lens. Theoretically, this will decrease the temperature of
the cornea when the device is active, and provide for a more comfortable experience.
From a functionality perspective, the antenna design played a role in choosing this
design. While 2 cm would allow us to achieve a higher efficiency, a 1.5 cm is a more
contemporary size for contact lenses. Many individuals do not wear scleral lenses as they are
larger and more uncomfortable. From a manufacturing perspective, while 1.5 cm lenses are
smaller, the difference is negligible in terms of modern machining tolerances and would be
supported as well as a 2cm design.
Cost Breakdown
The costs below do not include materials supplied by Cal Poly such as screens, PET,
PDMS, or access to facilities. This project was focused heavily on R&D.
●
●
●
$500 CNC tech hours + materials
$1744.95 R&D graphene supercapacitor
$323.94 Electrical components and thin film battery
Total Expenses: $2568.89
25
Safety Considerations
With any biomedical device, biocompatibility is a top priority. Ensuring that the body
does not harm the device and the device does not cause harm is the foundation of medical
engineering, and in our case this applies even more readily since the contact lens is interfaced
with a highly sensitive organ. Materials selection should be noted here as a point of special
consideration. PET was found to be, according to a study conducted in 2010, an endocrine
disruptor when exposed to a biological system [17]. This is a definite problem when applied to
all three designs. A design that incorporates a barrier between the PET and the surface of the
eye would be the most desirable to counteract this finding.
While literature reviews of PDMS have indicated it is a biocompatible material with
minimal side effects we do not advise usage of lens for extended periods of time. Both in vitro
and in vivo testing must be done before the lens can be safely used. These tests are also
critical, because they account for parameters that are difficult to measure, such as the leaching
behavior of the PET, electrochromic ink, lithium battery, or substances from the bare die IC
using simulation. At this point, safety concerns have been addressed as much as possible to
ensure a device that is both safe and effective.
Maintenance and Repair
Since this device is acting as a contact lens, user maintenance of the device is critical.
Traditional contact lenses are stored in a saline solution overnight to prevent build-up of proteins
over time. Additionally, while PDMS is oxygen permeable, the eye still does not receive as much
oxygen as if it were exposed to the atmosphere. Therefore, the user will likely be instructed to
remove the lenses after 8 hours or less similar to traditional lenses. While the user may be able
to wear the device for up to 8 hours, the device will only be functional for 1 hour, after which it
will need to be recharged. Recharging time is estimated to be less than 1 hour, and could
increase in the future as battery capacities grow. Lastly, the electronics in the device are
completely encapsulated, rendering any attempts at repair unlikely. The device will likely act as
a monthly or yearly lens to reduce cost of replacement.
Differences from Planned Design
Our final prototype departed from our final design due to a delay in the IC fabrication
process. Without the IC, it was impossible to assemble a complete circuit. As a result, focus was
placed on developing the manufacturing process used to create the lens. The final lens
incorporated the PET antenna, battery, and PDMS encapsulation to come as close as possible
to the final design within our resources.
26
VII Project Realization
Manufacturing Process
The lens manufacturing process was designed to be simple and robust. The process
overview involves four major steps: printing the lens, curving the lens, attaching the
components, and embedding the circuit in PDMS. To carry out this process, manufacturing
techniques from across disciplines were used. Screen-printing machines in the GrC department
were used to fabricate the antennas. The molds were designed using SolidWorks in the BMED
computer lab and manufactured in the Bonderson Machine shop. The PDMS was mixed and
degassed using the Micro fabrication lab facilities.
Outline of Manufacturing Process
1.
2.
3.
4.
5.
6.
Screen Print Antenna
Attach PET Substrate to Male Mold 1
Place Mold in oven at 180 degrees Celsius for 10 minutes
Attach components using thermo-epoxy
Use Mold 2 to build top layer of PDMS
Use Mold 3 to build bottom layer of PDMS
27
Step 1:
The first step in screen printing is to develop a mask of the design you want. The mask,
seen below in figure 1, is designed using Adobe Illustrator and printed using the Cyrel Digital
Imager, seen below in figure 2.
Figure 1: Mask for antenna fabrication.
Figure 2: Cryel Digital imaging machine printing antenna mask.
28
The next step requires applying an emulsion layer to a 420 lines per linear inch mesh.
This is very fine screen size that allows features of up to 10 um to be printed. The emulsion was
then cured using UV light. The mask was placed in between the screen and the UV. The areas
of the mask that block the UV light prevent the emulsion from curing. This allows it to be washed
away, as seen in figure 3 below. This process must be done in a dark room to prevent light from
curing emulsion and preventing its removal. If emulsion cures it will remain on the screen and
create holes in the final printed antenna design.
Figure 3: Removal of emulsion to create screen features.
29
In the final step of this process, the PET is placed on the machine, on top of which the
screen is placed. The ink is then applied and the automatic applicator is engaged. It is critical to
apply the silver ink quickly and engage the applicator, otherwise the ink will dry and ruin the
features of the antenna. This process can be seen below in figure 4. The final product can be
seen in figure 5 on the next page.
Figure 4: Application of the silver ink using the ATMAS machine.
30
Figure 5: Printed antennas on PET substrate.
31
Step 2 and 3:
The PET with the printed antenna is a planar object. To properly orient the antenna on
the mold, the antenna is manually centered on the male component of the mold and taped into
place on the side. The Female Mold 1 is then placed on top of the Male Mold 1 and placed in
the oven at 180 degrees Celsius. After 10 minutes, the mold is removed from the oven. The
oven and mold are both hot, so it is important to insulate all body parts from the heat. The mold
is then left to cool for 15 minutes down to room temperature. The mold is then separated and a
representative image of the result can be seen in figure 6 below.
Figure 6: Molded PET antenna.
Step 4:
The components are then attached to the device using a thermo-epoxy. This step was
not completed during the course of this project, as not all components were available. This will
be completed in the near-future.
32
Step 5 and 6:
The encapsulation of the lens in PDMS occurs in two stages with two separate molds.
For the first mold, PDMS is poured into the bottom of Female Mold 2. The PET lens is then
attached to the center of the male component, and placed component side down into the
PDMS. A common issue with PDMS is the formation of bubbles during the pouring process. To
overcome this, the mold is then placed in the degassing chamber to remove all bubbles that
have formed. The mold is then place into the oven at 70 degrees Celsius for 4 hours to cure.
For the second stage of the process, the lens is placed PDMS side down into the third mold,
and the PDMS is poured onto the PET side making a small pool and can be seen below in
figure 7. The male component is then inserted, and the cavity formed creates a 50um layer of
PDMS on top. The lens is then cured again for 4 hours at 70 degrees Celsius.
Figure 7: Pouring of PDMS to form external lens layer.
33
Difficulties and Recommendations for Future Designs:
Curving of the PET lens proved to be a much easier process than expected, however it
still had its challenged. Wrinkle formation was a problem, as many times the edges of the
antenna would have a fold or two. However, the center of the lens usually stayed free from
wrinkles. In the future to prevent this, I would use a much large male sphere over which to pull
the PET. This would help to ensure that the antenna at the center of the lens was free from
wrinkles and damage.
Forming the PDMS encapsulation layer provided multiple challenges including
separation of the mold following degassing and rouge edges. Following curing of the mold, it
was impossible to separate the male and female components by hand. Eventually, a razor blade
and a hammer were used to pry it apart. It was found that excess PDMS had escaped from the
center of the mold and dried in between the male and female mold faces. It was likely the bond
between the PDMS and the mold is what made it difficult to separate them. Apparently this is
not an uncommon occurrence in inject molding, and a simple solution is to machine a bevel in
the side similar to ones on a laptop to make it easier to pry open. To remove the rough edges of
the PDMS, manufacturing a separator similar to a cookie cutter would eliminate many of the
problems associated with manual removal via exact knife, such as rough edges and damage to
the mold itself.
34
VIII Design Verification and Results
To verify the final design, testing on both individual components and the final
manufactured lens was conducted. Each test was designed with the sole purpose of evaluating
the specific components ability to meet the design requirements put forth at the beginning of the
project. Table I below summarizes the testable design requirements.
Table I: Design requirements
Clear Vision
The focus of testing performed to assess the clear vision requirement was to ensure that
the product caused the least interference with the users vision at all times. As such, elements
such as refractive index, circuit geometry, and contact angle were analyzed.
Refractive Index Testing
In refractive index testing, a simple fixture was created using equipment found in the
Biomedical Engineering Lab of Engineering IV. The following materials were used: 650 nm, 10
mW laser pointer, solder assist clamps, tape measurer, a level, protractor, a white board, and
whiteboard markers. Figure 1 on the next page shows some of the materials used in proving the
fixture.
35
Figure 1: Materials used in designing refractive index test fixture.
The fixture involved attaching the laser pointer to a protractor, placing it on a level
surface, and fixing the test subject at a point at least 2 feet from the laser pointer. The laser
would then be directed using the protractor at the subject material, in this case the contact lens,
at varying angles. A white board placed behind the lens would be used to mark the location of
refraction through the lens, and after measuring the x and y distances from the lens, the angle
of refraction would be computed using trigonometry. Using Snell’s law, the refractive index of
the lens would then be determined. After 15 different attempts, data collected was discarded
due to a standard error that exceeded a tolerance of more than 3 standard deviations away from
the average.
This test ended inconclusively for a number of reasons. The first most jarring reason was
the difficulty in controlling the absorption of the laser as it entered the lens. Since the lens is a
compound material composed of two PDMS layers around a PET skeleton, there is a high
chance that some of the light will be absorbed and dispersed as it passes through the lens. This
effect was so noticeable, in fact, that the prototype incurred thermal damage when examined
immediately following the first test. To account for this effect, future tests should include
calculations to determine if the effect of absorption will damage the lens, and therefore make
determining refractive index impossible. Another challenge associated with ascertaining the
refractive index is the nature of the test itself. At the time when the tests were performed, it was
assumed that the light exiting the lens would be refracted. This is, however, not correct, as light
exiting the lens will retain the angle at which it entered, since it is reentering the medium in
which it originated. This is in accordance with Snell’s law of refraction. As such, it may be more
appropriate to purchase a refractive index tester, such as one manufactured by Presidium
Electronic, to determine the refractive index accurately.
36
Circuitry and Antenna Printing Verification
This phase of verification involved determining whether or not the screen printing
method of manufacturing the antenna produced consistent inner diameter free of obstruction.
Materials used in this test include calipers accurate to 3/1000th of an inch, 20 screen printed
antennas, and the prototype lens. Figure 2 below shows an example of this test using a ruler.
Figure 2: Example of antenna printing verification.
The result of measuring many antennas including the prototype lens revealed that the
printing process is a remarkably accurate manufacturing method. Nearly all antennas were
printed with an inner diameter of 8 mm, within the tolerance threshold for the specification. Out
of twenty printed antennas, only 2 had defects associated with the arduous process manually
screen printing the antennas. These defects caused a slight smearing of ink into the center of
the lens. Since the sample size is very small, no statistically powerful conclusion can be made
regarding the accuracy of the printing process, but for our design, an error rate of 10% that
creates visibly distinguishable defects is very good.
Since all other circuit components will be laid on top of the antenna and outside the inner
circle of the contact lens, this test remains a good measure of determining how strongly the
circuitry will interfere with vision, if at all. A more in depth test could be performed with a
contemporary contact lens in order to verify the minimum clearance diameter of 9 mm. Using a
non-toxic marker, a circle of 9 mm would be drawn onto the surface of a contact lens prior to
application. Then, the subject would wear the lens and enter a dark room to ensure maximum
pupil dilation. If the subject is able to distinguish the colored circle in their vision, then our
chosen 9 mm threshold radius is too small.
37
Antenna Testing and Characterization
The three printed antenna designs were tested for performance and far-field
characterization using the microwave lab equipment and a test antenna.
Equipment:
17cm reference
antenna/transmitter
Anritsu MS4622B
Network Analyzer
CXA Signal Analyzer
Synthesized Sweep
Generator
38
Figure 3. Antennas under testing. AUT1, AUT2, AUT3 from left to right.
List of test parameters:
1. Gain
2. Impedance and Bandwidth
3. Polarization and Beam width
4. Radiation pattern (magnitude)
5. Efficiency
1. Gain Test
The transmitter has an unknown gain, which can be determined by testing the received
power of each AUT and reference antenna. This test will also determine the gains for all
3 AUTs. Using a set of four conservation of power equations and four unknowns, we
can calculate the gains of each antenna.
𝑃𝑅𝑟𝑒𝑓−1 = 𝑃𝑇𝑋 + 𝐺𝑟𝑒𝑓 + 𝐺1 − 𝑃𝐹
𝑃𝑅𝑟𝑒𝑓−2 = 𝑃𝑇𝑋 + 𝐺𝑟𝑒𝑓 + 𝐺2 − 𝑃𝐹
𝑃𝑅𝑟𝑒𝑓−3 = 𝑃𝑇𝑋 + 𝐺𝑟𝑒𝑓 + 𝐺3 − 𝑃𝐹
𝑃𝑅1−2 = 𝑃𝑇𝑋 + 𝐺1 + 𝐺2 − 𝑃𝐹
In order to accurately test the antennas under far-field condition, we must first determine
2
the far-field range, which is 𝑅 ≥ 2𝐷 ⁄𝜆, where D is the length of our source reference
antenna and λ is our operating wavelength, 12.5cm.
2(17𝑐𝑚)2⁄
𝑅≥
12.5𝑐𝑚
𝑅 ≥ 46.24𝑐𝑚
39
Figure 4. Antenna gain test setup for 2.4GHz far-field range of 50cm
We will use a distance of 50cm for the entirety of our tests. At 50cm, the path loss at 2.4
GHz is calculated with free space approximation.
4𝜋𝑑 2
)
𝐹𝑃(𝑑𝐵) = 10𝑙𝑜𝑔 (
𝜆
FP = 34dB
Our power transmission test setup includes the reference antenna mounted 50cm away
from the test antennas and transmitting 1dB from the frequency generator. The test
antennas are mounted on the SMA adapter for easy swapping. The measured received
power from the reference antenna to antenna 1 to be -44.77dBm, as seen in Figure 5.
The measured received power from the reference antenna to AUT 2 to be -42.19dBm,
as seen in Figure 6. The measured received power from the reference antenna to AUT
3 to be -44.57dBm, as seen in Figure 7. The measured received power from AUT 2 to
AUT 1 to be -57.55dBm, as seen in Figure 8. Before any calculations, we can already
tell that our reference antenna has a higher gain than AUT 2 when comparing the
received power at AUT 1.
40
Figure 5. Reference antenna transmitting 1dB to AUT 1 at 50cm
Figure 6. Reference antenna transmitting 1dB to AUT 2 at 50cm
41
Figure 7. Reference antenna transmitting 1dB to AUT 3 at 50cm
Figure 8. Antenna 2 transmitting 1dB to AUT 1 at 50cm
42
Solving the conservation of power equations with the received power data gives us the
gains for each antenna.
−44.77 = 1 + 𝐺𝑟𝑒𝑓 + 𝐺1 − 34
−42.19 = 1 + 𝐺𝑟𝑒𝑓 + 𝐺2 − 34
−44.57 = 1 + 𝐺𝑟𝑒𝑓 + 𝐺3 − 34
−57.55 = 1 + 𝐺1 + 𝐺2 − 34
𝐺𝑟𝑒𝑓 = 1.7950 𝑑𝐵𝑖, 𝐺1 = −13.565 𝑑𝐵𝑖, 𝐺2 = −10.985 𝑑𝐵𝑖, 𝐺3 = −13.365 𝑑𝐵𝑖
2. Impedance and Bandwidth Test
Using the Anritsu MS4622B Network Analyzer, we measure the antennas’ impedance
and bandwidth from their reflection coefficient (Γ), S11 parameter. The measured
impedance and reflection coefficient for UAT 1 obtained from figures 9 and 10. The
measured impedance and reflection coefficient for UAT 2 obtained from figures 11 and
12. The measured impedance and reflection coefficient for UAT 3 obtained from figures
13 and 14. These measured values and calculated power-ratio, VSWR, and percent
reflected power are compiled in Table 2 below.
Parameter
AUT1
AUT2
AUT3
S11 (Γ) [dB]
-5.722
-2.346
-3.476
Power-ratio
0.2678
0.5826
0.4492
VSWR [dB]
1.73
3.79
2.63
Reflected Power [%]
7.2
33.9
20.2
Impedance
19.744-j22.993
13.003-j49.497
194.772+j281.877
Table 2. Antenna parameters calculated from reflection coefficient
Sample calculations:
𝛤
𝑃𝑜𝑤𝑒𝑟 − 𝑟𝑎𝑡𝑖𝑜 = 1010
𝑉𝑆𝑊𝑅 =
1+𝑝𝑜𝑤𝑒𝑟−𝑟𝑎𝑡𝑖𝑜
1−𝑝𝑜𝑤𝑒𝑟−𝑟𝑎𝑡𝑖𝑜
𝑅𝑒𝑓𝑙𝑒𝑐𝑡𝑒𝑑 𝑃𝑜𝑤𝑒𝑟(%) = 100 ∗ |Γ|2
For AUT1, 10
−5.722
10
= 0.2678
For AUT1, 𝑉𝑆𝑊𝑅 =
1+0.2678
1−0.2678
= 1.73
For AUT1, Refl. Power = 100%*0.26782 = 7.2%
43
Figure 9. AUT 1 S11 smith chart
Figure 10. AUT 1 S11 magnitude
44
Figure 11. AUT 2 S11 smith chart
Figure 12. AUT 2 S11 magnitude
45
Figure 13. AUT 3 S11 smith chart
Figure 14. AUT 3 S11 magnitude
46
3. Polarization and Beam Width
We tested polarization of the AUTs by transmitting a 1dB Sine wave at 50cm with the
reference antenna and measuring the received power as the AUT is rotated along the xaxis. Received power remained constant at every angle. From this we can concluded
that the AUT and the reference antenna are circularly polarized, which contradicts our
theory that most dipole antennas are directional and our printed designs are basically
soiled dipoles. If either the reference antenna or AUT were directional, then we would
not be able to receive any signals. Thus, beam width measurements are not applicable
to circularly polarized antennas.
Figure 15. The AUT is rotated around the horizontal axis for polarization measurement
47
4. Radiation Pattern (magnitude)
To characterize radiation patterns for the AUTs, we transmit 1 dBm signal from
the AUT to the reference antenna at 65cm. We measure the received signal and then
rotate the AUT about the reference antenna, taking measurements every 20 degrees.
The setup required both antennas to be connected to the frequency generator or VNA
while orbiting the AUT. The setup was improved three times to find a consistent and
efficient method of orbiting the antennas.
Figure 16. The first attempt at measuring radiation pattern required one person to manually holding the mount and
ruler and 1 person to record the data. The mount was very heavy. This did not work.
48
Figure 17. The second setup used a stool to hold the mount instead of a person, but this ran into trouble when the
table impedes the testing of the other 180 degrees.
Figure 18. The last setup uses a swivel stool to rotate the reference antenna, which is attached to a meter stick,
around the AUT.
49
120°
60°
150°
60
50
30°
40
30
20
10
0
10
20
30
40
50
210°
60
330°
240°
300°
Figure 19. AUT1 Radiation pattern and measured received power
50
120°
60°
150°
80
30°
60
40
20
0
20
40
60
210°
80
330°
240°
300°
Figure 20. AUT2 Radiation pattern and measured received power
51
120°
60°
150°
80
30°
60
40
20
0
20
40
60
210°
80
330°
240°
300°
Figure 21. AUT3 radiation pattern and measured received power
From figures 19 and 21, both radiation patterns are similar since both antennas are based
around the same coupled loop design. AUT 2 has a much more circular radiation pattern as
seen in Figure 18.
52
5. Efficiency
Efficiency is calculated using measurements that we’ve already collected. The gain of
the transmitting reference antenna is 1.8dBi, and the signal is 1dBm. The required data for
calculation is shown in Table 3 below.
S11 (Γ) [dB]
Received power [dBm]
Received power [nW]
Reflected Power [%]
gain [dBi]
pathloss [dB]
pathloss [nW]
AUT1
AUT2
AUT3
-5.722 -2.346 -3.476
-43.66
-58.9
-38
279
1.3
158
7.2
33.9
20.2
13.656 10.985 13.365
-48.1
-45.5
-47.9
-78.1
-75.5
-77.9
Table 3. Measured power parameters for efficiency calculation
𝑃𝑜𝑢𝑡 = 𝑃𝑇𝑋 + 𝐺𝑟𝑒𝑓 = 1𝑑𝐵𝑚 + 1.8𝑑𝐵𝑖 = 2.8𝑑𝐵𝑚
𝑃65𝑐𝑚 = 𝑃𝑜𝑢𝑡 − 𝑃𝐹 = 2.8𝑑𝐵𝑚 − 48.1𝑑𝐵𝑚 = −45.3𝑑𝐵𝑚 = 3𝑛𝑊
279𝑒 −9
279𝑒 −9
𝑃𝑟𝑒𝑐𝑖𝑒𝑣𝑒𝑑
=
=
= 300.6𝑛𝑊
𝑃𝑖𝑛 =
1 − %𝑟𝑒𝑓𝑒𝑐𝑡𝑒𝑑 1 − 7.2% 92.8%
𝑃𝑜𝑢𝑡
3𝑛𝑊
𝜂=
=
= 1%
𝑃𝑖𝑛
300𝑛𝑊
AUT1
AUT2
AUT3
Efficiency
1%
Less than 1%
15.6%
Table 4. Summary of tested antenna efficiency
Our tested efficiency does for AUT2 and AUT3 is close to the theoretical values from our
simulation for Design 1 and 2. On the other hand, AUT1 fails to meet the expected efficiency
because it was not properly tuned to 2.4GHz during the test. The outer ring needed to be
shorted by scrubbing off the silver. Due to time constraints and failed attempts, we had to make
due with testing a close to-but-not-perfect antenna.
Contact Angle Testing
Contact angle testing was used to primarily determine the wettability of the surface of
PDMS, the exterior component of the final lens design. The materials used for this test included
a sample of PDMS encased PET (the finals lens, sans circuitry,) a VCA Optima contact angle
system, saline, and DI water. Figure 22 below depicts this test in progress.
53
Figure 22: Contact angle test in progress using VCA Optima system.
The results of these tests were very clear: PDMS is not nearly hyrophonic enough to
provide a comfortable, clear adhesion of the prototype lens to the human eye. Contact lenses
rely on the surface tension created by the tear layer between the eye and the lens to remain in
place. If a material with insufficient wettability is placed on the eye, the tear layer will glob up
and areas of the lens will be in direct contact with the eye, causing irritation and blurry vision.
Results from this test were computed and are presented in figure 23 below.
Figure 23: Bar chart of average contact angles for PET and PDMS, including standard error bars. (n = 15)
As can be seen in the figure above, PDMS has a contact angle of around 105o, far
exceeding the upper limit of 65o set in the final specification. The most straightforward way to
remedy this large contact angle would be the use of a surface hydrogel coating for contact lens.
Silicone hydrogels are contemporary materials developed in 1999 to solve the problem of
contact lens induced ocular hypoxia. They are highly permeable, and have a desirable contact
angle for use in contact lens applications. Traditionally, they are cut using a lathe from a solid
54
block of material. This would make applying to the surface of PET difficult. Recently, however, a
patent filed under the number WO2009073401 A2 proposes the creation of a novel silicon
hydrogel spray for use in tissue dressing application [20]. Should this technology come to
fruition, it would be an excellent substitute for PDMS as an exterior lens material.
Comfortability
Biocompatibility Assessment
The purpose of biocompatibility assessment was to determine whether the final design
improved upon previous designs which exposed PET to the surface of the eye. PET, as stated
in our preliminary analysis, has been shown to yield endocrine disruptive properties.
Additionally, PET is traditionally characterized a “hard” contact lens material, with a high elastic
modulus and characteristically rough surface. This makes PET fairly uncomfortable when
placed on the human eye for long periods of time. In our final design, we purposely
encapsulated PET with PDMS to circumvent some of the uncomfortable aspects of PET with a
softer, more flexible polymer. An illustration of this design is shown in figure 24 below.
Figure 24: PET lens base completely encapsulated by PDMS. In general, layer one is much thicker than layer two.
We define a material as biocompatible when it a) does no harm to the human body and
b) when the human body does no harm to it. While this definition appears archaic, it is the first
requirement that is of utmost importance. Irritation, hypoxia, and hormonal disruption are all
undesirable and would readily fall under the definition of “harmful” in our scenario. After
surveying tissue engineering literature, it was determined that PDMS is readily used in biological
scenarios and displays biocompatible qualities both in vitro [21] and in vivo [22]. Thus we can
conclude that it is suitable in preventing long term exposure to PET in physiological
temperatures. To strengthen this conclusion, in vitro assessment can be performed over a long
period of time to determine cell adhesion and cell toxicity properties. The former would involve
colonizing cells on the surface of substrate material and attempting to wash cells off at various
time points. The number of cells remaining versus the number of cells washed away can be
used as a metric for cell adhesion. The latter would involve placing a layer of agar over a pre
colonized layer of cells, and placing a layer of PDMS above the agar, allowing the PDMS to
diffuse slowly into the cell layer. Microscopy to examine cell death could then be used as a
metric for cellular toxicity in vitro.
55
Thickness Testing
The total thickness of the prototype was determined using calipers accurate to
3/1000th’s of an inch. The primary goal of this test was to verify that the after the overall
manufacturing process was completed, the lens thickness remained within the desired
specification. This test was performed knowing that the thickness of the final design would
greatly exceed the planned specification due to time and cost constraints. To account for
variability in thickness over the surface of the lens, 15 separate points were chosen for
thickness measurements. Figure 25 below contains the materials used in this test.
Figure 25: Materials used to characterize thickness of manufactured substrate and prototype.
From this test, it was determined that the overall thickness of the lens was about 800
microns with a standard error of 71.12 microns (n = 15.) As stated before, this far exceeds the
desired specification of 400 microns. The solution to this is a much smaller integrated circuit
fabrication process to achieve smaller feature sizes and an overall thinner integrated circuit.
This, ideally, would bring overall thickness down to around 450 microns.
Substrate Breathability Testing
As with biocompatibility, quantifying substrate breathability is difficult without expensive
equipment and large amounts of time. Systems built to determine the permeability of a material
to oxygen, such as the LabThink OX2/231 Oxygen Permeability Tester, are not readily available
for students and are designed for industrial lab usage. As such, the primary means of verifying
oxygen permeability for our substrate materials was through a survey of literature and research
on polymer oxygen permeability.
While PET is not permeable to oxygen whatsoever (a property which has earned it a
throne in the plastic bottling industry,) PDMS has a very high oxygen permeability at
physiological temperature [23]. Since PDMS encapsulates the PET and subsequently the lens,
oxygen can readily diffuse throughout the exterior layer and circumvent the PET boundary.
Additionally, areas which are covered by PET but do not contain any functional components can
potentially be removed in future iterations of the device, further improving oxygen permeability
of the lens. In order to strengthen information obtained in these surveys, simulations in
56
COMSOL can be performed to determine transient oxygen concentration on the ocular side of
the lens. These simulations, however, are not straightforward, and require large amounts of
data regarding nonlinear oxygen metabolism of ocular tissue as well as diffusion coefficients for
ocular tissues. Further complexity is introduced when considering the air layers between the
PET and PDMS.
Bioheat Analysis
Further bioheat analysis was performed via simulations in COMSOL by introducing a
nonlinear conductivity model for ocular tissue with thermally significant vessels. This method
was first described by Weinbaum et al. in a landmark study on the effects of blood flow on local
average tissue temperature [24]. In this method, tissues that are perfused with blood experience
a nonlinear increase in conductivity, which creates “resistance” to heat flow into the tissues. A
surface plot of the temperature distribution using this model is shown in figure 27 below.
Figure 27: Nonlinear conductivity bioheat model transient temperature analysis.
The most significant observation of this result is its similarity to the previous model’s
result. Using simply the Penne bioheat equation, a nearly identical temperature distribution was
achieved in preliminary analysis. There are a few likely causes for this similarity. The first cause
is the presence of a large artery at the bottom of the model that acts as an outflow condition.
When microvasculature is in close proximity to the large vessel, such as the one shown above,
the Penne bioheat equation tends to dominate, as the large vessel acts as a heat sink for which
excess thermal energy can exit the system. While a microvascular model was implemented in
the domain of the Iris near the cornea, the effect of the perfusion was minimal when compared
to blood flow via the retinal arteries. The second is the difficulty associated with measuring
microvascular perfusion rates in ocular tissues. Since the new model relies on using blood
perfusion in tissue to calculate new values of thermal conductivity, a constant value of perfusion
57
for each tissue is not sufficient. The value must be dynamic, based upon available experimental
data of transient blood flow in the human eye. In the time our team had to create and use this
model, developing such a realistic simulation of ocular blood flow was not feasible.
IX Impact, Conclusion, and Recommendations
At the onset of the Clear Active Contact Lens project, our group set out to do four things:
modify the project to include a medical perspective, establish a quantitative set of design
requirements based on this new perspective, develop a robust and safe process for
manufacturing the new product, and test and evaluate our products biomedical requirements.
We looked to previous designs and attempts at creating an active contact lens and ask
ourselves “how can we improve that, and if we can, how do we know it’s an improvement?” At
the end of the road, with a prototype in hand, it is safe to say that we have successfully created
a platform for an active contact lens that makes the device more sustainable, isolated, and safe.
We believe that the combination of biomedical ingenuity and novel functionality associated with
our design sends a strong message to future groups of students looking to make active contact
lenses. Our acceptance into the Engineering in Medicine and Biology Conference in Chicago,
Illinois, as well as the feedback from our sponsor confirms that our work has made a definitive
impact on the cutting edge research of active contact lenses.
At this point, we hope that future teams will take this work as a reference for the
embedding of a functional platform into a contemporary contact lenses. There is still much work
that needs to be done to ensure that our design meets the high standard set by the FDA for
medical devices. Our heartfelt recommendation is that the work we performed in verifying the
biomedical requirements of the device are improved upon, such that the device can one day be
examined for in vivo and clinical study. Active contact lenses represent a milestone in
communicating visual information to an increasingly diverse and technically apt population. It is
our desire to see a device similar to ours one day contributing to the overall safety of the
population.
58
Appendix A - Senor Project Analysis
1. Summary of Function Requirements
This project functions as a research and development stepping stone for future works in active
contact lens. It lays the groundwork for a manufacturing process of PET and PDMS lens that
are clear and comfortable. The IC design is a basis for signal processing and charge control at
for nano size devices. The antenna provides a stepping stone for effective and efficient signal
transfer at a millimeter scale as well as an attempt at wireless power transfer. The battery
research is an ongoing effort to produce a nonhazardous, flexible, and long lasting power
source for the contact lens. This proof of concept is one great leap towards a future technology.
2. Primary Constraints
The first requirement we could gather was that the lens needed to be clear. Our initial target
customer base is largely situated in hazardous waste control. In this occupation, unclouded
vision is absolutely necessary. The individual must be completely aware of what is in front of
them and around them through the use of visual recognition, since environmental suits will
obstruct other senses, such as smell, hearing, and touch. Another requirement is comfortability,
since the individual will more than likely not be able to perform any field adjustments on the lens
itself. This is by far one of the most difficult requirements to incorporate into device design. As
such, any materials used will need to be inspected for minimal adverse reactions, with the most
important being inflammation, when in contact with the body. Finally, the terminal requirement is
isolated functionality. To remain competitive in the young market place of active wearable
devices, the clear active contact lens must not require any input or command from the
customer. The device must maintain autonomy from the user so that he or she can focus on the
task at hand. When the device has meaningful data to provide, it will do so with a designated
stimulus, in our case an LED. Otherwise, it will remain off and out of the way.
3. Economic
We hope to present our project at the EMBS conference (Engineering Medicine Biology Society
of IEEE) in August 2014, BioCas, and ISSCC. We are also planning to submit our research to
publications such as IEEE Journal of Biomedical and Health Informatics, IEEE Transactions on
Biomedical Circuits and Systems, and IEEE Transactions on Consumer Electronics. There has
not been much research regarding clear circuits for contact lens applications so our work would
lay additional groundwork for future miniature circuit project. Table 3 shows our initial cost
projections. Our final cost of everything is $2568.89, which includes $500 for CNC tech hours
and materials, $1744.95 R&D graphene supercapacitor, and $323.94 Electrical components
and thin film battery. Much of our thin film printing process was donated by the GrC department.
A timeline of our project development is shown in Figure 1.
59
60
Figure 1: Updated Gantt chart to reflect most recent project status.
4. Environmental
The circuit uses translucent conductive ink printed on PET and PDMS sheet. These use natural
resources that the manufacturers have not released as public information. The contact lens
design has humans as the primary market and do not directly affect other species. Whether or
not the manufacturers’ of the components affect other species is not public knowledge.
61
5. Manufacturability
The lens manufacturing process was designed to be simple and robust. The process overview
involves four major steps: printing the lens, curving the lens, attaching the components, and
embedding the circuit in PDMS. To carry out this process, manufacturing techniques from
across disciplines were used. Screen-printing machines in the GrC department were used to
fabricate the antennas. The molds were designed using SolidWorks in the BMED computer lab
and manufactured in the Bonderson Machine shop. The PDMS was mixed and degassed using
the Micro fabrication lab facilities.
Outline of Manufacturing Process
1. Screen Print Antenna
2. Attach PET Substrate to Male Mold 1
3. Place Mold in oven at 180 degrees Celsius for 10 minutes
4. Attach components using thermo-epoxy
5. Use Mold 2 to build top layer of PDMS
6. Use Mold 3 to build bottom layer of PDMS
6. Sustainability
The contact lens is not a sustainable product and would not last longer than the life of the
printed battery. Future improvements to the design include additional sensors, transparent
transistors, more efficient antenna, and more pixels than a single LED. Because of the size
constraint, any upgrade to the design will reduce some existing aspect unless smaller
components are invented. Practicality of contact lens also requires transparency, which is not
feasible with some hardware. Health issues with being in direct contact with the eye also
constrains the heat, chemical, and radiation effects of the circuit design.
7. Ethical
This design has the user’s health as the highest priority. All specifications and possible side
effects of use of the contact lens will be disclosed with 100% transparency. We will anticipate
hazards and design safety measures. This project will be presented at EMBS conference to
share our design with other professionals in the printed electronics field.
8. Health and Safety
As a product that is in contact with the eye, safety is our top concern. The contact lens material
is designed to a sufficient degree of clearness for unimpaired visibility, flexibility for a nonirritable wear, biocompatibility to prevent harm to the eye. We’ve simulated the heat dissipation
at the peak of power consumption of the IC and confirmed its safety.
9. Development
This project has pushed thin film printing technology to being able to print fine line designs of
250μm. We have also streamlined the mold process of the materials.
62
Appendix B - References
References
[1] "Fatal Occupational Injuries in 2012 - Chart Package." Occupational Health and Safety
Administration. U.S. Department of Labor, 2012. Web. 20 Jan. 2014.
[2] Lingley, Andrew R., et al. "A single-pixel wireless contact lens display." Journal of
Micromechanics and Microengineering 21.12 (2011): 125014.
[3] J. Pandey, Y. Liao, and A. Lingley, “A fully integrated RF-powered contact lens with a single
element display,” … Circuits Syst. …, vol. 4, no. 6, pp. 454–461, 2010.
[4] B. Parviz, “For your eye only,” Spectrum, IEEE, no. september 2009, pp. 36–41, 2009
[5] H. Yao, A. J. Shum, M. Cowan, I. Lähdesmäki, and B. a Parviz, “A contact lens with
embedded sensor for monitoring tear glucose level.,” Biosens. Bioelectron., vol. 26, no. 7, pp.
3290–6, Mar. 2011.
[6] A. Kagie, D. K. Bishop, J. Burdick, J. T. La Belle, R. Dymond, R. Felder, and J. Wang,
“Flexible Rolled Thick‐ Film Miniaturized Flow‐ Cell for Minimally Invasive Amperometric
Sensing,” Electroanalysis, vol. 20, no. 14, pp. 1610–1614, Jul. 2008.
[7] S. Iguchi, H. Kudo, T. Saito, M. Ogawa, H. Saito, K. Otsuka, A. Funakubo, and K.
Mitsubayashi, “A flexible and wearable biosensor for tear glucose measurement.,” Biomed.
Microdevices, vol. 9, no. 4, pp. 603–9, Aug. 2007.
[8] P. C. Nicolson and J. Vogt, “Soft contact lens polymers: an evolution.” Biomaterials, vol. 22,
no. 24, pp. 3273–83, Dec. 2001.
[9] "Medical Devices." Radio Frequency Wireless Technology in. FDA, 13 Aug. 2013. Web. 26
Jan. 2014.
[10] Naoki Sadayori and Yuji Hotta "Polycarbodiimide having high index of refraction and
production method thereof" US patent 2004/0158021 A1 (2004)
[11] L. B. Kish, "End of Moore's Law: Thermal (Noise) Death of Integration in Micro and Nano
Electronics," Phys. Lett. A. 305 144 (2002).
[12]Gokul, K. C., D. B. Gurung, and P. R. Adhikary. "Effect of Blood Perfusion and Metabolism
in Temperature Distribution in Human Eye." Advances in Applied Mathematical Biosciences
4.1 (2013): 13-23.
63
[13] Cvetkovic, Mario, Dragan Poljak, and Andres Peratta. "Thermal modelling of the human eye
exposed to laser radiation." Software, Telecommunications and Computer Networks, 2008.
SoftCOM 2008. 16th International Conference on. IEEE, 2008.
[14] Liu, Chenguang, Zhenning Yu, David Neff, Aruna Zhamu, and Bor Z. Jang. "GrapheneBased Supercapacitor with an Ultrahigh Energy Density." Nano Letters 10.12 (2010): 4863-868.
Print.
[15] “Graphene.” Wikipedia. Wikimedia Foundation, 25 Jan. 2014. Web. 27 Jan. 2014.
[16] S.J. Wang, Y. Geng, Q. Zheng and J.k. Kim, “Fabrication of highly conducting
and transparent graphene films,” Carbon, 48, 1815–1823 (2010).
[17]Leonard Sax (2010). "Polyethylene Terephthalate May Yield Endocrine Disruptors".
Environmental Health Perspectives 118 (4): 445–8.
[18] R.M. Ford and C.S. Coulston, Design for Electrical and Computer Engineers, McGraw-Hill,
2007.
[19] El-Kady, Maher. "Laser Scribing of High-Performance and Flexible Graphene-Based
Electrochemical Capacitors - Supporting Online Material." Laser Scribing of High-Performance
and Flexible Graphene-Based Electrochemical Capacitors - Supporting Online Material. Science
Magazine, 16 Mar. 2012. Web. 3 Nov. 2013.
[20] McCrea, Keith R., Yuan Tian, and Robert S. Ward. "Silicone hydrogels for tissue adhesives
and tissue dressing applications." U.S. Patent Application 12/745,509.
[21] Seitz, Helmut, et al. "Biocompatibility of polyethylene terephthalate (Trevira< sup>®</sup>
hochfest) augmentation device in repair of the anterior cruciate ligament." Biomaterials 19.1
(1998): 189-196.
[22] Seitz, H., et al. "Biocompatibility of polyethylene terephthalate--PET--(Trevira strong)--an in
vivo study of the sheep knee." Biomedizinische Technik. Biomedical engineering 41.6 (1996):
178.
[23] 10] Merkel, T. C., et al. "Gas sorption, diffusion, and permeation in poly (dimethylsiloxane)."
Journal of Polymer Science Part B: Polymer Physics 38.3 (2000): 415-434.
[24] Weinbaum, S., and L. M. Jiji. "A new simplified bioheat equation for the effect of blood flow
on local average tissue temperature." Journal of biomechanical engineering 107.2 (1985): 131139.
[24]"Flat Spiral Coil Design Calculator." Flat Spiral Coil Design Calculator. Circuits, 9 Dec. 2010.
Web. 27 Jan. 2014.
64
Appendix C - Final Drawings
Figure 1: Drawing of Female Mold 1
65
Figure 2: Drawing of Female Mold 2.
66
Figure 3: Drawing of Female Mold 3.
67
Figure 4: Drawing of Male Mold 1.
68
Figure 5: Drawing of Male Mold 2.
69
Figure 6: Drawing of Male Mold 3.
Appendix D - List of Vendors, Contact Information, Pricing
●
Cymbet CBC005 Thin Film battery: Qty = 143, Total Cost = $250
Cymbet™ Corporation, 18326 Joplin Street NW, Elk River, MN, 55330-1773
Phone: 763.633.1780
●
●
Rohm Semiconductor Red LED: Qty = 10, Total cost = $4.20
Rohm Semiconductor ML8511 UV Sensor: Qty = 5, Total Cost = $38.30
ROHM Semiconductor U.S.A.,LLC
6020 Cornerstone Court West, Suite 320, San Diego, CA 92121 U.S.A.
TEL : +1-858-625-3600 FAX : +1-858-625-3640
●
Fairchild MMG3002MT1 RF amplifier: Qty = 3, Total Cost =$14.94
Fairchild Semiconductor Corporation
3030 Orchard Parkway
San Jose, CA 95134
U.S.A.
Tel: 408-822-2000
70
●
Maxim MAX2750EUA+ 2.4GHz oscillator: Qty = 5, Total Cost = $16.50
Corporate Headquarters
Maxim Integrated
160 Rio Robles
San Jose, CA 95134 USA
408-601-1000
One Industrial Way
2705 McMillan Ave, San Luis Obispo, CA 93401
(805) 596-0645
1x 1/2" carbide, 3 flute, flat end mill for aluminum (estimate ~$30ea)
1x 1/4" carbide, 3 flute, flat end mill for aluminum (estimate ~$25ea)
2x 1/8" carbide, 3 flute, flat end mill for aluminum (estimate ~$20ea)
2x 1/8" carbide, 3 flute, ball end mill for aluminum (estimate ~$15ea)
2x 1/16" carbide, 3 flute, ball end mill for aluminum (estimate ~$15ea)
McMaster Carr
9630 Norwalk Blvd.
Santa Fe Springs, CA 90670-2932
(562) 692-5911
1 8975K239Multipurpose 6061
2
Aluminum, 1" Thick X 3" each
Width X 1' Length
today 24.34
each
71
48.